摘要
Objective. Magnetic resonance elastography (MRE) is widely adopted as a biomarker of liver fibrosis. However, in vivo MRE accuracy is difficult to assess. Approach. Finite element model (FEM) simulation was employed to evaluate liver MRE accuracy and inform methodological optimisation. MRE data was simulated in a 3D FEM of the human torso including the liver, and compared with spin-echo echo-planar imaging MRE acquisitions. The simulated MRE results were compared with the ground truth magnitude of the complex shear modulus ( divide G* divide ) for varying: (1) ground truth liver divide G* divide ; (2) simulated imaging resolution; (3) added noise; (4) data smoothing. Motion and strain-based signal-to-noise (SNR) metrics were evaluated on the simulated data as a means to select higher-quality voxels for preparation of acquired MRE summary statistics of divide G* divide . Main results. The simulated MRE accuracy for a given ground truth divide G* divide was found to be a function of imaging resolution, motion-SNR and smoothing. At typical imaging resolutions, it was found that due to under-sampling of the MRE wave-field, combined with motion-related noise, the reconstructed simulated divide G* divide could contain errors on the scale of the difference between liver fibrosis stages, e.g. 54% error for ground truth divide G* divide = 1 kPa. Optimum imaging resolutions were identified for given ground truth divide G* divide and motion-SNR levels. Significance. This study provides important knowledge on the accuracy and optimisation of liver MRE. For example, for motion-SNR <= 5, to distinguish between liver divide G* divide of 2 and 3 kPa (i.e. early-stage liver fibrosis) it was predicted that the optimum isotropic voxel size is 4-6 mm.
引用
收藏
页数:18
相关论文
共 50 条
  • [31] Magnetic Resonance Elastography as a Predictor of Insufficient Liver Enhancement on Gadoxetic Acid-Enhanced Hepatocyte-Phase Magnetic Resonance Imaging in Patients With Type C Hepatitis and Child-Pugh Class A Disease
    Motosugi, Utaroh
    Ichikawa, Tomoaki
    Muhi, Ali
    Sano, Katsuhiro
    Morisaka, Hiroyuki
    Ichikawa, Shintaro
    Araki, Tsutomu
    INVESTIGATIVE RADIOLOGY, 2012, 47 (10) : 566 - 570
  • [32] Role of magnetic resonance elastography in compensated and decompensated liver disease
    Asrani, Sumeet K.
    Talwalkar, Jayant A.
    Kamath, Patrick S.
    Shah, Vijay H.
    Saracino, Giovanna
    Jennings, Linda
    Gross, John B.
    Venkatesh, Sudhakar
    Ehman, Richard L.
    JOURNAL OF HEPATOLOGY, 2014, 60 (05) : 934 - 939
  • [33] Liver Magnetic Resonance Elastography: Focus on Methodology, Technique, and Feasibility
    Zerunian, Marta
    Masci, Benedetta
    Caruso, Damiano
    Pucciarelli, Francesco
    Polici, Michela
    Nardacci, Stefano
    De Santis, Domenico
    Iannicelli, Elsa
    Laghi, Andrea
    DIAGNOSTICS, 2024, 14 (04)
  • [34] Magnetic resonance elastography of liver: Technique, analysis, and clinical applications
    Venkatesh, Sudhakar K.
    Yin, Meng
    Ehman, Richard L.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 37 (03) : 544 - 555
  • [35] Magnetic resonance elastography
    Sack, I.
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2008, 133 (06) : 247 - 251
  • [36] Liver stiffness accuracy by magnetic resonance elastography in histologically proven non-alcoholic fatty liver disease patients: a Spanish cohort
    Romero, Carmen Lara
    Liang, Jia-Xu
    Fernandez-Lizaranzu, Isabel
    Herrojo, Javier Ampuero
    Castell, Javier
    Alba, Carmen del Prado
    Pascual, Inmaculada Dominguez
    Romero-Gomez, Manuel
    REVISTA ESPANOLA DE ENFERMEDADES DIGESTIVAS, 2023, 115 (04) : 162 - 167
  • [37] Normative values for magnetic resonance elastography-based liver stiffness in a healthy population
    Obrzut, Marzanna
    Atamaniuk, Vitaliy
    Obrzut, Bogdan
    Ehman, Richard
    Cholewa, Marian
    Rzucidlo, Mateusz
    Pozaruk, Andrii
    Gutkowski, Krzysztof
    POLISH ARCHIVES OF INTERNAL MEDICINE-POLSKIE ARCHIWUM MEDYCYNY WEWNETRZNEJ, 2019, 129 (05): : 321 - 326
  • [38] Quality Control of Magnetic Resonance Elastography Using Percent Measurable Liver Volume Estimation
    Ballard, David H.
    Ludwig, Daniel R.
    Fraum, Tyler J.
    Salter, Amber
    Narra, Vamsi R.
    Shetty, Anup S.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2022, 55 (06) : 1890 - 1899
  • [39] In vivo multifrequency magnetic resonance elastography of the human intervertebral disk
    Streitberger, Kaspar-Josche
    Diederichs, Gerd
    Guo, Jing
    Fehlner, Andreas
    Hamm, Bernd
    Braun, Juergen
    Sack, Ingolf
    MAGNETIC RESONANCE IN MEDICINE, 2015, 74 (05) : 1380 - 1387
  • [40] Diagnostic value of spleen stiffness by magnetic resonance elastography for prediction of esophageal varices in cirrhotic patients
    Jhang, Zih-En
    Wu, Kuan-Lin
    Chen, Chia-Bang
    Chen, Yao-Li
    Lin, Ping-Yi
    Chou, Chen-Te
    ABDOMINAL RADIOLOGY, 2021, 46 (02) : 526 - 533