Lithium heparin interference in the Abbott enzymatic creatinine assay: the significance of under-filled tubes

被引:4
|
作者
Squires, Maria [1 ]
Wise, Helen [1 ]
Holmes, Heather [2 ]
Hadfield, Katie [3 ]
机构
[1] NHS Lothian, Edinburgh Royal Infirm, Dept Clin Biochem, 51 Little France Crescent,Old Dalkeith Rd, Edinburgh EH16 4SA, Midlothian, Scotland
[2] NHS Fife, Victoria Hosp, Dept Clin Biochem, Kirkcaldy, Scotland
[3] Airedale NHS Fdn Trust, Airedale Hosp, Dept Clin Biochem, Keighley, England
关键词
lithium heparin; interference; enzymatic creatinine; under-filled;
D O I
10.1177/00045632211040673
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Spuriously high results using the Abbott Architect enzymatic creatinine assay were noted to be particularly associated with very small sample volumes. This led us to query the effect of under-filling lithium heparin tubes on the measured enzymatic creatinine result. Methods Blood was provided by 5 laboratory personnel and then decanted into 5 x1.2 mL Sarstedt S-Monovette tubes, giving final blood volumes of 200, 400, 600, 800 and 1200 mu L. Plasma was analysed using Abbott Architect Jaffe, enzymatic creatinine, Beckman Coulter (AU500) enzymatic creatinine and Roche (Cobas c702) enzymatic creatinine assays. Saline was also added to Sarstedt 1.2 mL and Teklab 2 mL tubes and analysed using the Abbott Jaffe and enzymatic creatinine methods. Results Increasing degrees of under-fill were associated with greater over-estimation of creatinine using the Abbott enzymatic assay, but no difference was noted using Jaffe methodology on the same platform or enzymatic assays provided by Roche or Beckman. On average, creatinine was 40.6% (+27.7 mu mol/L) higher when only 200 mu L of blood was present in the tube. Small volumes of saline added to lithium heparin tubes measured significant creatinine concentrations using the Abbott enzymatic method. Conclusions Lithium heparin directly interferes in the Abbott Architect enzymatic creatinine assay. Under-filling lithium heparin tubes can lead to clinically significant over-estimation of creatinine results by this assay. Users of this assay should be aware of the potential for spurious results in small sample volumes collected into lithium heparin tubes and implement robust procedures for identifying and reporting results on these samples.
引用
收藏
页码:653 / 656
页数:4
相关论文
共 16 条
  • [1] Interference on the Abbott i-STAT creatinine assay caused by hydroxyurea
    Paz, Matt Andrew
    Seegmiller, Jesse
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2022, 158 (SUPP 1) : S19 - S20
  • [2] Phenindione interference in enzymatic creatinine assay - A case report
    Natarajan, Balan
    Hart, Tanya
    Smith, Sophy
    Parry, Robin Geoffrey
    CLINICAL NEPHROLOGY, 2015, 83 (02) : 121 - 124
  • [3] Interference of dopamine in enzymatic creatinine assay: A Sri Lankan experience
    Gamage, E.
    Siriwardene, S.
    Perera, W.
    CLINICA CHIMICA ACTA, 2019, 493 : S217 - S217
  • [4] New enzymatic creatinine assay showing stable endpoint and no interference by creatine and lipids
    Thoenges, D
    Mueller, K
    Mueller, H
    Metzmann, E
    CLINICA CHIMICA ACTA, 2005, 355 : S209 - S209
  • [5] Pseudohypercreatininaemia in two patients caused by monoclonal IgM interference with enzymatic assay of creatinine
    Salter, Tracey
    Marsh, James
    Sood, Bhrigu
    Livingstone, Callum
    Gallagher, Hugh
    JOURNAL OF CLINICAL PATHOLOGY, 2015, 68 (10) : 854 - U105
  • [7] Interference in complement C4 immunoturbidimetric assay related to heparin tubes
    Zhu, Y.
    Brass, D.
    Jenkins, M.
    Horne, B.
    Drayton, N.
    Dean, S.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A223 - A223
  • [8] The calibrator value assignment protocol of the Abbott enzymatic creatinine assay is inadequate for ensuring suitable quality of serum measurements
    Pasqualetti, Sara
    Infusino, Ilenia
    Carnevale, Assunta
    Szoke, Dominika
    Panteghini, Mauro
    CLINICA CHIMICA ACTA, 2015, 450 : 125 - 126
  • [9] Evaluation of MULTIGENT® Creatinine (Enzymatic) liquid assay on the ABBOTT ARCHITECT® c8000 and AEROSET® Clinical Chemistry Systems
    Rota, F.
    Magro, F.
    Nance, B.
    Markle, H. V.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A204 - A204
  • [10] Evaluation of the interference by homogentisic acid and other organic acids on the enzymatic and Jaffe method creatinine assay
    Pauwels, Steven
    Cassiman, David
    Vermeersch, Pieter
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (04) : 749 - 750