A 6 DoF, Wearable, Compliant Shoe Sensor for Total Ground Reaction Measurement

被引:16
作者
Eng, Stephanie [1 ]
Al-Mai, Osama [1 ]
Ahmadi, Mojtaba [1 ]
机构
[1] Carleton Univ, Dept Mech & Aerosp Engn, Ottawa, ON K1S 5B6, Canada
关键词
Calibration; force measurement; force sensors; pressure measurement; sensor systems and applications; sensors; sensors systems; PLANTAR PRESSURE MEASUREMENT; REACTION FORCE; MEASUREMENT SYSTEM; BALANCE ASSESSMENT; FREQUENCY CONTENT; OLDER-ADULTS; GAIT; CALIBRATION; FALLS;
D O I
10.1109/TIM.2018.2829338
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, we present a compliant six-axis sensor for total ground reaction measurement, the calibration methodology, and preliminary data. The sensor is intended for gait analysis and is designed to have minimal effect on natural gait. A triaxial optical force sensor is combined with an array of pressure sensing films to form a wearable compliant six-axis force/moment sensor. Two sensor units were developed for the toe and heel and used in two types of experiments: stepping on the sensors that are mounted on the ground and attaching the sensors under the shoe while walking. The data from the sensors are compared with measurements obtained from a standard force plate. The deflections induced by the sensor compliance exhibit a slight nonlinear force-deflection relation. Regardless of the nonlinear effects, the sensor is accurately calibrated with a linear least squares method. To see how well these nonlinearities could be calibrated for, a nonlinear calibration with a neural network was used. For the sensors attached to the floor, a linear calibration achieved an RMSE of 4.49% while the neural network achieves 2.68%. For the wearable sensor, the linear calibration RMSE was 9.39% and the neural network RMSE was 5.21%. When the orientation data of the sensors (measured by a motion capture system) were added as an input to the neural network calibration, the RMSE was reduced to 3.25%.
引用
收藏
页码:2714 / 2722
页数:9
相关论文
共 26 条
[1]   A Compliant 3-Axis Fiber-Optic Force Sensor for Biomechanical Measurement [J].
Al-Mai, Osama ;
Ahmadi, Mojtaba ;
Albert, Jacques .
IEEE SENSORS JOURNAL, 2017, 17 (20) :6549-6557
[2]   THE FREQUENCY CONTENT OF GAIT [J].
ANTONSSON, EK ;
MANN, RW .
JOURNAL OF BIOMECHANICS, 1985, 18 (01) :39-47
[3]   The direct costs of fatal and non-fatal falls among older adults - United States [J].
Burns, Elizabeth R. ;
Stevens, Judy A. ;
Lee, Robin .
JOURNAL OF SAFETY RESEARCH, 2016, 58 :99-103
[4]   Detection of gait events using an F-Scan in-shoe pressure measurement system [J].
Catalfamo, Paola ;
Moser, David ;
Ghoussayni, Salim ;
Ewins, David .
GAIT & POSTURE, 2008, 28 (03) :420-426
[5]   Estimating the complete ground reaction forces with pressure insoles in walking [J].
Fong, Daniel Tik-Pui ;
Chan, Yue-Yan ;
Hong, Youlian ;
Yung, Patrick Shu-Hang ;
Fung, Kwai-Yau ;
Chan, Kai-Ming .
JOURNAL OF BIOMECHANICS, 2008, 41 (11) :2597-2601
[6]   Quantification of the path of center of pressure (COP) using an F-scan in-shoe transducer [J].
Han, TR ;
Paik, NJ ;
Im, MS .
GAIT & POSTURE, 1999, 10 (03) :248-254
[7]   Measuring of foot plantar pressure-possible applications in quantitative analysis of human body mobility [J].
Klimiec, E. ;
Jasiewicz, B. ;
Piekarski, J. ;
Zaraska, K. ;
Guzdek, P. ;
Kolaszczynski, G. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2017, 28 (05)
[8]   FACTORS DETERMINING THE FREQUENCY CONTENT OF THE ELECTROMYOGRAM [J].
KRANZ, H ;
WILLIAMS, AM ;
CASSELL, J ;
CADDY, DJ ;
SILBERSTEIN, RB .
JOURNAL OF APPLIED PHYSIOLOGY, 1983, 55 (02) :392-399
[9]   Reliability and validity of center of pressure measures for balance assessment in older adults [J].
Li, Zhen ;
Liang, Yan-Yi ;
Wang, Lei ;
Sheng, Jing ;
Ma, Shao-Jun .
JOURNAL OF PHYSICAL THERAPY SCIENCE, 2016, 28 (04) :1364-1367
[10]   Wearable force sensor with parallel structure for measurement of ground-reaction force [J].
Liu, Tao ;
Inoue, Yoshio ;
Shibata, Kyoko .
MEASUREMENT, 2007, 40 (06) :644-653