Hemolysis associated with Impella heart pump positioning: In vitro hemolysis testing and computational fluid dynamics modeling

被引:37
作者
Roberts, Nicholas [1 ]
Chandrasekaran, Uma [1 ]
Das, Soumen [1 ]
Qi, Zhongwei [1 ]
Corbett, Scott [1 ]
机构
[1] Abiomed Inc, 22 Cherry Hill Dr, Danvers, MA 01923 USA
关键词
Mechanical circulatory device; heart pump; Impella CP; heart pump positioning; hemolysis; BLOOD DAMAGE; SHEAR-STRESS; FAILURE;
D O I
10.1177/0391398820909843
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction: Short-term mechanical circulatory support devices provide temporary hemodynamic support in heart failure and are increasingly used to enable recovery or as a bridge to decision. Blood damage with mechanical circulatory support devices is influenced by many factors, including the magnitude and duration of shear stress and obstruction to blood flow. This study aimed to evaluate the effects of the Impella CP (R) heart pump positioning on hemolysis using in vitro hemolysis testing and computational fluid dynamics modeling. Methods: The in vitro hemolysis testing was conducted per the recommended Food and Drug Administration and American Society for Testing and Materials guidelines. The bench hemolysis testing and computational fluid dynamics simulation analysis were performed for both normal operating (outlet unobstructed) and outlet-obstructed condition of Impella CP (mimicking outlet on the aortic valve due to improper positioning). Results: The modified index of hemolysis was 2.78 +/- 0.69 at normal operating conditions compared to 18.7 +/- 7.8 when the Impella CP outlet was obstructed (p = 0.002). Computational fluid dynamics modeling showed about three times increase in exposure time to regions of high shear stress when the Impella CP outlet was obstructed compared to unobstructed condition, thus supporting the experimental observations. Conclusion: Based on these results, it is recommended to ensure proper placement of Impella CP via regular monitoring using echocardiographic guidance or other methods to minimize the risk of hemolysis associated with an obstructed outflow.
引用
收藏
页码:710 / 718
页数:9
相关论文
共 25 条
  • [1] Benefits of a novel percutaneous ventricular assist device for right heart failure: The prospective RECOVER RIGHT study of the Impel la RP device
    Anderson, Mark B.
    Goldstein, James
    Milano, Carmelo
    Morris, Lynn D.
    Kormos, Robert L.
    Bhama, Jay
    Kapur, Navin K.
    Barisal, Aditya
    Garcia, Jose
    Baker, Joshua N.
    Silvestry, Scott
    Holman, William L.
    Douglas, Pamela S.
    O'Neill, William
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2015, 34 (12) : 1549 - 1560
  • [2] Incidence of Hemolysis in Patients with Cardiogenic Shock Treated with Impella Percutaneous Left Ventricular Assist Device
    Badiye, Amit P.
    Hernandez, Gabriel A.
    Novoa, Italo
    Chaparro, Sandra V.
    [J]. ASAIO JOURNAL, 2016, 62 (01) : 11 - 14
  • [3] Notes on the role of viscosity, vorticity and dissipation in incompressible flows
    Buresti, Guido
    [J]. MECCANICA, 2009, 44 (04) : 469 - 487
  • [4] A randomized multicenter clinical study to evaluate the safety and efficacy of the TandemHeart percutaneous ventricular assist device versus conventional therapy with intraaortic balloon pumping for treatment of cardiogenic shock
    Burkhoff, Daniel
    Cohen, Howard
    Brunckhorst, Corinna
    O'Neill, William W.
    [J]. AMERICAN HEART JOURNAL, 2006, 152 (03) : 469.e1 - 469.e8
  • [5] Impella Induced Massive Hemolysis: Reemphasizing Echocardiographic Guidance for Correct Placement
    Cardozo, Shaun
    Ahmed, Tasneem
    Belgrave, Kevin
    [J]. CASE REPORTS IN CARDIOLOGY, 2015, 2015
  • [6] Increased Plasma-Free Hemoglobin Levels Identify Hemolysis in Patients With Cardiogenic Shock and a Trans valvular Micro-Axial Flow Pump
    Esposito, Michele L.
    Morine, Kevin J.
    Annamalai, Shiva K.
    O'Kelly, Ryan
    Aghili, Nima
    Pedicini, Robert
    Breton, Catalina
    Mullin, Andrew
    Hamadeh, Anas
    Kiernan, Michael S.
    DeNofrio, David
    Kapur, Navin K.
    [J]. ARTIFICIAL ORGANS, 2019, 43 (02) : 125 - 131
  • [7] Modeling and prediction of flow-induced hemolysis: a review
    Faghih, Mohammad M.
    Sharp, M. Keith
    [J]. BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2019, 18 (04) : 845 - 881
  • [8] A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index
    Fraser, Katharine H.
    Zhang, Tao
    Taskin, M. Ertan
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2012, 134 (08):
  • [9] Numerical estimation of blood damage in artificial organs
    Goubergrits, L
    Affeld, K
    [J]. ARTIFICIAL ORGANS, 2004, 28 (05) : 499 - 507
  • [10] Interagency Registry for Mechanically Assisted Circulatory Support National Heart Lung Blood Institute, 2013, APP ADV DEF INT ADV