Biopsy guided by real-time sonography fused with MRI: A phantom study

被引:17
作者
Ewertsen, Caroline [1 ]
Grossjohann, Hanne Sonder [1 ]
Nielsen, Kristina Rue [1 ]
Torp-Pedersen, Soren [2 ]
Nielsen, Michael Bachmann [1 ]
机构
[1] Univ Copenhagen Hosp, Rigshosp, Dept Radiol, Sect Ultrasound 10 4123, DK-2100 Copenhagen OE, Denmark
[2] Frederiksberg Univ Hosp, Parker Inst, Frederiksberg, Denmark
关键词
biopsy; image fusion; intervention; MRI; sonography;
D O I
10.2214/AJR.07.2587
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of our study was to test the accuracy of sonographically guided biopsies in a phantom of structures not visible on sonography but shown on MRI by using commercially available sonography systems with image fusion software. MATERIALS AND METHODS. A previously recorded MRI examination from a custom-made phantom was loaded into the sonography system. The phantom contained spheres that were invisible to sonography and contained red dye. The red dye was visible in the biopsy if it was successful. The images were coregistered using structures visible on both sonography and MRI, and biopsies were taken. The biopsy procedure was continued until a biopsy was successful, and the number of needle passes and time spent were registered. RESULTS. A total of 130 targets were hit. Ten minutes was used for loading the MRI data set and the coregistration; 94 of the 130 biopsies (72.3%) were successful at the first needle pass. The median number of needle passes until a successful biopsy was obtained was one (range, 1 - 7). CONCLUSION. The described method was successful in obtaining an adequate sample in a phantom.
引用
收藏
页码:1671 / 1674
页数:4
相关论文
共 13 条
[1]   Three-dimensional reconstruction of the coronary artery wall by image fusion of intravascular ultrasound and bi-plane angiography [J].
Cothren, RM ;
Shekhar, R ;
Tuzcu, EM ;
Nissen, SE ;
Cornhill, JF ;
Vince, DG .
INTERNATIONAL JOURNAL OF CARDIAC IMAGING, 2000, 16 (02) :69-85
[2]  
Ewertsen C, 2006, ULTRASCHALL MED, V27, P129
[3]   Image guided percutaneous splenic interventions [J].
Kang, Mandeep ;
Kalra, Naveen ;
Gulati, Madhu ;
Lal, Anupam ;
Kochhar, Rohit ;
Rajwanshi, Arvind .
EUROPEAN JOURNAL OF RADIOLOGY, 2007, 64 (01) :140-146
[4]   Real time MRI-ultrasound image guided stereotactic prostate biopsy [J].
Kaplan, I ;
Oldenburg, NE ;
Meskell, P ;
Blake, M ;
Church, P ;
Holupka, EJ .
MAGNETIC RESONANCE IMAGING, 2002, 20 (03) :295-299
[5]   Percutaneous liver biopsy: A cost-benefit analysis comparing sonographic and CT guidance [J].
Kliewer, MA ;
Sheafor, DH ;
Paulson, EK ;
Helsper, RS ;
Hertzberg, BS ;
Nelson, RC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (05) :1199-1202
[6]  
Lindseth F, 2001, INT CONGR SER, V1230, P247
[7]   Is the image guidance of ultrasonography beneficial for neurosurgical routine? [J].
Miller, Dorothea ;
Heinze, Stefan ;
Tirakotai, Wuttipong ;
Bozinov, Oliver ;
Sueruecue, Oguzkan ;
Benes, Ludwig ;
Bertalanffy, Helmut ;
Sure, Ulrich .
SURGICAL NEUROLOGY, 2007, 67 (06) :579-588
[8]   MRI/TRUS data fusion for prostate brachytherapy - Preliminary results [J].
Reynier, C ;
Troccaz, J ;
Fourneret, P ;
Dusserre, A ;
Gay-Jeune, C ;
Descotes, JL ;
Bolla, M ;
Giraud, JY .
MEDICAL PHYSICS, 2004, 31 (06) :1568-1575
[9]   Image fusion of MR images and real-time ultrasonography: evaluation of fusion accuracy combining two commercial instruments, a neuronavigation system and a ultrasound system [J].
Schlaier, JR ;
Warnat, J ;
Dorenbeck, U ;
Proescholdt, M ;
Schebesch, KM ;
Brawanski, A .
ACTA NEUROCHIRURGICA, 2004, 146 (03) :271-277
[10]  
Selli C, 2007, RADIOL MED, V112, P74, DOI 10.1007/s11547-007-0122-z