Ultrasonic anemometer for the measurement of respiratory flow in the forced oscillation technique

被引:0
|
作者
Araujo, Guilherme A. L. [1 ]
Junior, Reginardo T. L. [2 ]
Freire, Raimundo C. S. [2 ]
Silva, Ivan S. S. [1 ]
da Silva, Jose F. [3 ]
Catunda, Yuri S. C. [4 ]
Carvalho, Elyson A. N. [2 ]
机构
[1] UFPA CT DEEC, Rua Augusto Correa 01, BR-66075900 Belem, Para, Brazil
[2] UFCG DEE LIMC, BR-58109970 Campina Grande, Paraiba, Brazil
[3] UFPB CT DTM, BR-58051970 Joao Pessoa, Paraiba, Brazil
[4] UFMA CT DEE, BR-65080040 Sao Luis, Brazil
来源
MEMEA 2007: SECOND - IEEE INTERNATIONAL WORKSHOP ON MEDICAL MEASUREMENT AND APPLICATIONS | 2007年
关键词
respiratory flow; ultrasonic anemometer; forced oscillations technique;
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper it is proposed a method to the measurement of respiratory flow in the forced oscillations technique (FOT) using an ultrasonic anemometer. It is used the transit time method and the phase difference technique with the measurements of the phase angle between the signal at the transmission and at the receiver to detect changes in the transit time due to flow. The ultrasonic anemometer implementation is based on a microcontroller. The experimental results of the tests with the comparison among this flowmeter and a classical pneumotachometer used in the FOT for flow measurement will are presented
引用
收藏
页码:3 / +
页数:2
相关论文
共 5 条
  • [1] Measurement range expansion of continuous wave ultrasonic anemometer
    Han, Dongwoo
    Park, Sekwang
    MEASUREMENT, 2011, 44 (10) : 1909 - 1914
  • [2] Measurement errors and correction of the UAT-2 ultrasonic anemometer
    CHENG Xue Ling
    PENG Zhen
    HU Fei
    ZENG Qing Cun
    LUO Wei Dong
    ZHAO Yi Jun
    HONG Zhong Xiang
    Science China(Technological Sciences), 2015, (04) : 677 - 686
  • [3] Measurement errors and correction of the UAT-2 ultrasonic anemometer
    Cheng XueLing
    Peng Zhen
    Hu Fei
    Zeng QingCun
    Luo WeiDong
    Zhao YiJun
    Hong ZhongXiang
    SCIENCE CHINA-TECHNOLOGICAL SCIENCES, 2015, 58 (04) : 677 - 686
  • [4] Measurement errors and correction of the UAT-2 ultrasonic anemometer
    XueLing Cheng
    Zhen Peng
    Fei Hu
    QingCun Zeng
    WeiDong Luo
    YiJun Zhao
    ZhongXiang Hong
    Science China Technological Sciences, 2015, 58 : 677 - 686
  • [5] APPLICATION OF ULTRASONIC ANEMOMETER ARRAY TO FIELD MEASUREMENTS OF THE DOWNWASH FLOW OF AN AGRICULTURAL UNMANNED HELICOPTER
    Tang, Q.
    Zhang, R.
    Ding, C.
    Chen, L.
    Xu, M.
    Xu, G.
    Yi, T.
    Wen, Y.
    Li, L.
    Deng, W.
    TRANSACTIONS OF THE ASABE, 2019, 62 (05) : 1219 - 1230