Single-Breath Diffusing Capacity for Carbon Monoxide Instrument Accuracy Across 3 Health Systems

被引:7
作者
Hegewald, Matthew J. [1 ,2 ]
Markewitz, Boaz A. [2 ]
Wilson, Emily L. [1 ]
Gallo, Heather M. [1 ]
Jensen, Robert L. [1 ,2 ]
机构
[1] Intermt Med Ctr, Murray, UT 84157 USA
[2] Univ Utah, Salt Lake City, UT USA
关键词
accuracy; quality control; simulator; single-breath diffusing capacity for carbon monoxide; PULMONARY-FUNCTION; VARIABILITY;
D O I
10.4187/respcare.03512
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Measuring diffusing capacity of the lung for carbon monoxide (D-LCO) is complex and associated with wide intra- and inter-laboratory variability. Increased D-LCO variability may have important clinical consequences. The objective of the study was to assess instrument performance across hospital pulmonary function testing laboratories using a D-LCO simulator that produces precise and repeatable D-LCO values. METHODS: D-LCO instruments were tested with CO gas concentrations representing medium and high range D-LCO values. The absolute difference between observed and target D-LCO value was used to determine measurement accuracy; accuracy was defined as an average deviation from the target value of <2.0 mL/min/mm Hg. Accuracy of inspired volume measurement and gas sensors were also determined. RESULTS: Twenty-three instruments were tested across 3 healthcare systems. The mean absolute deviation from the target value was 1.80 mL/min/mm Hg (range 0.24-4.23) with 10 of 23 instruments (43%) being inaccurate. High volume laboratories performed better than low volume laboratories, although the difference was not significant. There was no significant difference among the instruments by manufacturers. Inspired volume was not accurate in 48% of devices; mean absolute deviation from target value was 3.7%. Instrument gas analyzers performed adequately in all instruments. CONCLUSIONS: D-LCO instrument accuracy was unacceptable in 43% of devices. Instrument inaccuracy can be primarily attributed to errors in inspired volume measurement and not gas analyzer performance. D-LCO instrument performance may be improved by regular testing with a simulator. Caution should be used when comparing D-LCO results reported from different laboratories.
引用
收藏
页码:430 / 436
页数:7
相关论文
共 20 条
[1]   Amiodarone pulmonary toxicity [J].
Camus, P ;
Martin, WJ ;
Rosenow, EC .
CLINICS IN CHEST MEDICINE, 2004, 25 (01) :65-+
[2]   Single-breath carbon monoxide diffusing capacity [J].
Crapo, RO ;
Jensen, RL ;
Wanger, JS .
CLINICS IN CHEST MEDICINE, 2001, 22 (04) :637-+
[3]  
Crapo Robert O, 2003, Respir Care, V48, P764
[4]   A prospective study of pulmonary function in patients treated with paclitaxel and carboplatin [J].
Dimopoulou, I ;
Galani, H ;
Dafni, U ;
Samakovli, A ;
Roussos, C ;
Dimopoulos, MA .
CANCER, 2002, 94 (02) :452-458
[5]   American College of Chest Physicians and Society of Thoracic Surgeons Consensus Statement for Evaluation and Management for High-Risk Patients With Stage I Non-small Cell Lung Cancer [J].
Donington, Jessica ;
Ferguson, Mark ;
Mazzone, Peter ;
Handy, John, Jr. ;
Schuchert, Matthew ;
Fernando, Hiran ;
Loo, Billy, Jr. ;
Lanuti, Michael ;
de Hoyos, Alberto ;
Detterbeck, Frank ;
Pennathur, Arjun ;
Howington, John ;
Landreneau, Rodney ;
Silvestri, Gerard .
CHEST, 2012, 142 (06) :1620-1635
[6]   Idiopathic pulmonary fibrosis - Prognostic value of changes in physiology and six-minute-walk test [J].
Flaherty, Kevin R. ;
Andrei, Adin-Cristian ;
Murray, Susan ;
Fraley, Chris ;
Colby, Thomas V. ;
Travis, William D. ;
Lama, Vibha ;
Kazerooni, Ella A. ;
Gross, Barry H. ;
Toews, Galen B. ;
Martinez, Fernando J. .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2006, 174 (07) :803-809
[7]   Practical guidelines for clinicians who treat patients with amiodarone [J].
Goldschlager, N ;
Epstein, AE ;
Naccarelli, G ;
Olshansky, B ;
Singh, B .
ARCHIVES OF INTERNAL MEDICINE, 2000, 160 (12) :1741-1748
[8]  
Hegewald MJ, 2011, AM J RESP CRIT CARE, V183
[9]   Quality control of DL,CO instruments in global clinical trials [J].
Jensen, R. ;
Leyk, M. ;
Crapo, R. ;
Muchmore, D. ;
Berclaz, P. Y. .
EUROPEAN RESPIRATORY JOURNAL, 2009, 33 (04) :828-834
[10]   Sources of long-term variability in measurements of lung function - Implications for interpretation and clinical trial design [J].
Jensen, Robert L. ;
Teeter, John G. ;
England, Richard D. ;
Howell, Heather M. ;
White, Heather J. ;
Pickering, Eve H. ;
Crapo, Robert O. .
CHEST, 2007, 132 (02) :396-402