A high sensitive fission neutron detector system with a lead slot collimator

被引:5
作者
Ouyang, XP [1 ]
Li, ZF
Wang, QS
Huo, YK
Ma, YL
Zhang, QM
Zhang, GG
Jin, YR
机构
[1] Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China
[2] NW Inst Nucl Technol, Xian 710024, Peoples R China
[3] Fudan Univ, Inst Modern Phys, Shanghai 200433, Peoples R China
关键词
neutron detection; high sensitivity; high eta/gamma discrimination; fission; Be backing;
D O I
10.7498/aps.54.4643
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
A fission neutron detector system with frontward- extended lead slot collimator has been developed and tested. By designing the slot collimator and developing large area PIN detectors as well as large area fissionable target with thin Be backing, this detector system can reach high neutron sensitivities up to 10-(16) C(.)cm(2) which is about 4-5 orders of magnitude higher than that of existing fission detectors, and high n/gamma discrimination ratio up to 30.
引用
收藏
页码:4643 / 4647
页数:5
相关论文
共 5 条
[1]  
FUDAN U, 1985, EXPT METHODS NUCL PH, V1
[2]  
JING YR, 2001, 2986 NINT B
[3]  
LIU QZ, 1994, DIAGNOSTIC TECHNOLOG
[4]   Development of large area silicon semiconductor detectors for use in the current mode [J].
Ouyang, XP ;
Li, ZF ;
Zhang, GG ;
Huo, YK ;
Zhang, QM ;
Zhang, XP ;
Song, XC ;
Jia, HY ;
Lei, JH ;
Sun, YC .
ACTA PHYSICA SINICA, 2002, 51 (07) :1502-1505
[5]  
OUYANG XP, 2002, STUDIES DETECTION PU