Mid-term outcomes of tantalum cup- a single centre study

被引:4
作者
Bawale, Rajesh [1 ]
Choudhry, Baseem [1 ]
Samsani, Srinivasa [1 ]
机构
[1] Medway Maritime Hosp, Dept Trauma & Orthopaed, Windmill Rd, Gillingham ME7 5NY, England
关键词
Hip joint; Revision arthroplasty; Reconstruction; Complications; TOTAL HIP-ARTHROPLASTY; POROUS METAL AUGMENTS; REVISION HIP; ACETABULAR REVISION; BONE INGROWTH; FOLLOW-UP; 10-YEAR; PERFORMANCE;
D O I
10.1186/s42836-021-00099-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction The cementless acetabular implants are commonly used in primary and revision hip arthroplasty. Reconstruction of acetabulum in case of bone defects can be challenging. The aims of this single center study are to review the mid-term outcomes of porous tantalum cups (TM) and evaluate complications. Methods The midterm outcome of a trabecular metal tantalum modular uncemented cup was evaluated in 59 hips in 58 patients. In our group, we had 23 males and 35 females. The mean age was 70.11 years (range, 30 to 87 years). Four patients were lost to follow-up and 13 died during the period without having further surgeries attributed to the hip arthroplasty. The remaining 41 patients (42 revision hip arthroplasties) had complete data available. Results The mean follow-up was 87 months, ranging from 24 to 144 months. Standard pelvic anteroposterior (AP) radiographs were used to assess and preoperatively classify acetabular defects as per Paprosky classification. The serial radiographs showed excellent stability, bone opposition and graft incorporation. Four patients had further surgeries. Two of these were due to infection (one superficial and one deep infection). One of the patients had washout and then removal of metal work, the other patient only had a washout and symptoms settled. One patient had vascular compromise and went for surgery to stem the bleeding. One patient had re-revision due to stem loosening and hence required surgery but the revision cup remained stable. We noted a 96% survival at an average of 7.2 years follow-up. Conclusion The mid-term results with the trabecular metal cementless cup appeared to be promising in both primary and revision hip arthroplasty, even in the presence of considerable bone loss which requires bone grafting and augments.
引用
收藏
页数:7
相关论文
共 29 条
  • [1] Combined trabecular metal acetabular shell and augment for acetabular revision with substantial bone loss
    Abolghasemian, M.
    Tangsataporn, S.
    Sternheim, A.
    Backstein, D.
    Safir, O.
    Gross, A. E.
    [J]. BONE & JOINT JOURNAL, 2013, 95B (02) : 166 - 172
  • [2] [Anonymous], 2020, The National Joint Registry 17th Annual Report 2020
  • [3] Porous tantalum structures for bone implants: Fabrication, mechanical and in vitro biological properties
    Balla, Vamsi Krishna
    Bodhak, Subhadip
    Bose, Susmita
    Bandyopadhyay, Amit
    [J]. ACTA BIOMATERIALIA, 2010, 6 (08) : 3349 - 3359
  • [4] Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial
    Bobyn, JD
    Stackpool, GJ
    Hacking, SA
    Tanzer, M
    Krygier, JJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (05): : 907 - 914
  • [5] Trabecular Metal Used for Major Bone Loss in Acetabular Hip Revision
    Davies, Jonah Hebert
    Laflamme, G. Yves
    Delisle, Josee
    Fernandes, Julio
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (08) : 1245 - 1250
  • [6] Middle term results of tantalum acetabular cups in total hip arthroplasty following pelvic irradiation
    De Paolis, Massimiliano
    Zucchini, Riccardo
    Romagnoli, Carlo
    Romantini, Matteo
    Mariotti, Federica
    Donati, Davide Maria
    [J]. ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2019, 53 (03) : 165 - 169
  • [7] Midterm Outcomes of Revision Total Hip Arthroplasty With the Use of a Multihole Highly-Porous Titanium Shell
    Delanois, Ronald E.
    Gwam, Chukwuweike U.
    Mohamed, Nequesha
    Khlopas, Anton
    Chughtai, Morad
    Malkani, Arthur L.
    Mont, Michael A.
    [J]. JOURNAL OF ARTHROPLASTY, 2017, 32 (09) : 2806 - 2809
  • [8] 20 years of porous tantalum in primary and revision hip arthroplasty-time for a critical appraisal
    Hailer, Nils
    [J]. ACTA ORTHOPAEDICA, 2018, 89 (03) : 254 - 255
  • [9] Bone Ingrowth in Well-Fixed Retrieved Porous Tantalum Implants
    Hanzlik, Josa A.
    Day, Judd S.
    [J]. JOURNAL OF ARTHROPLASTY, 2013, 28 (06) : 922 - 927
  • [10] Material-specific thrombin generation following contact between metal surfaces and whole blood
    Hong, J
    Azens, A
    Ekdahl, KN
    Granqvist, CG
    Nilsson, B
    [J]. BIOMATERIALS, 2005, 26 (12) : 1397 - 1403