Evaluation of Rebound Tonometer for Measuring Intraocular Pressure at Deviated Angle and Position

被引:9
作者
Takenaka, Joji [1 ]
Mochizuki, Hideki
Kunihara, Eriko
Tanaka, Junko [2 ]
Kiuchi, Yoshiaki
机构
[1] Hiroshima Univ, Dept Ophthalmol & Visual Sci, Grad Sch Biomed Sci, Minami Ku, Hiroshima 7348551, Japan
[2] Hiroshima Univ, Dept Epidemiol Infect Dis Control & Prevent, Grad Sch Biomed Sci, Hiroshima 7348551, Japan
关键词
Intraocular pressure; Perkins applanation tonometer; Rabbit; Rebound tonometer; Tonopen XL tonometer; Transducer; GLAUCOMA; CHILDREN; RABBITS;
D O I
10.3109/02713683.2010.534574
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate the influence of positional deviations on the rebound tonometer measurement. Materials and Methods: Ten white rabbits were used. A bottle of balanced salt solution was connected to the anterior chamber, and the IOP was altered by changing the height of the bottle. A pressure transducer was also connected to the anterior chamber. The IOP was measured by the rebound tonometer (RBT), the Tonopen XL, and the Perkins handheld applanation tonometer at optimal conditions and compared each tonometer. Then, the effect of placing the RBT probe at different distances, different lateral positions, and different angles to the cornea on the IOP was determined. Results: The RBT had the lowest variance in all three tonometers. The IOP measured with the RBT was not significantly altered when the distance from the cornea was 4 to 8 mm, or when the probe was displaced 2 mm to the central cornea, or when the probe was placed at an angle of 20 degrees from the central axis. Conclusions: The low variance of the RBT measurements and the non-significant changes in the IOP when the probe of the RBT was not on the central axis indicate that the RBT is useful for animal models of glaucoma. Further studies examining its use (repeatability and reliability) in human infants whose eyes and head are usually moving are required.
引用
收藏
页码:422 / 428
页数:7
相关论文
共 16 条
[1]  
Abrams LS, 1996, INVEST OPHTH VIS SCI, V37, P940
[2]  
AUSINSCH B, 1977, ANN OPHTHALMOL, V9, P1391
[3]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[4]  
Chiara G F, 1989, J Am Optom Assoc, V60, P105
[5]   The influence of cornea properties on rebound tonometry [J].
Chui, Wan-sang ;
Lam, Andrew ;
Chen, Davie ;
Chiu, Roger .
OPHTHALMOLOGY, 2008, 115 (01) :80-84
[6]  
HETHERIN.J, 1968, INVEST OPHTH VISUAL, V7, P134
[7]   Clinical performance of the Reichert AT550:: a new non-contact tonometer [J].
Jorge, J ;
Díaz-Rey, JA ;
González-Méijome, JM ;
Almeida, JB ;
Parafita, MA .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2002, 22 (06) :560-564
[8]   A new electromechanical method for measuring intraocular pressure [J].
Kontiola, A .
DOCUMENTA OPHTHALMOLOGICA, 1997, 93 (03) :265-276
[9]   The induction/impact tonometer: A new instrument to measure intraocular pressure in the rat [J].
Kontiola, AI ;
Goldblum, D ;
Mittag, T ;
Danias, J .
EXPERIMENTAL EYE RESEARCH, 2001, 73 (06) :781-785
[10]   Comparison of intraocular tonometry using three different non-invasive tonometers in children [J].
Lasseck, Julia ;
Jehle, Thomas ;
Feltgen, Nicolas ;
Lagreze, Wolf Alexander .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (10) :1463-1466