The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm

被引:26
作者
Ishihara, Hironori [1 ]
Sugo, Yoshihiro [2 ]
Tsutsui, Masato [3 ]
Yamada, Takashige [4 ]
Sato, Tetsufumi [5 ]
Akazawa, Toshimasa [6 ]
Sato, Nobukazu [7 ]
Yamashita, Koichi [8 ]
Takeda, Junzo [4 ]
机构
[1] Hirosaki Univ, Grad Sch Med, Dept Anesthesiol, Hirosaki, Aomori 0368562, Japan
[2] Nihon Kohden Corp, Monitoring Technol Ctr, Tokyo, Japan
[3] Natl Def Med Coll, Dept Anesthesiol, Saitama, Japan
[4] Keio Univ, Sch Med, Dept Anesthesiol, Tokyo, Japan
[5] Okayama Univ, Sch Med, Dept Anesthesiol & Resuscitol, Okayama 700, Japan
[6] Juntendo Univ, Sch Med, Dept Anesthesiol & Pain Med, Tokyo 113, Japan
[7] Toho Univ, Dept Anesthesiol, Tokyo, Japan
[8] Kochi Med Sch, Dept Anesthesiol & Crit Care Med, Kochi, Japan
关键词
Cardiac output; Measurement technique; Non-invasive calibration; Pulse contour analysis; Pulse wave transit time; WAVE TRANSIT-TIME; THERMODILUTION; FORM;
D O I
10.1007/s10877-012-9384-7
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A new non-invasive continuous cardiac output (esCCO) monitoring system solely utilizing a routine cardiovascular monitor was developed, even though a reference cardiac output (CO) is consistently required. Subsequently, a non-invasive patient information CO calibration together with a new automated exclusion algorithm was implemented in the esCCO system. We evaluated the accuracy and trending ability of the new esCCO system. Either operative or postoperative data of a multicenter study in Japan for evaluation of the accuracy of the original version of esCCO system were used to develop the new esCCO system. A total of 207 patients, mostly cardiac surgical patients, were enrolled in the study. Data were manually reviewed to formulate a new automated exclusion algorithm with enhanced accuracy. Then, a new esCCO system based on a patient information calibration together with the automated exclusion algorithm was developed. CO measured with a new esCCO system was compared with the corresponding intermittent bolus thermodilution CO (ICO) utilizing statistical methods including polar plots analysis. A total of 465 sets of CO data obtained using the new esCCO system were evaluated. The difference in the CO value between the new esCCO and ICO was 0.34 +/- 1.50 (SD) L/min (95 % confidence limits of -2.60 to 3.28 L/min). The percentage error was 69.6 %. Polar plots analysis showed that the mean polar angle was -1.6 degrees and radial limits of agreement were +/- 53.3 degrees. This study demonstrates that the patient information calibration is clinically useful as ICO, but trending ability of the new esCCO system is not clinically acceptable as judged by percentage error and polar plots analysis, even though it's trending ability is comparable with currently available arterial waveform analysis methods.
引用
收藏
页码:465 / 471
页数:7
相关论文
共 17 条
[1]   Effect of the degree of tricuspid regurgitation on cardiac output measurements by thermodilution [J].
Balik, M ;
Pachl, J ;
Hendl, J .
INTENSIVE CARE MEDICINE, 2002, 28 (08) :1117-1121
[2]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[3]   Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery [J].
Breukers, Rose-Marieke B. G. E. ;
Sepehrkhouy, Shahrzad ;
Spiegelenberg, Stefan R. ;
Groeneveld, A. B. Johan .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2007, 21 (05) :632-635
[4]   Performance of a minimally invasive uncalibrated cardiac output monitoring system (Flotrac™/Vigileo™) in haemodynamically unstable patients [J].
Compton, F. D. ;
Zukunft, B. ;
Hoffmann, C. ;
Zidek, W. ;
Schaefer, J. -H. .
BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) :451-456
[5]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[6]   Assessment of Trending Ability of Cardiac Output Monitors by Polar Plot Methodology [J].
Critchley, Lester A. ;
Yang, Xiao X. ;
Lee, Anna .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2011, 25 (03) :536-546
[7]   A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac Output [J].
Critchley, Lester A. ;
Lee, Anna ;
Ho, Anthony M. -H. .
ANESTHESIA AND ANALGESIA, 2010, 111 (05) :1180-1192
[8]   A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors: Comparison with the continuous thermodilution method early cardiovascular monitors [J].
Ishihara H. ;
Okawa H. ;
Tanabe K. ;
Tsubo T. ;
Sugo Y. ;
Akiyama T. ;
Takeda S. .
Journal of Clinical Monitoring and Computing, 2004, 18 (5-6) :313-320
[9]   Comparison of Uncalibrated Arterial Pressure Waveform Analysis with Continuous Thermodilution Cardiac Output Measurements in Patients Undergoing Elective Off-Pump Coronary Artery Bypass Surgery [J].
Jeong, Yong Bo ;
Kim, Tae Hee ;
Roh, Young Jin ;
Choi, In Cheol ;
Suh, Jeong Hun .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2010, 24 (05) :767-771
[10]   Relationship between aortic-to-radiad arterial pressure gradient after cardiopulmonary bypass and changes in arterial elasticity [J].
Kanazawa, M ;
Fukuyama, H ;
Kinefuchi, Y ;
Takiguchi, M ;
Suzuki, T .
ANESTHESIOLOGY, 2003, 99 (01) :48-53