Fracture of a ceramic liner in a total hip arthroplasty with a sandwich cup

被引:17
|
作者
Popescu, Dragos [1 ]
Gallart, X. [1 ]
Garcia, S. [1 ]
Bori, G. [1 ]
Tomas, X. [2 ]
Riba, J. [1 ]
机构
[1] Univ Barcelona, Dept Orthoped & Trauma Surg, Hip Unit, Hosp Clin, E-08036 Barcelona, Spain
[2] Univ Barcelona, Dept Radiol, Hosp Clin, E-08036 Barcelona, Spain
关键词
hip arthroplasty; liner fracture; complication;
D O I
10.1007/s00402-007-0536-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Alumina ceramic-on-ceramic total hip arthroplasty (THA) has been widely used due to its advantages such as low wear, scratch resistance, wettable surface and relatively low biological reactivity of the wear particles. Nevertheless, this material in THA still persists to be one of the major concerns about the risk of fracture, due to its brittleness. Many authors have reported a fracture of the ceramic head but few reported a fracture of the ceramic acetabular insert. In order to reduce the rigidity of the ceramic-on-ceramic coupling and prevent an impingement between the rim of the ceramic liner and the metal neck of the femoral stem, a modular acetabular component with a sandwich insertion (alumina/polyethylene/titanium) was proposed. We report the fracture of the ceramic acetabular liner of such a ceramic sandwich cup due to a slightly retroverted position of the cup that causes an impingement between the femoral stem and the rim of the insert. The fracture occurred 3 years after the operation without trauma. At revision the entire cup was replaced using a polyethylene liner without inner ceramic liner.
引用
收藏
页码:783 / 785
页数:3
相关论文
共 50 条
  • [31] Squeaking sound in total hip arthroplasty using ceramic-on-ceramic bearing surfaces
    Ki, Sung-Chan
    Kim, Byung-Hak
    Ryu, Ji-Hoon
    Yoon, Dae-Hyun
    Chung, Young-Yool
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2011, 16 (01) : 21 - 25
  • [32] Midterm Results of Fourth-Generation Ceramic-On-Ceramic Total Hip Arthroplasty
    Cho, Kyu-Jin
    Park, Kyung-Soon
    Jong, Won-Young
    Park, Chun-Gon
    Yoon, Taek-Rim
    INDIAN JOURNAL OF ORTHOPAEDICS, 2019, 53 (05) : 630 - 636
  • [33] Interposition of the Fracture Fragment of a Vitamin E-Blended, Highly Crosslinked Polyethylene Liner After Total Hip Arthroplasty: A Case Report
    Tanino, Hiromasa
    Mitsutake, Ryo
    Ito, Hiroshi
    CLINICAL CASE REPORTS, 2024, 12 (12):
  • [34] Ceramic-on-ceramic total hip arthroplasty: Is squeaking related to an inaccurate three-dimensional hip anatomy reconstruction?
    Sariali, E.
    Klouche, S.
    Mamoudy, P.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (04) : 437 - 440
  • [35] Incidence, Risk Factors, and Outcome of Ceramic-On-Ceramic Bearing Breakage in Total Hip Arthroplasty
    Hoskins, Wayne
    Rainbird, Sophia
    Peng, Yi
    Lorimer, Michelle
    Graves, Stephen E.
    Bingham, Roger
    JOURNAL OF ARTHROPLASTY, 2021, 36 (08): : 2992 - 2997
  • [36] FIRST CASE REPORT OF VANADIUM METALLOSIS AFTER CERAMIC-ON-CERAMIC TOTAL HIP ARTHROPLASTY
    Pesce, V.
    Maccagnano, G.
    Vicenti, G.
    Notarnicola, A.
    Lovreglio, P.
    Soleo, L.
    Pantalone, A.
    Salini, V.
    Moretti, B.
    JOURNAL OF BIOLOGICAL REGULATORS AND HOMEOSTATIC AGENTS, 2013, 27 (04): : 1063 - 1068
  • [37] Total hip and knee arthroplasty in nonagenarians
    Alfonso, Daniel T.
    Howell, Damani
    Strauss, Eric J.
    Di Cesare, Paul E.
    JOURNAL OF ARTHROPLASTY, 2007, 22 (06): : 807 - 811
  • [38] Total hip arthroplasty after failed treatment of proximal femur fracture
    Tomonori Tetsunaga
    Kazuo Fujiwara
    Hirosuke Endo
    Tomoyuki Noda
    Tomoko Tetsunaga
    Toru Sato
    Naofumi Shiota
    Toshifumi Ozaki
    Archives of Orthopaedic and Trauma Surgery, 2017, 137 : 417 - 424
  • [39] Total hip arthroplasty after failed treatment of proximal femur fracture
    Tetsunaga, Tomonori
    Fujiwara, Kazuo
    Endo, Hirosuke
    Noda, Tomoyuki
    Tetsunaga, Tomoko
    Sato, Toru
    Shiota, Naofumi
    Ozaki, Toshifumi
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2017, 137 (03) : 417 - 424
  • [40] Outcomes of dual-mobility acetabular cup for instability in primary and revision total hip arthroplasty
    Mohammed R.
    Hayward K.
    Mulay S.
    Bindi F.
    Wallace M.
    Journal of Orthopaedics and Traumatology, 2015, 16 (1) : 9 - 13