Determination of Entrance Surface dose for Three Common Planar Diagnostic X-Ray Examinations

被引:1
|
作者
Jeyasugiththan, Jeyasingam [1 ]
Ehalagasthanna, G. P. N. T. [1 ]
Hishaam, Urshella [1 ]
Satharasinghe, D. M. [1 ]
机构
[1] Univ Colombo, Fac Sci, Dept Nucl Sci, Colombo 03, Sri Lanka
关键词
RADIOGRAPHY; QUALITY;
D O I
10.1093/rpd/ncac177
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Patient dose during diagnostic radiography procedures primarily depends on the entrance surface dose (ESD) and the exposure area. The purpose of this study was to determine ESD for five common diagnostic X-ray examinations using dose area product and X-ray tube output. The ESDs were estimated in a sample of 340 patients using normalised X-ray output from Philips-PrimaryDiagnost digital X-ray machine equipped with digital flat panel scintillator detector. The resultant mean ESDs for chest postero-anterior/lateral, lumbar spine antero-posterior (AP)/lateral and abdomen AP were 0.49, 2.00, 5.74, 12.79 and 4.62 mGy, respectively. The mean and third quartile ESD values of this study were significantly higher than the values found in the literature. This can be attributed to the lower tube potential (kV(p)) values and the higher tube current-time products (mAs). The methodology used here is a better option for low-resources countries for routine dose monitoring.
引用
收藏
页码:1361 / 1367
页数:7
相关论文
共 50 条
  • [21] MANAGEMENT OF PATIENT DOSES FROM DIGITAL X-RAY CHEST SCREENING EXAMINATIONS
    Vodovatov, A. V.
    Drozdov, A. A.
    Telnova, A. U.
    Bernhardsson, C.
    RADIATION PROTECTION DOSIMETRY, 2016, 169 (1-4) : 232 - 239
  • [22] Assessment of radiation risk to pediatric patients undergoing conventional X-ray examinations
    Nahangi, H.
    Chaparian, A.
    RADIOPROTECTION, 2015, 50 (01) : 19 - 25
  • [23] Features of Images Obtained by Examinations of Large Animals on X-Ray Tomosynthesis Units
    Miroshnichenko, Sergii
    Nevgasymyi, Andrii
    Miroshnichenko, Oleksandra
    Khobta, Yurii
    Miroshnichenko, Natalia
    2019 IEEE 39TH INTERNATIONAL CONFERENCE ON ELECTRONICS AND NANOTECHNOLOGY (ELNANO), 2019, : 536 - 541
  • [24] Quantifying the effect of anode surface roughness on diagnostic x-ray spectra using Monte Carlo simulation
    Mehranian, A.
    Ay, M. R.
    Alam, N. Riyahi
    Zaidi, H.
    MEDICAL PHYSICS, 2010, 37 (02) : 742 - 752
  • [25] Fast method for the estimation of the absorbed dose in X-ray microtomography
    De Muynck, Amelie
    Bonte, Stijn
    Dhaene, Jelle
    Dierick, Manuel
    Bacher, Klaus
    Van Hoorebeke, Luc
    Boone, Matthieu N.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2019, 452 : 40 - 47
  • [26] DOSE EVALUATION FOR DIGITAL X-RAY IMAGING OF PREMATURE NEONATES
    Minkels, T. J. M.
    Jeukens, C. R. L. P. N.
    Andriessen, P.
    van der Linden, A. N.
    Dam, A. J.
    van Straaten, H. L. M.
    Cottaar, E. J. E.
    van Pul, C.
    RADIATION PROTECTION DOSIMETRY, 2017, 177 (04) : 440 - 449
  • [27] Radiography students' learning of plain X-ray examinations in clinical practice: An ethnographic study
    Holmstrom, Anneli
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2025, 56 (03)
  • [28] ESTABLISHMENT OF RADIATION DOSES FOR PEDIATRIC X-RAY EXAMINATIONS IN A LARGE PEDIATRIC HOSPITAL IN TURKEY
    Olgar, T.
    Sahmaran, T.
    RADIATION PROTECTION DOSIMETRY, 2017, 176 (03) : 302 - 308
  • [29] Diagnostic Imaging of Spinal Deformities Reducing Patients Radiation Dose With a New Slot-Scanning X-ray Imager
    Deschenes, Sylvain
    Charron, Guy
    Beaudoin, Gilles
    Labelle, Hubert
    Dubois, Josee
    Miron, Marie-Claude
    Parent, Stefan
    SPINE, 2010, 35 (09) : 989 - 994
  • [30] A sensitometric comparison of four dental X-ray films and their diagnostic accuracy
    Svenson, B
    Welander, U
    Shi, XQ
    Stamatakis, H
    Tronje, G
    DENTOMAXILLOFACIAL RADIOLOGY, 1997, 26 (04) : 230 - 235