Accuracy of pressure and flow capacities of four arthroscopic fluid management systems

被引:23
作者
Muellner, T
Menth-Chiari, WA
Reihsner, R
Eberhardsteiner, J
Engebretsen, L
机构
[1] Univ Vienna, Sch Med, Ctr Trauma, Vienna, Austria
[2] Oslo Orthopaed Univ Clin, Oslo, Norway
[3] Vienna Univ Technol, Inst Strenght Mat, Vienna, Austria
关键词
arthroscopic fluid delivery systems; infusion pumps; pressure characteristics; flow characteristics; experimental study;
D O I
10.1053/jars.2001.25340
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: The present study was conducted to evaluate 4 different fluid delivery systems: Arthrex AR-6450, Stryker 1.5L High Flow Pump, Arthro FMS 4, and Acufex InteliJet. Their flow and pressure capacity, as specified by the manufacturer, and their accuracy, as measured in a standardized fashion, were compared with directly measured pressure values. Type of Study: Experimental study. Methods: Two experiments were performed: (1) the achieved pressure was measured with the pressure sensor within a ball and compared with the preset pressure values. The pressure was increased constantly while the flow was kept constant. (2) maximum flow per minute was tested using a beaker and a stopwatch. Results: The highest measured pressures for the Arthrex and the Stryker pumps were 115 mm Hg (46% of published maximum pressure) and 113 mm Hg (57% of published maximum pressure), respectively. The highest measured pressures for the Arthro FMS 4 and the InteliJet pumps were 251 mm Hg (101% of published pressure) and 132 mm Hg (88% of published maximum pressure), respectively. The maximal flow values for the Arthrex and Stryker pumps were 900 mL/min (56.3%) and 675 mL/min (48%), respectively. The Arthro FMS 4 and InteliJet pumps reached a maximal flow of 450 mL/min (56%) and 1,250 mL/min (62.5%). Conclusions: In the simpler pumps (Arthrex AR-6450 and Stryker 1.5L High Flow Pump), the pressure on the display must be set to higher values compared with the outflow-controlling pumps (Arthro FMS 4 an InteliJet) to achieve the same intra-articular pressure. All pumps were able to maintain a pressure of 60 mm Hg very accurately. Therefore, a surgeon can trust all of the pumps when the pressure is set below 60 mm Hg.
引用
收藏
页码:760 / 764
页数:5
相关论文
共 18 条
  • [1] ABBUSHI W, 1991, ANASTH INTENSIVMED, V266, P348
  • [2] INTRAARTICULAR PRESSURES IN A GRAVITY-FED ARTHROSCOPY FLUID DELIVERY SYSTEM
    ARANGIO, G
    KOSTELNIK, KE
    [J]. ARTHROSCOPY, 1992, 8 (03): : 341 - 344
  • [3] Cardiac arrest as a result of intraabdominal extravasation of fluid during arthroscopic removal of a loose body from the hip joint of a patient with an acetabular fracture
    Bartlett, CS
    DiFelice, GS
    Buly, RL
    Quinn, TJ
    Green, DST
    Helfet, DL
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 1998, 12 (04) : 294 - 299
  • [4] Compartment syndrome of the leg after arthroscopic examination of a tibial Plateau fracture. Case report and review of the literature
    Belanger, M
    Fadale, P
    [J]. ARTHROSCOPY, 1997, 13 (05): : 646 - 651
  • [5] COMPLICATIONS ASSOCIATED WITH THE USE OF AN INFUSION PUMP DURING KNEE ARTHROSCOPY
    BOMBERG, BC
    HURLEY, PE
    CLARK, CA
    MCLAUGHLIN, CS
    [J]. ARTHROSCOPY, 1992, 8 (02): : 224 - 228
  • [6] RUPTURE OF THE KNEE CAPSULE FROM ARTICULAR HYPERPRESSURE - EXPERIMENTS IN CADAVER KNEES
    BURGAARD, P
    BLYME, PJH
    OLSEN, PMA
    KRISTENSEN, G
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1988, 59 (06): : 692 - 694
  • [7] COMPLICATIONS ASSOCIATED WITH THE USE OF AN INFUSION PUMP DURING KNEE ARTHROSCOPY
    CARPENTER, JF
    [J]. ARTHROSCOPY, 1993, 9 (02): : 244 - 244
  • [8] DOLK T, 1989, ARTHROSCOPY, V54, P307
  • [9] An experimental assessment of the risk of compartment syndrome during knee arthroscopy
    Ekman, EF
    Poehling, GG
    [J]. ARTHROSCOPY, 1996, 12 (02): : 193 - 199
  • [10] EWING JW, 1986, ARTHROSCOPY, V24, P264