Titanemia from total knee arthroplasty - A case resulting from a failed patellar component

被引:18
作者
vonSchroeder, HP
Smith, DC
Gross, AE
Pilliar, RM
Kandel, RA
Chernecky, R
Lugowski, SJ
机构
[1] Division of Orthopaedic Surgery, Mount Sinai Hospital, Toronto, Ont.
[2] Centre for Biomaterials, University of Toronto, Toronto, Ont.
[3] Division of Orthopaedic Surgery, Mt. Sinai Hospital, Toronto, Ont. M5G 1X5
关键词
titanium; total knee arthroplasty; metallosis; immune response; metallurgy;
D O I
10.1016/S0883-5403(96)80120-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The subject of this case report is a patient with elevated serum levels of titanium (77 parts/billion [ppb]; normal, 3.3 ppb) and vanadium (0.38 ppb; normal, 0.17 ppb) resulting from excessive wear of a metal-backed patellar component in a total knee arthroplasty. The patellar component was worn through both its polyethy lene and metal backing as a result of abnormal contact between the patellar and femoral components. Scanning electron microscopic examination of the ingrowth surface of the patellar component indicated that particle debonding occurred as a result of overloading of the sintered neck regions at the particle-substrate interface, suggesting a possible damage during initial insertion of the device, which may have predisposed it to loosening and abnormal contact with the femoral component. Wear particles resulted in staining of the tissues within the knee and an inflammatory and immune response in the synovium consisting of giant cells and T lymphocytes. The serum metal levels were reduced 22 weeks after replacing the patellar component; however, the titanium level was still slightly elevated (8 ppb).
引用
收藏
页码:620 / 625
页数:6
相关论文
共 26 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]   MALIGNANT FIBROUS HISTIOCYTOMA OF BONE AT THE SITE OF A TOTAL HIP-ARTHROPLASTY - A CASE-REPORT [J].
BAGOGRANELL, J ;
AGUIRRECANYADELL, M ;
NARDI, J ;
TALLADA, N .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1984, 66 (01) :38-40
[3]  
BARCKHAUS RH, 1990, METAL IONS BIOL MED, P284
[4]   METALLOSIS ASSOCIATED WITH A STABLE TITANIUM-ALLOY FEMORAL COMPONENT IN TOTAL HIP-REPLACEMENT - A CASE-REPORT [J].
BLACK, J ;
SHERK, H ;
BONINI, J ;
ROSTOKER, WR ;
SCHAJOWICZ, F ;
GALANTE, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (01) :126-130
[5]  
BRIEN WW, 1990, 36 ANN M ORTH RES SO
[6]  
DORR LD, 1990, CLIN ORTHOP RELAT R, P82
[7]   THE INCIDENCE OF CANCER FOLLOWING TOTAL HIP-REPLACEMENT [J].
GILLESPIE, WJ ;
FRAMPTON, CMA ;
HENDERSON, RJ ;
RYAN, PM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (04) :539-542
[8]  
HAHN TJ, 1989, BONE MINERAL RES, P219
[9]   RELEASE AND EXCRETION OF METAL IN PATIENTS WHO HAVE A TOTAL HIP-REPLACEMENT COMPONENT MADE OF TITANIUM-BASE ALLOY [J].
JACOBS, JJ ;
SKIPOR, AK ;
BLACK, J ;
URBAN, RM ;
GALANTE, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (10) :1475-1486
[10]  
JACOBS JJ, 1990, M SOC BIOM CHARL SC