Diagnostic reference levels (DRLs) for routine X-ray examinations in Lorestan province, Iran

被引:0
|
作者
Gholami, M. [1 ]
Maziar, A. [2 ]
Khosravi, H. R. [3 ]
Ebrahimzadeh, F. [4 ]
Mayahi, S. [1 ]
机构
[1] Lorestan Univ Med Sci, Fac Allied Med Sci, Dept Med Phys & Radiat Sci, Khorramabad, Iran
[2] Isfahan Univ Med Sci, Fac Med, Dept Med Phys & Med Engn, Esfahan, Iran
[3] Atom Energy Org Iran, Tehran, Iran
[4] Lorestan Univ Med Sci, Fac Hlth & Nutr, Dept Biostat, Khorramabad, Iran
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2015年 / 13卷 / 01期
关键词
Diagnostic radiology; ESAK; patient dose; Lorestan province; quality control; DOSE MEASUREMENTS; ENTRANCE SURFACE; EXPOSURES; BRAZIL; CHEST; UK;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: In diagnostic radiology there are two reasons for measuring or estimating radiation doses to patients. Firstly measurements provide a means for setting and checking standards of good practice as an aid to the optimization of patient protection. Secondly estimates of the absorbed dose to tissue and organs in the patients. Material and Methods: A total of 2382 patients were studied to calculate the Entrance Surface Air Kerma (ESAK) following seven radiographic examinations including: chest (PA, Lat), lumbar spines (AP, Lat), pelvis (AP), abdomen (AP), skull (PA, Lat), thoracic spine (AP, Lat) and cervical spines (AP, Lat). The ESAKs values were measured according to x- ray tube output, optimized exposure parameters and body thickness (tp) for each technique. Results: The parameters such as, 1st quartile, mean, median, 3rd quartile, minimum, maximum and standard deviation of each ESAK values are reported and compared to NRPB guide levels. The results showed that the ESAKs values in the lumbar spines and chest X-ray examinations were 30% above the guide levels. However, for the pelvis (AP), skull (PA) and abdomen (AP) examinations, these values were below than those reported by the NRPB. Conclusion: Periodic quality control and monitoring the technical performance of radiographers might effectively improve the image quality and eventually reducing the dose received by patients.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 50 条
  • [31] Diagnostic X-ray facilities as per quality control performances in Tanzania
    Sungita, Yesaya Y.
    Mdoe, Simon S. L.
    Msaki, Peter
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (04): : 66 - 73
  • [32] Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations
    Compagnone, Gaetano
    Pagan, Laura
    Bergamini, Carlo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2005, 6 (01): : 101 - 113
  • [33] A retrospective survey to establish institutional diagnostic reference levels for CT urography examinations based on clinical indications: preliminary results
    Efthymiou, Fotios O.
    Metaxas, Vasileios, I
    Dimitroukas, Christos P.
    Delis, Harry B.
    Zikou, Kiriaki D.
    Ntzanis, Epameinondas S.
    Zampakis, Petros E.
    Panayiotakis, George S.
    Kalogeropoulou, Christina P.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (06)
  • [34] Establishment of institutional diagnostic reference levels for 6 adult computed tomography examinations: Results from preliminary data collection
    Albahiti, Sarah K.
    Barnawi, Rashid A.
    Alsafi, Khalid
    Khafaji, Mawya
    Aljondi, Rowa
    Alghamdi, Salem Saeed
    Awan, Zuhier
    Sulieman, Abdelmoneim
    Jafer, Mustafa
    Tamam, Nissren
    Tajaldeen, Abdulrahman
    Mattar, Essam H.
    Al-Malki, Khaled Mudis
    Bradley, David
    RADIATION PHYSICS AND CHEMISTRY, 2022, 201
  • [35] Quality Control Monitoring of Medical Diagnostic X-Ray Units in Southern Tanzania
    Mammba H.P.
    Khamis F.O.
    Suleiman S.A.
    Edmund E.D.
    Salum S.K.
    Makoba A.
    Iranian Journal of Medical Physics, 2023, 20 (01) : 53 - 61
  • [36] QUANTITATIVE QUALITY CONTROL MEASUREMENTS OF A DIAGNOSTIC DIGITAL X-RAY IMAGE DETECTOR
    Ernits, Varmo
    Kepler, Kalle
    Vladimirov, Anatoli
    MEDICAL PHYSICS IN THE BALTIC STATES, 2017, : 45 - 48
  • [37] Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging system
    Sharma, Reena
    Sharma, Sunil Dutt
    Pawar, Shramika
    Chaubey, Ajay
    Kantharia, S.
    Babu, D. A. R.
    JOURNAL OF MEDICAL PHYSICS, 2015, 40 (01) : 29 - 37
  • [38] Assessment of patient radiation dose during skull X-ray examinations in King Khalid Hospital in Majmaah
    Saeed, Mohammed Khalil
    Abdallah, Yousif
    Sulieman, Abdelmoneim
    RADIATION PHYSICS AND CHEMISTRY, 2022, 201
  • [39] Physics audit of selected diagnostic X-Ray machines in Ghana: towards implementation of a national dosimetry audit programme
    Edem Kwabla Sosu
    Mary Boadu
    Francis Hasford
    Stephen Inkoom
    Shirazu Issahaku
    Theophilus Sackey
    Mark Pokoo–Aikins
    Theresa Dery
    Ernest Eduful
    Linda Osei-Poku
    Lawrence Akono Sarsah
    Kofi Okyere Akyea-Larbi
    Health and Technology, 2023, 13 : 791 - 798
  • [40] Physics audit of selected diagnostic X-Ray machines in Ghana: towards implementation of a national dosimetry audit programme
    Sosu, Edem Kwabla
    Boadu, Mary
    Hasford, Francis
    Inkoom, Stephen
    Issahaku, Shirazu
    Sackey, Theophilus
    Pokoo-Aikins, Mark
    Dery, Theresa
    Eduful, Ernest
    Osei-Poku, Linda
    Sarsah, Lawrence Akono
    Akyea-Larbi, Kofi Okyere
    HEALTH AND TECHNOLOGY, 2023, 13 (5) : 791 - 798