Stent expansion in curved vessel and their interactions: A finite element analysis

被引:110
作者
Wu, Wei [1 ]
Wang, Wei-Qiang [1 ]
Yang, Da-Zhi [1 ]
Qi, Min [1 ]
机构
[1] Dalian Univ Technol, Dept Mat Engn, Dalian 116024, Peoples R China
基金
中国国家自然科学基金;
关键词
curved vessel; coronary stent; finite element methods; in-stent restenosis; mechanical properties;
D O I
10.1016/j.jbiomech.2006.11.009
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Coronary restenosis after angioplasty has been reduced by stenting procedure, but in-stent restenosis (ISR) has not been eliminated yet, especially in tortuous vessels. In this paper, we proposed a finite element method (FEM) to study the expansion of a stent in a curved vessel (the CV model) and their interactions. A model of the same stent in a straight vessel (the SV model) was also studied and mechanical parameters of both models were researched and compared, including final lumen area, tissue prolapse between stent struts and stress distribution. Results show that in the CV model, the vessel was straightened by stenting and a hinge effect can be observed at extremes of the stent. The maximum tissue prolapse of the CV model was more severe (0.079 mm) than the SV model (0.048 mm), and the minimum lumen area of the CV was decreased (6.10 mm(2)), compared to that of the SV model (6.28 mm(2)). Tissue stresses of the highest level were concentrated in the inner curvature of the CV model. The simulations offered some explanations for the clinical results of ISR in curved vessels and gave design suggestions of the stent and balloon for tortuous vessels. This FEM provides a tool to study mechanisms of stents in curved vessels and can improve new stent designs especially for tortuous vessels. (C) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2580 / 2585
页数:6
相关论文
共 27 条
  • [1] Abhyankar AD, 1997, CATHETER CARDIO DIAG, V40, P261
  • [2] The cardiocoil stent-artery interaction
    Brand, M
    Ryvkin, M
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (02): : 337 - 344
  • [3] Computational hemodynamics of an implanted coronary Stent based on three-dimensional Cine angiography reconstruction
    Chen, MCY
    Lu, PC
    Chen, JSY
    Hwang, NHC
    [J]. ASAIO JOURNAL, 2005, 51 (04) : 313 - 320
  • [4] Selection of coronary stents
    Colombo, A
    Stankovic, G
    Moses, JW
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 40 (06) : 1021 - 1033
  • [5] Pathological mechanisms of fatal late coronary stent thrombosis in humans
    Farb, A
    Burke, AP
    Kolodgie, FD
    Virmani, R
    [J]. CIRCULATION, 2003, 108 (14) : 1701 - 1706
  • [6] Patterns and mechanisms of in-stent restenosis - A serial intravascular ultrasound study
    Hoffmann, R
    Mintz, GS
    Dussaillant, GR
    Popma, JJ
    Pichard, AD
    Satler, LF
    Kent, KM
    Griffin, J
    Leon, MB
    [J]. CIRCULATION, 1996, 94 (06) : 1247 - 1254
  • [7] Changes in the mechanical environment of stenotic arteries during interaction with stents: Computational assessment of parametric stent designs
    Holzapfel, G
    Stadler, M
    Gasser, TC
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (01): : 166 - 180
  • [8] Non-Newtonian blood flow in human right coronary arteries: Transient simulations
    Johnston, BM
    Johnston, PR
    Corney, S
    Kilpatrick, D
    [J]. JOURNAL OF BIOMECHANICS, 2006, 39 (06) : 1116 - 1128
  • [9] Radial force and wall apposition of balloon-expandable vascular stents in eccentric stenoses:: An in vitro evaluation in a curved vessel model
    Kalmár, G
    Hübner, F
    Voelker, W
    Hutzenlaub, J
    Teubner, J
    Poerner, T
    Süselbeck, T
    Borggrefe, M
    Haase, KK
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (05) : 499 - 508
  • [10] Circumferential vascular deformation after stent implantation alters wall shear stress evaluated with time-dependent 3D computational fluid dynamics models
    LaDisa, JF
    Olson, LE
    Guler, I
    Hettrick, DA
    Kersten, JR
    Warltier, DC
    Pagel, PS
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2005, 98 (03) : 947 - 957