Analysis of Run-in-Stage Wear Behavior and Contact Mechanics of Metal-on-Metal Hip Joint Bearings with Different Radial Clearances

被引:14
|
作者
Chen, Yan [1 ]
Li, Yunping [2 ,3 ]
Kurosu, Shingo [3 ]
Meng, Qingen [4 ]
Tang, Ning [3 ]
Koizumi, Yuichiro [3 ]
Chiba, Akihiko [3 ]
机构
[1] Tohoku Univ, Dept Mat Sci & Engn, Sendai, Miyagi 9808579, Japan
[2] Cent S Univ, Mat Sci & Engn, Changsha 410083, Hunan, Peoples R China
[3] Tohoku Univ, Inst Mat Res, Sendai, Miyagi 9808577, Japan
[4] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
关键词
joint prostheses; CoCrMo alloy; radial clearance; contact mechanics; ELASTOHYDRODYNAMIC LUBRICATION; RESURFACING ARTHROPLASTY; SIMULATOR; DIAMETER; REPLACEMENTS; PROSTHESES; DEBRIS; IMPLANTS; SURFACES; FRICTION;
D O I
10.2320/matertrans.M2014440
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
We systematically elucidate the effects of radial clearance on the wear behavior of hip joint prostheses bearings during run-in-stage. The results indicated that bearings with smaller radial clearances exhibited lower wear rate and less abrasive wear characterized by mild surface roughness and sphericity. The contact mechanics and lubrication regime of MOM bearings with different radial clearances were also analyzed, indicating that bearings with radial clearance of similar to 20 mu m was more beneficial for small contact pressures even the enlarged clearance made the bearings operate under a full-fluid-to-mixed lubrication during wear. The agreement between the experimental results of wear rate (mm(3)/Mc) and contact mechanics and lubrication analysis indicated that the radial clearance significantly affected the wear behavior of the MOM bearings.
引用
收藏
页码:826 / 834
页数:9
相关论文
共 45 条
  • [11] Explant analysis of the Biomet Magnum/ReCap metal-on-metal hip joint
    Scholes, S. C.
    Hunt, B. J.
    Richardson, V. M.
    Langton, D. J.
    Smith, E.
    Joyce, T. J.
    BONE & JOINT RESEARCH, 2017, 6 (02): : 113 - 122
  • [12] Wear of ceramic coated metal-on-metal bearings used for hip replacement
    Gispert, M. P.
    Serro, A. P.
    Colaco, R.
    Pires, E.
    Saramago, B.
    WEAR, 2007, 263 : 1060 - 1065
  • [13] Analysis of contact mechanics in McKee-Farrar metal-on-metal hip implants
    Yew, A
    Jagatia, M
    Ensaff, H
    Jin, ZM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2003, 217 (H5) : 333 - 340
  • [14] Determination of Low Wear Rates in Metal-On-Metal Hip Joint Replacements Based on Ultra Trace Element Analysis in Simulator Studies
    Kretzer, J. P.
    Krachler, M.
    Reinders, J.
    Jakubowitz, E.
    Thomsen, M.
    Heisel, C.
    TRIBOLOGY LETTERS, 2010, 37 (01) : 23 - 29
  • [15] Effect of microseparation on contact mechanics in metal-on-metal hip replacementsA finite element analysis
    Liu, Feng
    Williams, Sophie
    Fisher, John
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2015, 103 (06) : 1312 - 1319
  • [16] Influence of Lubrication Performance on Wear Factor in Metal-on-Metal Hip Joint Replacement using Numerical Analysis
    Harun, Muhamad Noor
    Jin, Zhongmin
    Syahrom, Ardiyansyah
    INTERNATIONAL TRIBOLOGY CONFERENCE MALAYSIA 2013, 2013, 68 : 95 - 101
  • [17] Quantitative analysis of wear and wear debris from metal-on-metal hip prostheses tested In a physiological hip joint simulator
    Firkins, PJ
    Tipper, JL
    Saadatzadeh, MR
    Ingham, E
    Stone, MH
    Farrar, R
    Fisher, J
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2001, 11 (02) : 143 - 157
  • [18] Effect of head size on wear properties of metal-on-metal bearings of hip prostheses, and comparison with wear properties of metal-on-polyethylene bearings using hip simulator
    Okazaki, Yoshimitsu
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 31 : 152 - 163
  • [19] Prediction of contact mechanics in metal-on-metal Total Hip Replacement for parametrically comprehensive designs and loads
    Donaldson, Finn E.
    Nyman, Edward, Jr.
    Coburn, James C.
    JOURNAL OF BIOMECHANICS, 2015, 48 (10) : 1828 - 1835
  • [20] Comparison of contact mechanics between a total hip replacement and a hip resurfacing with a metal-on-metal articulation
    Liu, F
    Udofia, IJ
    Jin, ZM
    Hirt, F
    Rieker, C
    Roberts, P
    Grigoris, P
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2005, 219 (07) : 727 - 732