Comparison of Simulated Parallel Transmit Body Arrays at 3 T Using Excitation Uniformity, Global SAR, Local SAR, and Power Efficiency Metrics

被引:61
作者
Guerin, Bastien [1 ]
Gebhardt, Matthias [2 ]
Serano, Peter [1 ]
Adalsteinsson, Elfar [3 ,4 ]
Hamm, Michael [2 ]
Pfeuffer, Josef [2 ]
Nistler, Juergen [2 ]
Wald, Lawrence L. [1 ,4 ]
机构
[1] Massachusetts Gen Hosp, Martinos Ctr Biomed Imaging, Dept Radiol, Charlestown, MA USA
[2] Siemens Healthcare, Erlangen, Germany
[3] MIT, Dept Elect Engn & Comp Sci, Cambridge, MA 02139 USA
[4] MIT, Harvard Mit Div Hlth Sci & Technol, Cambridge, MA 02139 USA
关键词
Local SAR; global SAR; power; parallel transmission; pTx coils; excitation fidelity; spokes; ABSORPTION RATE; COIL; DESIGN; OPTIMIZATION; CHANNELS; PULSES;
D O I
10.1002/mrm.25243
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeWe compare the performance of eight parallel transmit (pTx) body arrays with up to 32 channels and a standard birdcage design. Excitation uniformity, local specific absorption rate (SAR), global SAR, and power metrics are analyzed in the torso at 3 T for radiofrequency (RF)-shimming and 2-spoke excitations. MethodsWe used a fast cosimulation strategy for field calculation in the presence of coupling between transmit channels. We designed spoke pulses using magnitude least squares optimization with explicit constraint of SAR and power and compared the performance of the different pTx coils using the L-curve method. ResultsPTx arrays outperformed the conventional birdcage coil in all metrics except peak and average power efficiency. The presence of coupling exacerbated this power efficiency problem. At constant excitation fidelity, the pTx array with 24 channels arranged in three z-rows could decrease local SAR more than 4-fold (2-fold) for RF-shimming (2-spoke) compared to the birdcage coil for pulses of equal duration. Multi-row pTx coils had a marked performance advantage compared to single row designs, especially for coronal imaging. ConclusionPTx coils can simultaneously improve the excitation uniformity and reduce SAR compared to a birdcage coil when SAR metrics are explicitly constrained in the pulse design. Magn Reson Med 73:1137-1150, 2015. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1137 / 1150
页数:14
相关论文
共 37 条
  • [1] Degenerate mode band-pass birdcage coil for accelerated parallel excitation
    Alagappan, Vijayanand
    Nistler, Juergen
    Adalsteinsson, Elfar
    Setsompop, Kawin
    Fontius, Ulrich
    Zelinski, Adam
    Vester, Markus
    Wiggins, Graham C.
    Hebrank, Franz
    Renz, Wolfgang
    Schmitt, Franz
    Wald, Lawrence L.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (06) : 1148 - 1158
  • [2] Boyd S., 2004, CONVEX OPTIMIZATION
  • [3] Breuer FA, 2005, P 13 ANN M ISMRM MIA, P2668
  • [4] Impact of number of channels on RF shimming at 3T
    Childs, Alexander S.
    Malik, Shaihan J.
    O'Regan, Declan P.
    Hajnal, Joseph V.
    [J]. MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2013, 26 (04) : 401 - 410
  • [5] Local Specific Absorption Rate Control for Parallel Transmission by Virtual Observation Points
    Eichfelder, Gabriele
    Gebhardt, Matthias
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (05) : 1468 - 1476
  • [6] A radiofrequency coil to facilitate B1+ shimming and parallel imaging acceleration in three dimensions at 7 T
    Gilbert, Kyle M.
    Curtis, Andrew T.
    Gati, Joseph S.
    Klassen, L. Martyn
    Menon, Ravi S.
    [J]. NMR IN BIOMEDICINE, 2011, 24 (07) : 815 - 823
  • [7] Graesslin I, 2008, P ISMRM TOR CAN, P74
  • [8] Graesslin I., 2010, Proceedings of the 18th Scientific Meeting of ISMRM, P4932
  • [9] Spatial domain method for the design of RF pulses in multicoil parallel excitation
    Grissom, William
    Yip, Chun-yu
    Zhang, Zhenghui
    Stenger, V. Andrew
    Fessler, Jeffrey A.
    Noll, Douglas C.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (03) : 620 - 629
  • [10] Local Specific Absorption Rate (SAR), Global SAR, Transmitter Power, and Excitation Accuracy Trade-Offs in Low Flip-Angle Parallel Transmit Pulse Design
    Guerin, Bastien
    Gebhardt, Matthias
    Cauley, Steven
    Adalsteinsson, Elfar
    Wald, Lawrence L.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (04) : 1446 - 1457