Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations

被引:74
作者
Villa, T
Migliavacca, F
Gastaldi, D
Colombo, M
Pietrabissa, R
机构
[1] Politecn Milan, Dipartimento Ingn Strutturale, I-20133 Milan, Italy
[2] Politecn Milan, Dipartimento Bioingn, Lab Biol Struct Mech, I-20133 Milan, Italy
关键词
Fuji film; finite element method; rotating platform knee prosthesis; tibial tray; ISO standard;
D O I
10.1016/S0021-9290(03)00255-0
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The evaluation of contact areas and pressures in total knee prosthesis is a key issue to prevent early failure. The first part of this study is based on the hypothesis that the patterns of contact stresses on the tibial insert of a knee prosthesis at different stages of the gait cycle could be an indicator of the wear performances of a knee prosthesis. Contact stresses were calculated for a mobile bearing knee prosthesis by means of finite element method (FEM). Contact areas and stresses were also measured through in vitro tests using Fuji Prescale film in order to support the FEM findings. The second part of this study addresses the long-term structural integrity of metal tibial components in terms of fatigue life by means of experimental tests and FEM simulations. Fatigue experimental evaluations were performed on Cr-Co alloy tibial tray, based on ISO standards. FEM models were used to calculate the stress patterns. The failure risk was estimated with a standard fatigue criterion on the basis of the results obtained from the FEM calculations. Experimental and computational results showed a positive matching. (C) 2003 Elsevier Ltd. All rights reserved.
引用
收藏
页码:45 / 53
页数:9
相关论文
共 35 条
  • [1] Fracture of the metal tibial tray after kinematic total knee replacement - A common cause of early aseptic failure
    Abernethy, PJ
    Robinson, CM
    Fowler, RM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (02): : 220 - 225
  • [2] Evaluation of a testing method for the fatigue performance of total knee tibial trays
    Ahir, SP
    Blunn, GW
    Haider, H
    Walker, PS
    [J]. JOURNAL OF BIOMECHANICS, 1999, 32 (10) : 1049 - 1057
  • [3] Fracture of the tibial tray after total knee arthroplasty
    Altintas, F
    Sener, N
    Ugutmen, E
    [J]. JOURNAL OF ARTHROPLASTY, 1999, 14 (01) : 112 - 114
  • [4] Compression and tension fatigue resistance of medical grade ultra high molecular weight polyethylene: the effect of morphology, sterilization, aging and temperature
    Baker, DA
    Hastings, RS
    Pruitt, L
    [J]. POLYMER, 2000, 41 (02) : 795 - 808
  • [5] BELL CJ, 1999, T ORTHOP RES SOC, V24, P962
  • [6] Wear in retrieved condylar knee arthroplasties - A comparison of wear in different designs of 280 retrieved condylar knee prostheses
    Blunn, GW
    Joshi, AB
    Minns, RJ
    Lidgren, L
    Lilley, P
    Ryd, L
    Engelbrecht, E
    Walker, PS
    [J]. JOURNAL OF ARTHROPLASTY, 1997, 12 (03) : 281 - 290
  • [7] RADIATION STERILIZATION OF MEDICAL DEVICES - EFFECTS OF IONIZING-RADIATION ON ULTRA-HIGH-MOLECULAR-WEIGHT POLYETHYLENE
    BUCHALLA, R
    SCHUTTLER, C
    BOGL, KW
    [J]. RADIATION PHYSICS AND CHEMISTRY, 1995, 46 (4-6): : 579 - 585
  • [8] BUECHEL FF, 1990, CLIN ORTHOP RELAT R, P162
  • [9] Cemented rotating-platform total knee replacement - A nine to twelve-year follow-up study
    Callaghan, JJ
    Squire, MW
    Goetz, DD
    Sullivan, PM
    Johnston, RC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (05) : 705 - 711
  • [10] CHEN F, 1994, CLIN ORTHOP RELAT R, P249