Thrombogenicity of deflated intraaortic balloon: Impact of heparin coating

被引:4
作者
Mueller, XM [1 ]
Tevaearai, HT
Hayoz, T
von Segesser, LK
机构
[1] CHU Vaudois, Cardiovasc Surg Clin, CH-1011 Lausanne, Switzerland
[2] CHU Vaudois, Div Vasc Med, CH-1011 Lausanne, Switzerland
关键词
intraaortic balloon pumping; thrombosis; heparin coating;
D O I
10.1046/j.1525-1594.1999.06228.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In a calf model, 3 standard and 3 heparin coated intraaortic balloons (IAB) were inserted and left deflated in the descending aorta during 20 min to simulate balloon rupture. At the end of the experiment, the IAB were examined, and 3 samples of each were collected for scanning electron microscopy (SEM) analysis. None of the heparin coated IAB showed any sign of clot or deposit whereas all 3 standard IAB exhibited clots. SEM revealed no deposit on the heparin coated samples while 8 of the 9 standard samples disclosed deposits (p < 0.001). Morphometrically, 16.1 +/- 21.4% of the surfaces of the standard samples were covered with deposits whereas all the heparin coated samples were free (p = 0.004). The presence of clots on every standard IAB supports the hypothesis that other than local factors of the insertion site can play a role in the vascular complications of IAB. The absence of clots and deposits on the heparin coated IAB suggests a promising role of such devices in circumventing the tendency of clot formation on deflated balloons while avoiding the drawback of systemic anticoagulation.
引用
收藏
页码:195 / 198
页数:4
相关论文
共 21 条
[1]  
ALPERT J, 1976, ARCH SURG-CHICAGO, V111, P1190
[2]   Preoperative intraaortic balloon pump enhances cardiac performance and improves the outcome of redo CABG [J].
Christenson, JT ;
Badel, P ;
Simonet, F ;
Schmuziger, M .
ANNALS OF THORACIC SURGERY, 1997, 64 (05) :1237-1244
[3]   PLAQUE ABRASION AND INTRAAORTIC BALLOON LEAK [J].
COX, PM ;
GORAN, SF ;
MORTON, J ;
WOLVEK, S .
CHEST, 1995, 108 (06) :1495-1498
[4]   RESULTS OF INTRAAORTIC BALLOON PUMPING AFTER CARDIAC-SURGERY - EXPERIENCE WITH THE PERCOR BALLOON CATHETER [J].
DILELLO, F ;
MULLEN, DC ;
FLEMMA, RJ ;
ANDERSON, AJ ;
KLEINMAN, LH ;
WERNER, PH .
ANNALS OF THORACIC SURGERY, 1988, 46 (04) :442-446
[5]  
GOTT VL, 1963, SCIENCE, V142, P1297, DOI 10.1126/science.142.3597.1297
[6]  
Gravlee G P, 1994, J Cardiothorac Vasc Anesth, V8, P213, DOI 10.1016/1053-0770(94)90066-3
[7]   PERCUTANEOUS INTRA-AORTIC BALLOON COUNTERPULSATION - CLINICAL EFFECTIVENESS AND HAZARDS [J].
HAUSER, AM ;
GORDON, S ;
GANGADHARAN, V ;
RAMOS, RG ;
WESTVEER, DC ;
GARG, AK ;
TIMMIS, GC .
CHEST, 1982, 82 (04) :422-425
[8]   COMPLICATIONS OF THE INTRA-AORTIC BALLOON COUNTERPULSATION DEVICE - CLINICAL AND MORPHOLOGIC OBSERVATIONS IN 45 NECROPSY PATIENTS [J].
ISNER, JM ;
COHEN, SR ;
VIRMANI, R ;
LAWRINSON, W ;
ROBERTS, WC .
AMERICAN JOURNAL OF CARDIOLOGY, 1980, 45 (02) :260-268
[9]  
Kantrowitz A, 1968, JAMA-J AM MED ASSOC, V203, P135
[10]  
KVILEKVAL KHV, 1991, ARCH SURG-CHICAGO, V126, P621