Improved treatment planning for COMS eye plaques

被引:61
作者
Astrahan, MA [1 ]
机构
[1] Univ So Calif, Norris Canc Ctr, Dept Radiat Oncol, Los Angeles, CA 90033 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2005年 / 61卷 / 04期
关键词
eye plaque; seed carrier; silastic; dosimetry; I-125; Pd-103;
D O I
10.1016/j.ijrobp.2004.09.062
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A recent reanalysis of the Collaborative Ocular Melanoma Study (COMS) medium tumor trial concluded that incorporating factors to account for anisotropy, line source approximation, the gold plaque, and attenuation in the Silastic seed carrier into the dose calculations resulted in a significant and consistent reduction of calculated doses to structures of interest within the eye. The authors concluded that future eye plaque dosimetry should be "performed using the most up-to-date parameters available." The reason these factors are important is attributable to the low energy I-125 radiation (approximately 28 keV) that is primarily absorbed by the photoelectric process. Photoelectric absorption is quite dependent on the atomic composition of the absorbing material. Being 40% silicon by weight, the effective atomic number of Silastic is significantly greater than that of water. Although the AAPM TG43 brachytherapy formalism inherently addresses the issues of source anisotropy and geometry, its parameter that accounts for scatter and attenuation, the radial dose function g(r), assumes that the source is immersed in infinite homogeneous water. In this work, factors are proposed for I-125 that correct for attenuation in the Silastic carrier and scatter deficits resulting from the gold plaque and nearby air. The implications of using Pd-103 seeds in COMS plaques are also discussed. Methods and Materials: An existing TG43-based ophthalmic plaque planning system was modified to incorporate additional scatter and attenuation correction factors that better account for the path length of primary radiation in the Silastic seed carrier and the distance between the dose calculation point and the eye-air interface. Results: Compared with homogeneous water, the dose-modifying effects of the Silastic and gold are greatest near the plaque surface and immediately adjacent to the plaque, while being least near the center of the eye. The calculated dose distribution surrounding a single 1251 seed centered in a COMS 20 mm plaque was found to be consistent with previously published examples that used thermoluminescent dosimetry measurements and Monte Carlo methods. For fully loaded 12 and 20 mm plaques, calculated dose to critical ocular structures ranged from 16%-50% less than would have been reported using the standard COMS dose calculation protocol. Conclusions: Treatment planning for COMS eye plaques that accurately accounts for the presence of the gold, Silastic and extraocular air is both possible and practical. (c) 2005 Elsevier Inc.
引用
收藏
页码:1227 / 1242
页数:16
相关论文
共 29 条
  • [1] [Anonymous], 2003, PHYS RAD THERAPY
  • [2] Conformal episcleral plaque therapy
    Astrahan, MA
    Luxton, G
    Pu, Q
    Petrovich, Z
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (02): : 505 - 519
  • [3] OPTIMIZATION OF I-125 OPHTHALMIC PLAQUE BRACHYTHERAPY
    ASTRAHAN, MA
    LUXTON, G
    JOZSEF, G
    LIGGETT, PE
    PETROVICH, Z
    [J]. MEDICAL PHYSICS, 1990, 17 (06) : 1053 - 1057
  • [4] A patch source model for treatment planning of ruthenium ophthalmic applicators
    Astrahan, MA
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1219 - 1228
  • [5] AN INTERACTIVE TREATMENT PLANNING SYSTEM FOR OPHTHALMIC PLAQUE RADIOTHERAPY
    ASTRAHAN, MA
    LUXTON, G
    JOZSEF, G
    KAMPP, TD
    LIGGETT, PE
    SAPOZINK, MD
    PETROVICH, Z
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (03): : 679 - 687
  • [6] DOSIMETRY FOR I-125 SEED (MODEL-6711) IN EYE PLAQUES
    CHIUTSAO, ST
    ANDERSON, LL
    OBRIEN, K
    STABILE, L
    LIU, JC
    [J]. MEDICAL PHYSICS, 1993, 20 (02) : 383 - 389
  • [7] Collaborative Ocular Melanoma Study Group, 1995, COMS MAN PROC
  • [8] EFFECTS OF GOLD AND SILVER BACKINGS ON THE DOSE-RATE AROUND AN I-125 SEED
    CYGLER, J
    SZANTO, J
    SOUBRA, M
    ROGERS, DWO
    [J]. MEDICAL PHYSICS, 1990, 17 (02) : 172 - 178
  • [10] Diener-West M, 2001, ARCH OPHTHALMOL-CHIC, V119, P969