Fracture of a ceramic liner in a total hip arthroplasty with a sandwich cup

被引:17
|
作者
Popescu, Dragos [1 ]
Gallart, X. [1 ]
Garcia, S. [1 ]
Bori, G. [1 ]
Tomas, X. [2 ]
Riba, J. [1 ]
机构
[1] Univ Barcelona, Dept Orthoped & Trauma Surg, Hip Unit, Hosp Clin, E-08036 Barcelona, Spain
[2] Univ Barcelona, Dept Radiol, Hosp Clin, E-08036 Barcelona, Spain
关键词
hip arthroplasty; liner fracture; complication;
D O I
10.1007/s00402-007-0536-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Alumina ceramic-on-ceramic total hip arthroplasty (THA) has been widely used due to its advantages such as low wear, scratch resistance, wettable surface and relatively low biological reactivity of the wear particles. Nevertheless, this material in THA still persists to be one of the major concerns about the risk of fracture, due to its brittleness. Many authors have reported a fracture of the ceramic head but few reported a fracture of the ceramic acetabular insert. In order to reduce the rigidity of the ceramic-on-ceramic coupling and prevent an impingement between the rim of the ceramic liner and the metal neck of the femoral stem, a modular acetabular component with a sandwich insertion (alumina/polyethylene/titanium) was proposed. We report the fracture of the ceramic acetabular liner of such a ceramic sandwich cup due to a slightly retroverted position of the cup that causes an impingement between the femoral stem and the rim of the insert. The fracture occurred 3 years after the operation without trauma. At revision the entire cup was replaced using a polyethylene liner without inner ceramic liner.
引用
收藏
页码:783 / 785
页数:3
相关论文
共 50 条
  • [1] Fracture of a ceramic liner in a total hip arthroplasty with a sandwich cup
    Dragos Popescu
    X. Gallart
    S. Garcia
    G. Bori
    X. Tomas
    J. Riba
    Archives of Orthopaedic and Trauma Surgery, 2008, 128 : 783 - 785
  • [2] Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup
    Hasegawa, M
    Sudo, A
    Hirata, H
    Uchida, A
    JOURNAL OF ARTHROPLASTY, 2003, 18 (05): : 658 - 661
  • [3] Risk factors for ceramic liner fracture after total hip arthroplasty
    Traina, Francesco
    De Fine, Marcello
    Bordini, Barbara
    Toni, Aldo
    HIP INTERNATIONAL, 2012, 22 (06) : 607 - 614
  • [4] Ceramic Head Fracture in Ceramic-on-Polyethylene Total Hip Arthroplasty
    Nho, Jae-Hwi
    Park, Jong-Seok
    Song, Ui-Seoub
    Kim, Woo-Jong
    Suh, You-Sung
    YONSEI MEDICAL JOURNAL, 2013, 54 (06) : 1550 - 1553
  • [5] Fracture of the femoral head after ceramic-on-polyethylene total hip arthroplasty
    Callaway, GH
    Flynn, W
    Ranawat, CS
    Sculco, TP
    JOURNAL OF ARTHROPLASTY, 1995, 10 (06): : 855 - 859
  • [6] Fracture of a ceramic femoral head after ceramic-on-ceramic total hip arthroplasty
    Malillos Toran, Manuel
    Cuenca, Jorge
    Antonio Martinez, Angel
    Herrera, Antonio
    Vicente Thomas, Javier
    JOURNAL OF ARTHROPLASTY, 2006, 21 (07): : 1072 - 1073
  • [7] Long-term follow-up of total hip arthroplasty using polyethylene-ceramic composite (sandwich) liner
    Rodriguez-Perez, Daniel
    Carnaval, Thiago
    Rodriguez, Marcos-del-Carmen
    Coscujuela-Mana, Antonio
    Agullo, Jose-Luis
    Videla, Sebastian
    HIP INTERNATIONAL, 2024, 34 (04) : 467 - 475
  • [8] Spontaneous fracture of the ceramic ball in a ceramic-polyethylene total hip arthroplasty
    Michaud, RJ
    Rashad, SY
    JOURNAL OF ARTHROPLASTY, 1995, 10 (06): : 863 - 867
  • [9] Cup positioning in total hip arthroplasty
    Scheerlinck, Thierry
    ACTA ORTHOPAEDICA BELGICA, 2014, 80 (03): : 336 - 347
  • [10] Ceramic on ceramic bearings in total hip arthroplasty
    Buchanan, JM
    BIOCERAMICS 15, 2003, 240-2 : 793 - 796