210Pb activity by detection of bremsstrahlung in 210Bi β-decay

被引:5
|
作者
Nachab, A. [1 ]
Hubert, Ph [2 ]
机构
[1] Univ Strasbourg, CNRS, IN2P3, IPHC,UMR 7178, F-67037 Strasbourg 2, France
[2] Univ Bordeaux 1, CNRS, IN2P3, Ctr Etud Nucl Bordeaux Gradignan,UMR 5797, F-33175 Gradignan, France
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS | 2012年 / 274卷
关键词
gamma-Spectrometry; Lead shield; Pb-210; Bi-210; Bremsstrahlung;
D O I
10.1016/j.nimb.2011.11.020
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A method for the determination of the Pb-210 activity by HPGe spectrometry is proposed, consisting of measuring the bremsstrahlung generated in a sample by Bi-210 beta-emission (1.16 MeV). The detection efficiency is evaluated by GEANT3 simulation and is compared with experiment. A detection limit of 600 mBq was obtained. This technique has been used to select low background detector materials and lead shielding. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:188 / 190
页数:3
相关论文
共 28 条
  • [1] Contribution of 210Pb bremsstrahlung to the background of lead shielded gamma spectrometers
    Mrda, D.
    Bikit, I.
    Veskovic, M.
    Forkapic, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 572 (02): : 739 - 744
  • [2] Isolation of lead from water samples and determination of 210Pb
    Grahek, Z
    Macefat, MR
    Lulic, S
    ANALYTICA CHIMICA ACTA, 2006, 560 (1-2) : 84 - 93
  • [3] Calibration and measurement of 210Pb using two independent techniques
    Villa-Alfageme, Marie
    Hurtado, S.
    Manjon, G.
    Garcia-Tenorio, R.
    RADIATION MEASUREMENTS, 2007, 42 (09) : 1552 - 1560
  • [4] A comparative study of alternative methods for 210Pb determination in environmental samples
    Cuesta, E.
    Barba-Lobo, A.
    Lozano, R. L.
    San Miguel, E. G.
    Mosqueda, F.
    Bolivar, J. P.
    RADIATION PHYSICS AND CHEMISTRY, 2022, 191
  • [5] 210Pb determination by gamma spectrometry in voluminal samples (cylindrical geometry)
    San Miguel, EG
    Pérez-Moreno, JP
    Bolívar, JP
    García-Tenorio, R
    Martin, JE
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 493 (1-2): : 111 - 120
  • [6] A decade of 7Be and 210Pb activity in surface aerosols measured over the Western Iberian Peninsula
    Carvalho, Ana Cristina
    Reis, Mario
    Silva, Lidia
    Madruga, Maria Jose
    ATMOSPHERIC ENVIRONMENT, 2013, 67 : 193 - 202
  • [7] Recent Bremsstrahlung-based assays of 210Pb in lead and comments on current availability of low-background lead in North America
    Keillor, Martin E.
    Aalseth, C. E.
    Arnquist, I. J.
    Eggemeyer, T. A.
    Fuller, E. S.
    Glasgow, B. D.
    Hoppe, E. W.
    Morley, S. M.
    Myers, A. W.
    Orrell, J. L.
    Overman, C. T.
    Seifert, A.
    Shaff, S. M.
    Thommasson, K. S.
    APPLIED RADIATION AND ISOTOPES, 2017, 126 : 185 - 187
  • [8] Low-level gamma-ray spectrometry for the determination of 210Pb
    Nikola Marković
    Per Roos
    Sven P. Nielsen
    Journal of Radioanalytical and Nuclear Chemistry, 2017, 311 : 1473 - 1478
  • [9] Low-level gamma-ray spectrometry for the determination of 210Pb
    Markovic, Nikola
    Roos, Per
    Nielsen, Sven P.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 311 (02) : 1473 - 1478
  • [10] 210Pb determination in lead shields for low-level γ-spectrometry applying two independent radiometric techniques
    Hurtado, S
    García-Tenorio, R
    García-León, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 497 (2-3): : 381 - 388