Evaluation of three indirect calorimetry devices in mechanically ventilated patients: Which device compares best with the Deltatrac II®? A prospective observational study

被引:71
作者
Graf, Severine [1 ,2 ]
Karsegard, Veronique Laurie [1 ]
Viatte, Valerie [1 ]
Heidegger, Claudia Paula [2 ]
Fleury, Yvan [2 ]
Pichard, Claude [1 ]
Genton, Laurence [1 ]
机构
[1] Univ Hosp, Clin Nutr Unit, CH-1211 Geneva, Switzerland
[2] Univ Hosp, Intens Care Unit, CH-1211 Geneva, Switzerland
关键词
Indirect calorimetry; Intensive care unit; Energy expenditure; Mechanical ventilation; CRITICALLY-ILL PATIENTS; INTENSIVE-CARE-UNIT; RESTING METABOLIC-RATE; ENERGY-EXPENDITURE; PREDICTION EQUATIONS; PARENTERAL-NUTRITION; ENTERAL NUTRITION; INSTRUMENTS; GUIDELINES; PROVISION;
D O I
10.1016/j.clnu.2014.01.008
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Background & aims: Indirect calorimetry (IC) is the gold standard to measure energy expenditure (EE) in hospitalized patients. The popular 30 year-old Deltatrac II(R) (Datex) IC is no more commercialized, but other manufacturers have developed new devices. This study aims at comparing for the first time simultaneously, two new IC, the CCM express(R) (Medgraphics) and the Quark RMR(R) (Cosmed) with the Deltatrac II(R) to assess their potential use in intensive care unit (ICU) patients. Methods: ICU patients on mechanical ventilation, with positive end-expiratory pressure <9 cm H2O and fraction of inspired oxygen <60%, underwent measurements by the three IC simultaneously connected during 20 min to the ventilator (Evita XL, Drager). Patients' characteristics, VO2 consumption, VCO2 production, respiratory quotient and EE were recorded. Data were presented as mean (SD) and compared by linear regression, repeated measure one-way ANOVA and Bland & Altman diagrams. Results: Forty patients (23 males, 60(17) yrs, BMI 25.4(7.0) kg/m(2)) were included. For the Deltatrac II(R), VO2 was 227(61) ml/min, VCO2 189(52) ml/min and EE 1562(412) kcal/d. VO2, VCO2, and EE differed significantly between Deltatrac II(R) and CCM express(R) (p < 0.001), but not between Deltatrac II(R) and Quark RMR(R). For EE, diagrams showed a mean difference (2SD) of 25.2(441) kcal between Deltatrac II(R) vs. the Quark RMR(R), and -273 (532) kcal between Deltatrac II(R) vs CCM express(R). Conclusion: Quark RMR(R) compares better with Deltatrac II(R) than CCM express(R), but it suffers an EE variance of 441 kcal, which is not acceptable for clinical practice. New indirect IC should be further improved before recommending their clinical use in ICU. (C) 2014 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. All rights reserved.
引用
收藏
页码:60 / 65
页数:6
相关论文
共 31 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Assessing Validity and Reliability of Resting Metabolic Rate in Six Gas Analysis Systems [J].
Cooper, Jamie A. ;
Watras, Abigail C. ;
O'Brien, Matthew J. ;
Luke, Amy ;
Dobratz, Jennifer R. ;
Earthman, Carrie P. ;
Schoeller, Dale A. .
JOURNAL OF THE AMERICAN DIETETIC ASSOCIATION, 2009, 109 (01) :128-132
[3]   Combination enteral and parenteral nutrition in critically ill patients: harmful or beneficial? A systematic review of the evidence [J].
Dhaliwal, R ;
Jurewitsch, B ;
Harrietha, D ;
Heyland, DK .
INTENSIVE CARE MEDICINE, 2004, 30 (08) :1666-1671
[4]   Computerized energy balance and complications in critically ill patients: An observational study [J].
Dvir, D ;
Cohen, J ;
Singer, P .
CLINICAL NUTRITION, 2006, 25 (01) :37-44
[5]   Assessment of resting energy expenditure in mechanically ventilated patients [J].
Faisy, C ;
Guerot, E ;
Diehl, JL ;
Labrousse, J ;
Fagon, JY .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 2003, 78 (02) :241-249
[6]  
Flancbaum L, 1999, AM J CLIN NUTR, V69, P461
[7]   Adaptive alterations in metabolism: practical consequences on energy requirements in the severely ill patient [J].
Fontaine, Eric ;
Mueller, Manfred J. .
CURRENT OPINION IN CLINICAL NUTRITION AND METABOLIC CARE, 2011, 14 (02) :171-175
[8]   Energy Estimation and Measurement in Critically Ill Patients [J].
Fraipont, Vincent ;
Preiser, Jean-Charles .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2013, 37 (06) :705-713
[9]   On heat, respiration, and calorimetry [J].
Frankenfield, David C. .
NUTRITION, 2010, 26 (10) :939-950
[10]   Analysis of Estimation Methods for Resting Metabolic Rate in Critically Ill Adults [J].
Frankenfield, David C. ;
Coleman, Abigail ;
Alam, Shoaib ;
Cooney, Robert N. .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2009, 33 (01) :27-36