Unicondylar Knee Retrieval Analysis

被引:35
作者
Manson, Theodore T. [2 ]
Kelly, Natalie H. [1 ]
Lipman, Joseph D. [1 ]
Wright, Timothy M. [1 ]
Westrich, Geoffrey H. [1 ]
机构
[1] Hosp Special Surg, Dept Adult Reconstruct & Joint Replacement, Weill Cornell Med Coll, New York, NY 10021 USA
[2] Univ Maryland, Dept Orthopaed, R Adams Cowley Shock Trauma Ctr, Baltimore, MD 21201 USA
关键词
articular wear; mobile-bearing implants; delamination; unicondylar knee; retrieval analysis; ARTHROPLASTY; BEARINGS;
D O I
10.1016/j.arth.2010.05.004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Unicondylar knee arthroplasty (UKA) is considered an alternative to total knee arthroplasty for patients who have arthritis limited to one compartment of the knee. This study examined surface damage of 3 contemporary UKA designs that were retrieved at revision surgery. Two of the UKA designs were fixed bearing and one was mobile bearing. Demographic information was collected, as well as information about the implants used at revision surgery. Articular surface damage was greater in the fixed-bearing designs as compared to the mobile bearing, although the mobile-bearing implants had significantly shorter length of implantation. Backside damage was also graded for the mobile bearing and when combined with articular wear resulted in overall damage scores higher than both fixed-bearing designs. The fixed-bearing designs showed delamination and surface deformation, whereas the mobile bearing had no evidence of these damage modes. However, mobile-bearing components showed other types of wear, and significant wear damage was present on the bearing surfaces of the mobile-bearing implants despite a short time of implantation. At the time of conversion to a total knee arthroplasty, more than 50% of cases required the use of stems, augments, or constrained inserts for the tibial reconstruction. In conclusion, wear modes differed among UKA prosthesis designs. Revision of a UKA to a total knee arthroplasty remains complex with the tibial preparation more complicated than in the primary setting.
引用
收藏
页码:108 / 111
页数:4
相关论文
共 11 条
[1]   Results of unicompartmental knee arthroplasty at a minimum of ten years of follow-up [J].
Berger, RA ;
Meneghini, RM ;
Jacobs, JJ ;
Sheinkop, MB ;
Della Valle, CJ ;
Rosenberg, AG ;
Galante, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (05) :999-1006
[2]   Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty [J].
Collier, Matthew B. ;
Engh, C. Anderson, Jr. ;
McAuley, James P. ;
Engh, Gerard A. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (06) :1306-1314
[3]   RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES [J].
HOOD, RW ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05) :829-842
[4]   A 5-YEAR TO 7-YEAR FOLLOW-UP OF UNICONDYLAR ARTHROPLASTY [J].
INSALL, J ;
AGLIETTI, P .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (08) :1329-1337
[5]  
Insall J, 1976, CLIN ORTHOP RELAT RE, V83
[6]   UNICOMPARTMENTAL TIBIOFEMORAL RESURFACING ARTHROPLASTY [J].
LASKIN, RS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1978, 60 (02) :182-185
[7]   Medial unicompartmental knee arthroplasty with the Miller-Galante prosthesis [J].
Naudie, D ;
Guerin, J ;
Parker, DA ;
Bourne, RB ;
Rorabeck, CH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (09) :1931-1935
[8]   REVISION TOTAL KNEE ARTHROPLASTY FOR FAILED UNICOMPARTMENTAL REPLACEMENT [J].
PADGETT, DE ;
STERN, SH ;
INSALL, JN .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (02) :186-190
[9]  
PRICE AJ, 2005, CLIN ORTHOP RELAT RE, V171
[10]   Wear of congruent meniscal bearings in unicompartmental knee arthroplasty - A retrieval study of 16 specimens [J].
Psychoyios, V ;
Crawford, RW ;
O'Connor, JJ ;
Murray, DW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (06) :976-982