Monte Carlo calculation of scattered radiation from applicators in low energy clinical electron beams

被引:0
|
作者
Jabbari, Nasrollah
Hashemi-Malayeri, Bijan [1 ]
Farajollahi, Ali Reza
Kazemnejad, Anoshiravan
机构
[1] Tarbiat Modares Univ, Dept Med Phys, Tehran, Iran
[2] Tarbiat Modares Univ, Dept Biostat, Tehran, Iran
关键词
Monte Carlo; radiation therapy; medical linear accelerator; electron applicator; scattered radiation; treatment planning;
D O I
暂无
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In radiotherapy with electron beams, scattered radiation from an electron applicator influences the dose distribution in the patient. The contribution of this radiation to the patient dose is significant, even in modern accelerators. In most of radiotherapy treatment planning systems, this component is not explicitly included. In addition, the scattered radiation produced by applicators varies based on the applicator design as well as the field size and distance from the applicators. The aim of this study was to calculate the amount of scattered dose contribution from applicators. We also tried to provide an extensive set of calculated data that could be used as input or benchmark data for advanced treatment planning systems that use Monte Carlo algorithms for dose distribution calculations. Electron beams produced by a NEPTUN 10PC medical linac were modeled using the BEAMnrc system. Central axis depth dose curves of the electron beams were measured and calculated, with and without the applicators in place, for different field sizes and energies. The scattered radiation from the applicators was determined by subtracting the central axis depth dose curves obtained without the applicators from that with the applicator. The results of this study indicated that the scattered radiation from the electron applicators of the NEPTUN 10PC is significant and cannot be neglected in advanced treatment planning systems. Furthermore, our results showed that the scattered radiation depends on the field size and decreases almost linearly with depth.
引用
收藏
页码:97 / 103
页数:7
相关论文
共 50 条
  • [1] Evaluation of a new commercial Monte Carlo dose calculation algorithm for electron beams
    Vandervoort, Eric J.
    Tchistiakova, Ekaterina
    La Russa, Daniel J.
    Cygler, Joanna E.
    MEDICAL PHYSICS, 2014, 41 (02) : 021711
  • [2] Evaluation of the RayStation electron Monte Carlo dose calculation algorithm
    Richmond, Neil
    Allen, Vincent
    Wyatt, Jonathan
    Codling, Richard
    MEDICAL DOSIMETRY, 2020, 45 (02) : 159 - 167
  • [3] A Monte Carlo study of the energy spectra and transmission characteristics of scattered radiation from x-ray computed tomography
    Platten, David John
    JOURNAL OF RADIOLOGICAL PROTECTION, 2014, 34 (02) : 445 - 456
  • [4] Monte Carlo calculation of the dose distributions of two 106Ru eye applicators
    Sánchez-Reyes, A
    Tello, JJ
    Guix, B
    Salvat, F
    RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) : 191 - 196
  • [5] Monte Carlo modelling of clinical accelerator beams and estimation of primary electron beam parameters
    Horova, S.
    Judas, L.
    RADIOPROTECTION, 2018, 53 (01) : 61 - 66
  • [6] Monte Carlo treatment planning for photon and electron beams
    Reynaert, N.
    van der Marck, S. C.
    Schaart, D. R.
    Van der Zee, W.
    Van Vliet-Vroegindeweij, C.
    Tomsej, M.
    Jansen, J.
    Heijmen, B.
    Coghe, M.
    De Wagter, C.
    RADIATION PHYSICS AND CHEMISTRY, 2007, 76 (04) : 643 - 686
  • [7] Monte Carlo calculation of electron Rutherford backscattering spectra and high-energy reflection electron energy loss spectra
    Li, Y. G.
    Ding, Z. J.
    Zhang, Z. M.
    Tokesi, K.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2009, 267 (02) : 215 - 220
  • [8] Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
    Siantar, CLH
    Walling, RS
    Daly, TP
    Faddegon, B
    Albright, N
    Bergstrom, P
    Chuang, C
    Garrett, D
    House, RK
    Knapp, D
    Wieczorek, DJ
    Verhey, LJ
    MEDICAL PHYSICS, 2001, 28 (07) : 1322 - 1337
  • [9] Performance of a Commercial Macro Monte Carlo Dose Calculation Algorithm for Determining Output Factors of Clinical Electron Fields
    Popple, Richard A.
    Brezovich, Ivan A.
    Pareek, Prem N.
    Duan, Jun
    Shen, Sui
    Wu, Xingen
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2009, 8 (04) : 307 - 314
  • [10] Monte Carlo calculation of the TG-43 dosimetry parameters for the INTRABEAM source with spherical applicators
    Alvarez, David Santiago Ayala
    Watson, Peter G. F.
    Popovic, Marija
    Heng, Veng Jean
    Evans, Michael D. C.
    Seuntjens, Jan
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (21)