Novel needle cutting edge geometry for end-cut biopsy

被引:49
作者
Moore, Jason Z. [1 ]
McLaughlin, Patrick W. [2 ]
Shih, Albert J. [1 ,3 ]
机构
[1] Penn State Univ, Dept Mech & Nucl Engn, State Coll, PA 16802 USA
[2] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Dept Biomed Engn, Ann Arbor, MI 48109 USA
基金
中国国家自然科学基金; 美国国家科学基金会;
关键词
biopsy; tissue cutting; needle geometry; INSERTION;
D O I
10.1118/1.3665253
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To introduce and determine the biopsy length performance of the novel enhanced cutting edge (ECE) needle tip design, which contains high inclination angles that allow for more efficient tissue cutting. Methods: ECE and regular two-plane symmetric needle tip's biopsy performance and cutting force are compared over a series of needle insertion experiments into bovine liver under varying levels of internal needle vacuum. An earlier developed needle tip force model is also applied. From these experiments and force model, the effect of needle tip geometry and vacuum on biopsy performance and force is studied. Results: Biopsy sample length is on average 22%, 30%, and 49% longer for ECE needles compared to that of regular needles for the internal pressures of 0, -33.9, and -67.7 kPa, respectively. For ECE needles the vacuum level of -67.7 kPa produces on average biopsy lengths that are 41%, 31%, 29%, 45%, and 42% longer compared to no vacuum for two-plane needle tip bevel angles of 10 degrees, 15 degrees, 20 degrees, 25 degrees, and 30 degrees, respectively. The force results show the ECE needle can be inserted with less initial insertion force than the regular two-plane needle for needles where the needle tip is fully contacting the tissue upon insertion. Vacuum is also showed to help lower insertion forces. Conclusions: The novel ECE needle tip design outperforms the regular two-plane symmetric needle by yielding longer biopsy samples and lower insertion forces, thereby demonstrating the benefits of needle geometries that contain higher inclination angles. The use of vacuum further improves the ECE needle tip biopsy sample length and lowers insertion forces. VC 2012 American Association of Physicists in Medicine. [DOI: 10.1118/1.3665253]
引用
收藏
页码:99 / 108
页数:10
相关论文
共 22 条
[1]   Prostate biopsy outcome using 29 mm cutting length [J].
Fink, KG ;
Hutarew, G ;
Pytel, A ;
Schmeller, NT .
UROLOGIA INTERNATIONALIS, 2005, 75 (03) :209-212
[2]  
Gravlee JF, 1998, US patent, Patent No. 5733266
[3]  
Heverly M, 2005, IEEE INT CONF ROBOT, P1646
[4]  
Huber R.L., 1946, P1, Patent No. 2409979
[5]   Needle core length in sextant biopsy influences prostate cancer detection rate [J].
Iczkowski, KA ;
Casella, G ;
Seppala, RJ ;
Jones, GL ;
Mishler, BA ;
Qian, JQ ;
Bostwick, DG .
UROLOGY, 2002, 59 (05) :698-703
[6]  
Kataoka H., 2001, P MEDICAL IMAGE COMP, P966
[7]  
Kucklick T.R., 2006, MED DEVICE R D HDB
[8]   MR-guided vacuum-assisted breast biopsy: Accuracy of targeting and success in sampling in a model [J].
Lehman, CD ;
Aikawa, T .
RADIOLOGY, 2004, 232 (03) :911-914
[9]   In vivo study of forces during needle insertions [J].
Maurin, B ;
Barbe, L ;
Bayle, B ;
Zanne, P ;
Gangloff, J ;
De Mathelin, M ;
Gangi, A ;
Soler, L ;
Forgione, A .
PERSPECTIVE IN IMAGE-GUIDED SURGERY, 2004, :415-422
[10]  
MOORE J, P I MECH B IN PRESS