Parameter optimization for reproducible cardiac 1H-MR spectroscopy at 3 Tesla

被引:16
|
作者
de Heer, Paul [1 ]
Bizino, Maurice B. [2 ]
Lamb, Hildo J. [2 ]
Webb, Andrew G. [1 ]
机构
[1] Leiden Univ, Med Ctr, CJ Gorter Ctr High Field MR, Dept Radiol, Leiden, Netherlands
[2] Leiden Univ, Med Ctr, Dept Radiol, Leiden, Netherlands
基金
欧洲研究理事会;
关键词
cardiac proton magnetic resonance spectroscopy; reproducibility; triglycerides; myocardial triglyceride content; motion compensation; water suppression; MAGNETIC-RESONANCE-SPECTROSCOPY; MYOCARDIAL TRIGLYCERIDE CONTENT; LOCALIZED PROTON SPECTROSCOPY; DIABETES-MELLITUS; HUMAN HEART; QUANTIFICATION; SUPPRESSION; HUMANS; 1.5T;
D O I
10.1002/jmri.25254
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo optimize data acquisition parameters in cardiac proton MR spectroscopy, and to evaluate the intra- and intersession variability in myocardial triglyceride content. Materials and MethodsData acquisition parameters at 3 Tesla (T) were optimized and reproducibility measured using, in total, 49 healthy subjects. The signal-to-noise-ratio (SNR) and the variance in metabolite amplitude between averages were measured for: (i) global versus local power optimization; (ii) static magnetic field (B-0) shimming performed during free-breathing or within breathholds; (iii) post R-wave peak measurement times between 50 and 900 ms; (iv) without respiratory compensation, with breathholds and with navigator triggering; and (v) frequency selective excitation, Chemical Shift Selective (CHESS) and Multiply Optimized Insensitive Suppression Train (MOIST) water suppression techniques. Using the optimized parameters intra- and intersession myocardial triglyceride content reproducibility was measured. Two cardiac proton spectra were acquired with the same parameters and compared (intrasession reproducibility) after which the subject was removed from the scanner and placed back in the scanner and a third spectrum was acquired which was compared with the first measurement (intersession reproducibility). ResultsLocal power optimization increased SNR on average by 22% compared with global power optimization (P = 0.0002). The average linewidth was not significantly different for pencil beam B-0 shimming using free-breathing or breathholds (19.1 Hz versus 17.5 Hz; P = 0.15). The highest signal stability occurred at a cardiac trigger delay around 240 ms. The mean amplitude variation was significantly lower for breathholds versus free-breathing (P = 0.03) and for navigator triggering versus free-breathing (P = 0.03) as well as for navigator triggering versus breathhold (P = 0.02). The mean residual water signal using CHESS (1.1%, P = 0.01) or MOIST (0.7%, P = 0.01) water suppression was significantly lower than using frequency selective excitation water suppression (7.0%). Using the optimized parameters an intrasession limits of agreement of the myocardial triglyceride content of -0.11% to +0.04%, and an intersession of -0.15% to +0.9%, were achieved. The coefficient of variation was 5% for the intrasession reproducibility and 6.5% for the intersession reproducibility. ConclusionUsing approaches designed to optimize SNR and minimize the variation in inter-average signal intensities and frequencies/phases, a protocol was developed to perform cardiac MR spectroscopy on a clinical 3T system with high reproducibility. J. Magn. Reson. Imaging 2016;44:1151-1158.
引用
收藏
页码:1151 / 1158
页数:8
相关论文
共 50 条
  • [11] 1H-MR Spectroscopy in the Human Spinal Cord
    Hock, A.
    Henning, A.
    Boesiger, P.
    Kollias, S. S.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (09) : 1682 - 1689
  • [12] Elucidating Metabolic Maturation in the Healthy Fetal Brain Using 1H-MR Spectroscopy
    Evangelou, I. E.
    du Plessis, A. J.
    Vezina, G.
    Noeske, R.
    Limperopoulos, C.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (02) : 360 - 366
  • [13] In vivo 1H-MR spectroscopy of the human heart
    Kreis, Roland
    Felblinger, Jacques
    Jung, Bruno
    Boesch, Chris
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 1998, 6 (2-3) : 164 - 167
  • [14] Brain 1H-MR spectroscopy in clinical neuroimaging at 3T
    Tchofo, P. Jissendi
    Baleriaux, D.
    JOURNAL OF NEURORADIOLOGY, 2009, 36 (01) : 24 - 40
  • [15] Methods and reproducibility of cardiac/respiratory double-triggered 1H-MR spectroscopy of the human heart
    Felblinger, J
    Jung, B
    Slotboom, J
    Boesch, C
    Kreis, R
    MAGNETIC RESONANCE IN MEDICINE, 1999, 42 (05) : 903 - 910
  • [16] Reproducibility of Cerebral Phenylalanine Levels in Patients With Phenylketonuria Determined by 1H-MR Spectroscopy
    Kreis, R.
    Zwygart, K.
    Boesch, C.
    Nuoffer, J-M.
    MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (01) : 11 - 16
  • [17] The influence of cardiac triggering time and an optimization strategy for improved cardiac MR spectroscopy
    Carlsson, Asa
    Sohlin, Maja C.
    Lagerstrand, Kerstin M.
    Aronsson, Eva Forsell
    Ljungberg, Maria
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (04): : 310 - 317
  • [18] Reproducibility of relaxometry of human lumbar vertebrae at 3 Tesla using 1H MR spectroscopy
    Neumayer, Bernhard
    Widek, Thomas
    Stollberger, Rudolf
    Scheurer, Eva
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2018, 48 (01) : 153 - 159
  • [19] 1H-MR spectroscopy of adult-onset autosomal dominant leukodystrophy with autonomic symptoms
    Finnsson, J.
    Melberg, A.
    Raininko, R.
    NEURORADIOLOGY, 2013, 55 (08) : 933 - 939
  • [20] Detection of Liquid Phase Cholesteryl Ester in Carotid Atherosclerosis by 1H-MR Spectroscopy in Humans
    Duivenvoorden, Raphael
    van Wijk, Diederik
    Klimas, Michael
    Kastelein, John J. P.
    Stroes, Erik S. G.
    Nederveen, Aart J.
    JACC-CARDIOVASCULAR IMAGING, 2013, 6 (12) : 1277 - 1284