Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study

被引:20
作者
Racine, Damien [1 ]
Ryckx, Nick [1 ]
Ba, Alexandre [1 ]
Becce, Fabio [2 ]
Viry, Anais [1 ]
Verdun, Francis R. [1 ]
Schmidt, Sabine [2 ]
机构
[1] Lausanne Univ Hosp, Inst Radiat Phys, Rue Grand Pre 1, CH-1007 Lausanne, Switzerland
[2] Lausanne Univ Hosp, Dept Diagnost & Intervent Radiol, Rue Bugnon 46, CH-1011 Lausanne, Switzerland
关键词
Abdominal computed tomography; Image quality; Model observer; Standardisation; Task-based assessment; STATISTICAL ITERATIVE RECONSTRUCTION; LOW-CONTRAST DETECTABILITY; FILTERED BACK-PROJECTION; LOW-TUBE-VOLTAGE; NOISE; OPTIMIZATION; PERFORMANCE; RESOLUTION; ALGORITHM; PHANTOM;
D O I
10.1007/s00330-018-5518-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveWe investigated the variability in diagnostic information inherent in computed tomography (CT) images acquired at 68 different CT units, with the selected acquisition protocols aiming to answer the same clinical question.MethodsAn anthropomorphic abdominal phantom with two optional rings was scanned on 68 CT systems from 62 centres using the local clinical acquisition parameters of the portal venous phase for the detection of focal liver lesions. Low-contrast detectability (LCD) was assessed objectively with channelised Hotelling observer (CHO) using the receiver operating characteristic (ROC) paradigm. For each lesion size, the area under the ROC curve (AUC) was calculated and considered as a figure of merit. The volume computed tomography dose index (CTDIvol) was used to indicate radiation dose exposure.ResultsThe median CTDIvol used was 5.8 mGy, 10.5 mGy and 16.3 mGy for the small, medium and large phantoms, respectively. The median AUC obtained from clinical CT protocols was 0.96, 0.90 and 0.83 for the small, medium and large phantoms, respectively.ConclusionsOur study used a model observer to highlight the difference in image quality levels when dealing with the same clinical question. This difference was important and increased with growing phantom size, which generated large variations in patient exposure. In the end, a standardisation initiative may be launched to ensure comparable diagnostic information for well-defined clinical questions. The image quality requirements, related to the clinical question to be answered, should be the starting point of patient dose optimisation.Key Points center dot Model observers enable to assess image quality objectively based on clinical tasks.center dot Objective image quality assessment should always include several patient sizes.center dot Clinical diagnostic image quality should be the starting point for patient dose optimisation.center dot Dose optimisation by applying DRLs only is insufficient for ensuring clinical requirements.
引用
收藏
页码:5203 / 5210
页数:8
相关论文
共 34 条
  • [1] Human- and model-observer performance in ramp-spectrum noise: effects of regularization and object variability
    Abbey, CK
    Barrett, HH
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2001, 18 (03): : 473 - 488
  • [2] [Anonymous], 1993, INTRO BOOTSTRAP
  • [3] Contrast-to-Noise Ratio and Low-Contrast Object Resolution on Full- and Low-Dose MDCT: SAFIRE Versus Filtered Back Projection in a Low-Contrast Object Phantom and in the Liver
    Baker, Mark E.
    Dong, Frank
    Primak, Andrew
    Obuchowski, Nancy A.
    Einstein, David
    Gandhi, Namita
    Herts, Brian R.
    Purysko, Andrei
    Remer, Erick
    Vachani, Neil
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (01) : 8 - 18
  • [4] Evaluation of the channelized Hotelling observer with an internal-noise model in a train-test paradigm for cardiac SPECT defect detection
    Brankov, Jovan G.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (20) : 7159 - 7182
  • [5] Impact of Reduced Patient Life Expectancy on Potential Cancer Risks from Radiologic Imaging
    Brenner, David J.
    Shuryak, Igor
    Einstein, Andrew J.
    [J]. RADIOLOGY, 2011, 261 (01) : 193 - 198
  • [6] Noise-reducing algorithms do not necessarily provide superior dose optimisation for hepatic lesion detection with multidetector CT
    Dobeli, K. L.
    Lewis, S. J.
    Meikle, S. R.
    Thiele, D. L.
    Brennan, P. C.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1023)
  • [7] Reducing the Radiation Dose for CT Colonography Using Adaptive Statistical Iterative Reconstruction: A Pilot Study
    Flicek, Kristina T.
    Hara, Amy K.
    Silva, Alvin C.
    Wu, Qing
    Peter, Mary B.
    Johnson, C. Daniel
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) : 126 - 131
  • [8] Combination of a Low-Tube-Voltage Technique With Hybrid Iterative Reconstruction (iDose) Algorithm at Coronary Computed Tomographic Angiography
    Funama, Yoshinori
    Taguchi, Katsuyuki
    Utsunomiya, Daisuke
    Oda, Seitaro
    Yanaga, Yumi
    Yamashita, Yasuyuki
    Awai, Kazuo
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2011, 35 (04) : 480 - 485
  • [9] Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies
    Habibzadeh, M. A.
    Ay, M. R.
    Asl, A. R. Kamali
    Ghadiri, H.
    Zaidi, H.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (03): : 191 - 199
  • [10] Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility Study
    Hara, Amy K.
    Paden, Robert G.
    Silva, Alvin C.
    Kujak, Jennifer L.
    Lawder, Holly J.
    Pavlicek, William
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (03) : 764 - 771