Coincidence summing effects in gamma-ray spectrometry using a Marinelli beaker

被引:10
作者
Lee, Mosung [2 ]
Park, Tae Soon [1 ]
Woo, Jong-Kwan [3 ]
机构
[1] Korea Res Inst Stand & Sci, Taejon 305340, South Korea
[2] Cheong Ju Univ, Dept Nano Sci, CheongJu 360764, South Korea
[3] Univ Seoul, Dept Phys, Seoul 130743, South Korea
关键词
coincidence summing; peak-to-total ratio; (133)Ba; (152)Eu; Marinelli beaker;
D O I
10.1016/j.apradiso.2008.02.017
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The coincidence summing effects on a HPGe spectrometer have been studied by using Marinelli beakers containing (133)Ba and (152)Eu solutions. The coincidence summing effects were calculated from the ratio of full-energy peak efficiency and total efficiency. The peak efficiency values for (133)Ba and (152)Eu corrected by using the peak-to-total ratio were found to be in good agreement (within 3%) as compared with the efficiency values from Marinelli beakers containing single gamma-ray emitters. (c) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:799 / 803
页数:5
相关论文
共 6 条
[1]  
*AM NAT STAND I, 1999, ANSIN4214
[2]  
Be M.-M., 2004, Monographie BIPM-5
[3]   The influence of peak area determination bias in quantitative gamma-ray spectrometry in the presence of true coincidence summing [J].
Blaauw, M ;
Gelsema, SJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3) :417-422
[4]  
Debertin K., 1988, Gamma and X-Ray Spectrometry with Semiconductor Detectors, P258
[5]  
GILMORE G, 1995, PRACTICAL GAMMA RAY, P148
[6]   MEASUREMENT OF RADIOACTIVE SAMPLES IN MARINELLI BEAKERS BY GAMMA-RAY SPECTROMETRY [J].
PARK, TS ;
JEON, WJ .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1995, 193 (01) :133-144