Accelerated Photo-Induced Hydrophilicity Promotes Osseointegration: An Animal Study

被引:44
作者
Jimbo, Ryo [2 ]
Ono, Daisuke [1 ]
Hirakawa, Yuko [1 ]
Odatsu, Tetsurou [1 ]
Tanaka, Tsunehiro [1 ]
Sawase, Takashi [1 ]
机构
[1] Nagasaki Univ, Grad Sch Biomed Sci, Div Appl Prosthodont, Nagasaki 8528588, Japan
[2] Nagasaki Univ, Grad Sch Biomed Sci, Div Regenerat Oral Surg, Nagasaki 8528588, Japan
基金
日本学术振兴会;
关键词
anatase TiO2; enhanced bone apposition; fluoride; osseointegration; photo-induced hydrophilicity; ORAL IMPLANT SURFACES; DENTAL IMPLANTS; PLASMA FIBRONECTIN; TITANIUM IMPLANTS; BONE APPOSITION; IN-VIVO; RESPONSES; BEHAVIOR;
D O I
10.1111/j.1708-8208.2009.00179.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: In the previous in vitro study, fluoride-modified, anodized porous titanium was proven to have enhanced its photo-induced hydrophilicity, which induced the hyperactivation of initial cell response. Purpose: The purpose of the present study was to investigate in vivo bone apposition during the early stages of osseointegration in rabbit tibiae. Materials and Methods: Anodized porous titanium implants (TiU, TiUnite (R), Nobel Biocare AB, Goteborg, Sweden) were modified with 0.175 wt% ammonium hydrogen fluoride solution (NH4F-HF2). Twenty-four hours prior to the experiments, the surface-modified implants were ultraviolet-irradiated (modTiU). Blinded and unpackaged TiU implants were used as controls. Thereafter, the implants were placed in the rabbit tibial metaphyses and histomorphometrically analyzed at 2 and 6 weeks after insertion. Results: ModTiU demonstrated a significantly greater degree of bone-to-metal contact than TiU after 2 and 6 weeks of healing. Conclusion: The results proved that the enhanced photo-induced hydrophilicity of the NH4F-HF2-modified anodized implants promoted bone apposition during the early stages of osseointegration.
引用
收藏
页码:79 / 85
页数:7
相关论文
共 30 条
[11]  
Dean J W 3rd, 1995, Int J Oral Maxillofac Implants, V10, P721
[12]   Biomechanical evaluation of the interfacial strength of a chemically modified sandblasted and acid-etched titanium surface [J].
Ferguson, S. J. ;
Broggini, N. ;
Wieland, M. ;
de Wild, M. ;
Rupp, F. ;
Geis-Gerstorfer, J. ;
Cochran, D. L. ;
Buser, D. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2006, 78A (02) :291-297
[13]   ENVIRONMENTAL APPLICATIONS OF SEMICONDUCTOR PHOTOCATALYSIS [J].
HOFFMANN, MR ;
MARTIN, ST ;
CHOI, WY ;
BAHNEMANN, DW .
CHEMICAL REVIEWS, 1995, 95 (01) :69-96
[14]   Enhanced initial cell responses to chemically modified anodized titanium [J].
Jimbo, Ryo ;
Sawase, Takashi ;
Baba, Koumei ;
Kurogi, Tadafumi ;
Shibata, Yasuaki ;
Atsuta, Mitsuru .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2008, 10 (01) :55-61
[15]   Enhanced osseointegration by the chemotactic activity of plasma fibronectin for cellular fibronectin positive cells [J].
Jimbo, Ryo ;
Sawase, Takashi ;
Shibata, Yasuaki ;
Hirata, Kazunari ;
Hishikawa, Yoshitaka ;
Tanaka, Yasuhiro ;
Bessho, Kazuhisa ;
Ikeda, Tohru ;
Atsuta, Mitsuru .
BIOMATERIALS, 2007, 28 (24) :3469-3477
[16]   Adsorption and dissolution behavior of human plasma fibronectin on thermally and chemically modified titanium dioxide particles [J].
MacDonald, DE ;
Deo, N ;
Markovic, B ;
Stranick, M ;
Somasundaran, P .
BIOMATERIALS, 2002, 23 (04) :1269-1279
[17]  
Ohko Y, 2001, J BIOMED MATER RES, V58, P97, DOI 10.1002/1097-4636(2001)58:1<97::AID-JBM140>3.0.CO
[18]  
2-8
[19]  
Orsini G, 2000, INT J ORAL MAX IMPL, V15, P779
[20]   Cortical bone healing following laser osteotomy using 6.1 μm wavelength [J].
Payne, JT ;
Peavy, GM ;
Reinisch, L ;
Van Sickle, DC .
LASERS IN SURGERY AND MEDICINE, 2001, 29 (01) :38-43