Parameter optimization for reproducible cardiac 1H-MR spectroscopy at 3 Tesla

被引:16
|
作者
de Heer, Paul [1 ]
Bizino, Maurice B. [2 ]
Lamb, Hildo J. [2 ]
Webb, Andrew G. [1 ]
机构
[1] Leiden Univ, Med Ctr, CJ Gorter Ctr High Field MR, Dept Radiol, Leiden, Netherlands
[2] Leiden Univ, Med Ctr, Dept Radiol, Leiden, Netherlands
基金
欧洲研究理事会;
关键词
cardiac proton magnetic resonance spectroscopy; reproducibility; triglycerides; myocardial triglyceride content; motion compensation; water suppression; MAGNETIC-RESONANCE-SPECTROSCOPY; MYOCARDIAL TRIGLYCERIDE CONTENT; LOCALIZED PROTON SPECTROSCOPY; DIABETES-MELLITUS; HUMAN HEART; QUANTIFICATION; SUPPRESSION; HUMANS; 1.5T;
D O I
10.1002/jmri.25254
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo optimize data acquisition parameters in cardiac proton MR spectroscopy, and to evaluate the intra- and intersession variability in myocardial triglyceride content. Materials and MethodsData acquisition parameters at 3 Tesla (T) were optimized and reproducibility measured using, in total, 49 healthy subjects. The signal-to-noise-ratio (SNR) and the variance in metabolite amplitude between averages were measured for: (i) global versus local power optimization; (ii) static magnetic field (B-0) shimming performed during free-breathing or within breathholds; (iii) post R-wave peak measurement times between 50 and 900 ms; (iv) without respiratory compensation, with breathholds and with navigator triggering; and (v) frequency selective excitation, Chemical Shift Selective (CHESS) and Multiply Optimized Insensitive Suppression Train (MOIST) water suppression techniques. Using the optimized parameters intra- and intersession myocardial triglyceride content reproducibility was measured. Two cardiac proton spectra were acquired with the same parameters and compared (intrasession reproducibility) after which the subject was removed from the scanner and placed back in the scanner and a third spectrum was acquired which was compared with the first measurement (intersession reproducibility). ResultsLocal power optimization increased SNR on average by 22% compared with global power optimization (P = 0.0002). The average linewidth was not significantly different for pencil beam B-0 shimming using free-breathing or breathholds (19.1 Hz versus 17.5 Hz; P = 0.15). The highest signal stability occurred at a cardiac trigger delay around 240 ms. The mean amplitude variation was significantly lower for breathholds versus free-breathing (P = 0.03) and for navigator triggering versus free-breathing (P = 0.03) as well as for navigator triggering versus breathhold (P = 0.02). The mean residual water signal using CHESS (1.1%, P = 0.01) or MOIST (0.7%, P = 0.01) water suppression was significantly lower than using frequency selective excitation water suppression (7.0%). Using the optimized parameters an intrasession limits of agreement of the myocardial triglyceride content of -0.11% to +0.04%, and an intersession of -0.15% to +0.9%, were achieved. The coefficient of variation was 5% for the intrasession reproducibility and 6.5% for the intersession reproducibility. ConclusionUsing approaches designed to optimize SNR and minimize the variation in inter-average signal intensities and frequencies/phases, a protocol was developed to perform cardiac MR spectroscopy on a clinical 3T system with high reproducibility. J. Magn. Reson. Imaging 2016;44:1151-1158.
引用
收藏
页码:1151 / 1158
页数:8
相关论文
共 50 条
  • [41] 1H-MR spectroscopy indicates prominent cerebellar dysfunction in benign adult familial myoclonic epilepsy
    Striano, Pasquale
    Caranci, Ferdinando
    Di Benedetto, Raffaella
    Tortora, Fabio
    Zara, Federico
    Striano, Salvatore
    EPILEPSIA, 2009, 50 (06) : 1491 - 1497
  • [42] Are there differences in the biochemical profile of bilateral normal testes? A 3.0 T 1H-MR spectroscopy study
    Tsili, Athina C.
    Astrakas, Loukas G.
    Sofikitis, Nikolaos
    Argyropoulou, Maria I.
    ANDROLOGIA, 2020, 52 (05)
  • [43] Quantitative Cardiac 31P Spectroscopy at 3 Tesla Using Adiabatic Pulses
    El-Sharkawy, AbdEl-Monem
    Schar, Michael
    Ouwerkerk, Ronald
    Weiss, Robert G.
    Bottomley, Paul A.
    MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (04) : 785 - 795
  • [44] Regional apparent metabolite concentrations in young adult brain measured by 1H MR spectroscopy at 3 tesla
    Baker, Eva H.
    Basso, Gianpaolo
    Barker, Peter B.
    Smith, Mari A.
    Bonekamp, David
    Horska, Alena
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (03) : 489 - 499
  • [45] Evaluation of cerebral glioma grade by using normal side creatine as an internal reference in multi-voxel 1H-MR spectroscopy
    Yerli, Hasan
    Agildere, Ahmet Muhtesem
    Ozen, Ozlem
    Geyik, Elif
    Atalay, Basar
    Elhan, Atilla Halil
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2007, 13 (01) : 3 - 9
  • [46] Succinate detection using in vivo 1H-MR spectroscopy identifies germline and somatic SDHx mutations in paragangliomas
    Charlotte Lussey-Lepoutre
    Alexandre Bellucci
    Nelly Burnichon
    Laurence Amar
    Alexandre Buffet
    Tom Drossart
    Sébastien Fontaine
    Olivier Clement
    Paule Benit
    Pierre Rustin
    Lionel Groussin
    Tchao Meatchi
    Anne-Paule Gimenez-Roqueplo
    Bertrand Tavitian
    Judith Favier
    European Journal of Nuclear Medicine and Molecular Imaging, 2020, 47 : 1510 - 1517
  • [47] Short- and long-term reproducibility of marrow adipose tissue quantification by 1H-MR spectroscopy
    Vibha Singhal
    Karen K. Miller
    Martin Torriani
    Miriam A. Bredella
    Skeletal Radiology, 2016, 45 : 221 - 225
  • [48] An investigation into potential gender-specific differences in myocardial triglyceride content assessed by 1H-Magnetic Resonance Spectroscopy at 3Tesla
    Petritsch, Bernhard
    Koestler, Herbert
    Gassenmaier, Tobias
    Kunz, Andreas S.
    Bley, Thorsten A.
    Horn, Michael
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2016, 44 (03) : 585 - 591
  • [49] Short- and long-term reproducibility of marrow adipose tissue quantification by 1H-MR spectroscopy
    Singhal, Vibha
    Miller, Karen K.
    Torriani, Martin
    Bredella, Miriam A.
    SKELETAL RADIOLOGY, 2016, 45 (02) : 221 - 225
  • [50] 7.0 T high-resolution 1H-MR spectroscopy of metabolic changes induced by chronic codeine phosphate in rat hippocampus
    Cao, Zhen
    Lin, Pei-yin
    Shen, Zhi-wei
    Xiao, Ye-yu
    Wu, Ren-hua
    NEUROREPORT, 2015, 26 (13) : 735 - 739