Magnetic microbead detection using the planar Hall effect

被引:66
作者
Ejsing, L
Hansen, MF
Menon, AK
Ferreira, HA
Graham, DL
Freitas, PP
机构
[1] Tech Univ Denmark, Dept Micro & Nanotechnol, DK-2800 Lyngby, Denmark
[2] INESC, MN, P-1000029 Lisbon, Portugal
关键词
Hall effect (planar); micromer-M; magnetic sensor; biosensor; permalloy (Ni80Fe20); MnIr (Mn76Ir24); detection of microspheres;
D O I
10.1016/j.jmmm.2005.02.071
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Magnetic sensors based on the planar Hall effect of exchanged-biased permalloy have been fabricated and characterized. It is demonstrated that the sensors are feasible for detecting just a few commercial 2.0 mu m magnetic beads commonly used for bioseparation (Micromer-M, Micromod, Germany) and that the sensor sense current is sufficient to generate a signal from the beads. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:677 / 684
页数:8
相关论文
共 16 条
  • [1] A biosensor based on magnetoresistance technology
    Baselt, DR
    Lee, GU
    Natesan, M
    Metzger, SW
    Sheehan, PE
    Colton, RJ
    [J]. BIOSENSORS & BIOELECTRONICS, 1998, 13 (7-8) : 731 - 739
  • [2] Detection of a single magnetic microbead using a miniaturized silicon Hall sensor
    Besse, PA
    Boero, G
    Demierre, M
    Pott, V
    Popovic, R
    [J]. APPLIED PHYSICS LETTERS, 2002, 80 (22) : 4199 - 4201
  • [3] The BARC biosensor applied to the detection of biological warfare agents
    Edelstein, RL
    Tamanaha, CR
    Sheehan, PE
    Miller, MM
    Baselt, DR
    Whitman, LJ
    Colton, RJ
    [J]. BIOSENSORS & BIOELECTRONICS, 2000, 14 (10-11) : 805 - 813
  • [4] Planar Hall effect sensor for magnetic micro- and nanobead detection
    Ejsing, L
    Hansen, MF
    Menon, AK
    Ferreira, HA
    Graham, DL
    Freitas, PP
    [J]. APPLIED PHYSICS LETTERS, 2004, 84 (23) : 4729 - 4731
  • [5] EJSING L, 2003, EUROSENSORS 2003, P1095
  • [6] Biodetection using magnetically labeled biomolecules and arrays of spin valve sensors (invited)
    Ferreira, HA
    Graham, DL
    Freitas, PP
    Cabral, JMS
    [J]. JOURNAL OF APPLIED PHYSICS, 2003, 93 (10) : 7281 - 7286
  • [7] FREITAS PP, 2004, MAGNETORESISTIVE BIO
  • [8] Ion beam deposition of Mn-Ir spin valves
    Gehanno, V
    Freitas, PP
    Veloso, A
    Ferreira, J
    Almeida, B
    Sousa, JB
    Kling, A
    Soares, JC
    da Silva, MF
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 1999, 35 (05) : 4361 - 4367
  • [9] High sensitivity detection of molecular recognition using magnetically labelled biomolecules and magnetoresistive sensors
    Graham, DL
    Ferreira, HA
    Freitas, PP
    Cabral, JMS
    [J]. BIOSENSORS & BIOELECTRONICS, 2003, 18 (04) : 483 - 488
  • [10] Single magnetic microsphere placement and detection on-chip using current line designs with integrated spin valve sensors: Biotechnological applications
    Graham, DL
    Ferreira, H
    Bernardo, J
    Freitas, PP
    Cabral, JMS
    [J]. JOURNAL OF APPLIED PHYSICS, 2002, 91 (10) : 7786 - 7788