A deflectable guiding catheter for real-time MRI-guided interventions

被引:29
|
作者
Bell, Jamie A. [1 ]
Saikus, Christina E. [1 ]
Ratnayaka, Kanishka [1 ,2 ]
Wu, Vincent [1 ]
Sonmez, Merdim [1 ]
Faranesh, Anthony Z. [1 ]
Colyer, Jessica H. [1 ,2 ]
Lederman, Robert J. [1 ]
Kocaturk, Ozgur [1 ]
机构
[1] NHLBI, Cardiovasc & Pulm Branch, Div Intramural Res, NIH, Bethesda, MD 20892 USA
[2] Childrens Natl Med Ctr, Div Cardiol, Washington, DC 20010 USA
基金
美国国家卫生研究院;
关键词
interventional MRI; structural heart disease; catheter engineering; active MRI catheters; REMOTE-CONTROL; TIP; OPPORTUNITY; DESIGN; MODEL;
D O I
10.1002/jmri.23520
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To design a deflectable guiding catheter that omits long metallic components yet preserves mechanical properties to facilitate therapeutic interventional MRI procedures. Materials and Methods: The catheter shaft incorporated Kevlar braiding. A 180 degrees deflection was attained with a 5-cm nitinol slotted tube, a nitinol spring, and a Kevlar pull string. We tested three designs: passive, passive incorporating an inductively coupled coil, and active receiver. We characterized mechanical properties, MRI properties, RF induced heating, and in vivo performance in swine. Results: Torque and tip deflection force were satisfactory. Representative procedures included hepatic and azygos vein access, laser cardiac septostomy, and atrial septal defect crossing. Visualization was best in the active configuration, delineating profile and tip orientation. The passive configuration could be used in tandem with an active guidewire to overcome its limited conspicuity. There was no RF-induced heating in all configurations under expected use conditions in vitro and in vivo. Conclusion: Kevlar and short nitinol component substitutions preserved mechanical properties. The active design offered the best visibility and usability but reintroduced metal conductors. We describe versatile deflectable guiding catheters with a 0.057'' lumen for interventional MRI catheterization. Implementations are feasible using active, inductive, and passive visualization strategies to suit application requirements.
引用
收藏
页码:908 / 915
页数:8
相关论文
共 50 条
  • [41] Real-time radial reconstruction with domain transform manifold learning for MRI-guided radiotherapy
    Waddington, David E. J.
    Hindley, Nicholas
    Koonjoo, Neha
    Chiu, Christopher
    Reynolds, Tess
    Liu, Paul Z. Y.
    Zhu, Bo
    Bhutto, Danyal
    Paganelli, Chiara
    Keall, Paul J. J.
    Rosen, Matthew S. S.
    MEDICAL PHYSICS, 2023, 50 (04) : 1962 - 1974
  • [42] Comparing Transformer training strategies for real-time tumor tracking in MRI-guided radiotherapy
    Lombardo, Elia
    Velezmoro, Laura
    Marschner, Sebastian
    Reiner, Michael
    Corradini, Stefanie
    Belka, Claus
    Riboldi, Marco
    Kurz, Christopher
    Landry, Guillaume
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4286 - S4288
  • [43] Real-time MRI-guided right heart catheterization in adults using passive catheters
    Ratnayaka, Kanishka
    Faranesh, Anthony Z.
    Hansen, Michael S.
    Stine, Annette M.
    Halabi, Majdi
    Barbash, Israel M.
    Schenke, William H.
    Wright, Victor J.
    Grant, Laurie P.
    Kellman, Peter
    Kocaturk, Ozgur
    Lederman, Robert J.
    EUROPEAN HEART JOURNAL, 2013, 34 (05) : 380 - 389
  • [44] Safety of MRI-guided vascular interventions
    Buecker, Arno
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (02) : 65 - 70
  • [45] MRI-guided interventions of the musculoskeletal system
    Smith, Kevin A.
    Carrino, John
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (02) : 339 - 346
  • [46] MR systems for MRI-guided interventions
    Hushek, Stephen G.
    Martin, Alastair J.
    Steckner, Michael
    Bosak, Elyakim
    Debbins, Josef
    Kucharzyk, Walter
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (02) : 253 - 266
  • [47] Quantitative analysis of MRI-guided radiotherapy treatment process time for tumor real-time gating efficiency
    Placidi, Lorenzo
    Cusumano, Davide
    Boldrini, Luca
    Votta, Claudio
    Pollutri, Veronica
    Antonelli, Marco Valerio
    Chiloiro, Giuditta
    Romano, Angela
    De Luca, Viola
    Catucci, Francesco
    Indovina, Luca
    Valentini, Vincenzo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 70 - 79
  • [48] A Stacked Multi-Sensor Platform for Real-Time MRI Guided Interventions
    Zolfaghari, Parviz
    Erden, Oguz K.
    Tumer, Murat
    Yalcinkaya, Arda D.
    Ferhanoglu, Onur
    OPTICS AND LASERS IN ENGINEERING, 2023, 161
  • [49] Phase II Study of Real-Time MRI-Guided Adaptive Accelerated Partial Breast Irradiation
    Harari, Colin
    Yadav, Poonam
    Glide-Hurst, Carri
    Anderson, Bethany
    AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2024, 47 (10): : S37 - S38
  • [50] Real-time probe tracking using EM-optical sensor for MRI-guided cryoablation
    Gao, Wenpeng
    Jiang, Baichuan
    Kacher, Daniel F.
    Fetics, Barry
    Nevo, Erez
    Lee, Thomas C.
    Jayender, Jagadeesan
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2018, 14 (01):