The use of megavoltage CT (MVCT) images for dose recomputations

被引:144
作者
Langen, KM [1 ]
Meeks, SL
Poole, DO
Wagner, TH
Willoughby, TR
Kupelian, PA
Ruchala, KJ
Haimerl, J
Olivera, GH
机构
[1] MD Anderson Canc Ctr, Dept Radiat Oncol, Orlando, FL 32806 USA
[2] TomoTherapy Inc, Madison, WI USA
关键词
D O I
10.1088/0031-9155/50/18/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Megavoltage CT (MVCT) images of patients are acquired daily on a helical tomotherapy unit (TomoTherapy, Inc., Madison, WI). While these images are used primarily for patient alignment, they can also be used to recalculate the treatment plan for the patient anatomy of the day. The use of MVCT images for dose computations requires a reliable CT number to electron density calibration curve. In this work, we tested the stability of the MVCT numbers by determining the variation of this calibration with spatial arrangement of the phantom, time and MVCT acquisition parameters. The two calibration curves that represent the largest variations were applied to six clinical MVCT images for recalculations to test for dosimetric uncertainties. Among the six cases tested, the largest difference in any of the dosimetric endpoints was 3.1% but more typically the dosimetric endpoints varied by less than 2%. Using an average CT to electron density calibration and a thorax phantom, a series of end-to-end tests were run. Using a rigid phantom, recalculated dose volume histograms (DVHs) were compared with plan DVHs. Using a deformed phantom, recalculated point dose variations were compared with measurements. The MVCT field of view is limited and the image space outside this field of view can be filled in with information from the planning kVCT. This merging technique was tested for a rigid phantom. Finally, the influence of the MVCT slice thickness on the dose recalculation was investigated. The dosimetric differences observed in all phantom tests were within the range of dosimetric uncertainties observed due to variations in the calibration curve. The use of MVCT images allows the assessment of daily dose distributions with an accuracy that is similar to that of the initial kVCT dose calculation.
引用
收藏
页码:4259 / 4276
页数:18
相关论文
共 15 条
  • [1] Applications of simulator computed tomography number for photon dose calculations during radiotherapy treatment planning
    Chu, JCH
    Ni, B
    Kriz, R
    Saxena, VA
    [J]. RADIOTHERAPY AND ONCOLOGY, 2000, 55 (01) : 65 - 73
  • [2] Accuracy of inhomogeneity correction in photon radiotherapy from CT scans with different settings
    Guan, HQ
    Yin, FF
    Kim, JH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (17) : N223 - N231
  • [3] Tomotherapeutic portal imaging for radiation treatment verification
    Hesse, BM
    Spies, L
    Groh, BA
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12) : 3607 - 3616
  • [4] Radiation characteristics of helical tomotherapy
    Jeraj, R
    Mackie, TR
    Balog, J
    Olivera, G
    Pearson, D
    Kapatoes, J
    Ruchala, K
    Reckwerdt, P
    [J]. MEDICAL PHYSICS, 2004, 31 (02) : 396 - 404
  • [5] On the verification of the incident energy fluence in tomotherapy IMRT
    Kapatoes, JM
    Olivera, GH
    Ruchala, KJ
    Mackie, TR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11) : 2953 - 2965
  • [6] On the accuracy and effectiveness of dose reconstruction for tomotherapy
    Kapatoes, JM
    Olivera, GH
    Balog, JP
    Keller, H
    Reckwerdt, PJ
    Mackie, TR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04) : 943 - 966
  • [7] A feasible method for clinical delivery verification and dose reconstruction in tomotherapy
    Kapatoes, JM
    Olivera, GH
    Ruchala, KJ
    Smilowitz, JB
    Reckwerdt, PJ
    Mackie, TR
    [J]. MEDICAL PHYSICS, 2001, 28 (04) : 528 - 542
  • [8] Delivery verification in sequential and helical tomotherapy
    Kapatoes, JM
    Olivera, GH
    Reckwerdt, PJ
    Fitchard, EE
    Schloesser, EA
    Mackie, TR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) : 1815 - 1841
  • [9] Tolerance levels for quality assurance of electron density values generated from CT in radiotherapy treatment planning
    Kilby, W
    Sage, J
    Rabett, V
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (09) : 1485 - 1492
  • [10] MACKIE TR, 2003, INTENSITY MODULATED