Performance of heated wire humidifiers: An in vitro study

被引:17
作者
Pelosi, Paolo [2 ,3 ]
Chiumello, Davide [1 ]
Severgnini, Paolo [3 ]
De Grandis, Cinzia Elisabetta [1 ]
Landi, Laura [1 ]
Chierichetti, Laura Maria [2 ,3 ]
Frigerio, Alice [2 ,3 ]
Munaro, Marco [3 ]
Park, Gilbert [4 ]
Chiaranda, Maurizio [2 ,3 ]
机构
[1] Univ Milan, Fondaz Osped Maggiore Policlin, IRCCS, Inst Anesthesia & Crit Care, Milan, Italy
[2] Univ Insubria Circolo, Dept Environm Hlth Safety, Varese, Italy
[3] Fondaz Macchi Hosp, Varese, Italy
[4] Addenbrookes Hosp, Dept Intens Care Res, Cambridge, England
关键词
condensation; heated wire humidifier; mechanical ventilation;
D O I
10.1016/j.jcrc.2006.12.009
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Purpose: We assessed the performance of heated wire humidifiers (HWHs), which should avoid water condensation in the circuit. Methods: We evaluated the efficiency of 3 HWEs, MR850 (Fisher & Paykel, Auckland, New Zealand), CONCHATHERM IV (Hudson RCI, Temecula, Calif), and DAR HC 2000 (Mallinckrodt DAR, Mirandola, Italy), in comparison with that of the MR730 heated humidifier (HH), which has a standard circuit. We measured gas temperature and absolute humidity (AH) at the Y piece of the ventilatory circuit using a test lung ventilated at 2 minute ventilation volumes (5 and 15 L/min). Temperature levels at the Y piece of the ventilatory circuit of the HHs were set at 35 degrees C, 37 degrees C, and 39 degrees C with different gradients (-2 degrees C, 0 degrees C, and +2 degrees C) between the outlet chamber and the Y piece of the ventilatory circuit. Results: At the set temperature levels of 35 degrees C, 37 degrees C, and 39 degrees C with a gradient of 0 degrees C, the MR850 and CONCHATHERM IV had lower gas temperature and AH levels as compared with the DAR HC 2000 and MR730 HH. With increasing temperature gradient, gas temperature increased only with the CONCHATHERM IV but AH increased with all the HWHs. The MR850 showed lower gas temperature and AH levels as compared with CONCHATHERM IV The condensate was abolished inside the inspiratory circuit with the HWHs. Conclusions: Heated wire humidifiers eliminate water condensation but present significant differences in gas temperature and AH levels that are lower than the expected settings. (C) 2007 Elsevier Inc. All rights reserved.
引用
收藏
页码:258 / 264
页数:7
相关论文
共 24 条
[21]   Energy balance in the intubated human airway is an indicator of optimal gas conditioning [J].
Ryan, SN ;
Rankin, N ;
Meyer, E ;
Williams, R .
CRITICAL CARE MEDICINE, 2002, 30 (02) :355-361
[22]   A REVIEW OF THE MECHANISMS AND METHODS OF HUMIDIFICATION OF INSPIRED GASES [J].
SHELLY, MP ;
LLOYD, GM ;
PARK, GR .
INTENSIVE CARE MEDICINE, 1988, 14 (01) :1-9
[23]   Heat and moisture exchangers in mechanically ventilated intensive care unit patients:: A plea for an independent assessment of their performance [J].
Thiéry, G ;
Boyer, A ;
Pigné, E ;
Salah, A ;
de Lassence, A ;
Dreyfuss, D ;
Ricard, JD .
CRITICAL CARE MEDICINE, 2003, 31 (03) :699-704
[24]   Relationship between the humidity and temperature of inspired gas and the function of the airway mucosa [J].
Williams, R ;
Rankin, N ;
Smith, T ;
Galler, D ;
Seakins, P .
CRITICAL CARE MEDICINE, 1996, 24 (11) :1920-1929