Fully automated 96-well liquid-liquid extraction for analysis of biological samples by liquid chromatography with tandem mass spectrometry

被引:79
|
作者
Peng, SX [1 ]
Branch, TM [1 ]
King, SL [1 ]
机构
[1] Procter & Gamble Pharmaceut, Hlth Care Res Ctr, Mason, OH 45040 USA
关键词
D O I
10.1021/ac001036c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fully automated high-throughput liquid-liquid extraction (LLE) methodology has been developed for preparation of biological samples using a 96-well LLE plate and a 96-channel robotic liquid handling workstation. The 96-well LLE plate is made of a 96-well filter plate filled with inert diatomaceous earth particles, allowing continuous and efficient extraction of analytes between the aqueous biological sample and the organic extraction solvent. Two carboxylic acid-based protease inhibitor compounds with high and low levels of plasma protein binding were chosen for the development and application of the automated methodology. The LLE extracts of the plasma samples of the two compounds were analyzed by high-performance liquid chromatography with electrospray (ESI) tandem mass spectrometry (LC-MS/MS), The LC-MS/MS method was developed using a rapid gradient LC separation, followed by sample introduction through an ionspray interface in the negative ion mode and tandem mass spectrometric detection with selected reaction monitoring. In the optimized LLE method, a formate buffer solution was first loaded into a 96-well filter plate packed with inert diatomaceous earth material. Then crude plasma samples and a water-immiscible organic solvent, methyl ethyl ketone, were sequentially added to the LLE plate so that LLE would occur in the interface between the two liquid phases on the surface of individual particles in each well. The organic eluate containing extracted analytes was evaporated and reconstituted for LC-MS/MS analysis. This fully automated LLE methodology avoids several disjointed steps involved in a manual or semiautomated LLE method, leading to significantly reduced sample preparation time, increased sample throughput, and clean sample extracts for improved ESI-MS/MS detection. The automated LLE methodology is universal and can be employed for sample preparation of other biological fluids, The complete bioanalytical method, based on the automated LLE and fast gradient LC-IMS/MS, was validated and successfully applied to the quantitative analysis of protease inhibitors in rat plasma.
引用
收藏
页码:708 / 714
页数:7
相关论文
共 50 条
  • [31] Direct injection versus liquid-liquid extraction for plasma sample analysis by high performance liquid chromatography with tandem mass spectrometry
    Jemal, M
    Huang, M
    Jiang, XH
    Mao, Y
    Powell, ML
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1999, 13 (21) : 2125 - 2132
  • [32] Determination of a metabolite of nifursol in foodstuffs of animal origin by liquid-liquid extraction and liquid chromatography with tandem mass spectrometry
    Wang, Chuanxian
    Qu, Li
    Liu, Xia
    Zhao, Chaomin
    Zhao, Fengjuan
    Huang, Fuzhen
    Zhu, Zhenou
    Han, Chao
    JOURNAL OF SEPARATION SCIENCE, 2017, 40 (03) : 671 - 676
  • [33] Determination of (R)- and (S)-ketoprofen in human plasma by liquid chromatography/tandem mass spectrometry following automated solid-phase extraction in the 96-well format
    Eichhold, TH
    Bailey, RE
    Tanguay, SL
    Hoke, SH
    JOURNAL OF MASS SPECTROMETRY, 2000, 35 (04): : 504 - 511
  • [34] Validation and application of a high-performance liquid chromatography-tandem mass spectrometric method for simultaneous quantification of lopinavir and ritonavir in human plasma using semi-automated 96-well liquid-liquid extraction
    Wang, Perry G.
    Wei, Jack S.
    Kim, Grace
    Chang, Min
    EI-Shourbagy, Tawakol
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1130 (02) : 302 - 307
  • [35] Simultaneous quantitation of enalapril and enalaprilat in human plasma by 96-well solid-phase extraction and liquid chromatography/tandem mass spectrometry
    Lee, J
    Son, J
    Lee, M
    Lee, KT
    Kim, DH
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (11) : 1157 - 1162
  • [36] An improved high-throughput liquid chromatographic/tandem mass spectrometric method for terbinafine quantification in human plasma, using automated liquid-liquid extraction based on 96-well format plates
    Dotsikas, Yannis
    Apostolou, Constantinos
    Kousoulos, Constantinos
    Tsatsou, Georgia
    Loukas, Yannis L.
    BIOMEDICAL CHROMATOGRAPHY, 2007, 21 (02) : 201 - 208
  • [37] Quantitative determination of paclitaxel in human plasma using semi-automated liquid-liquid extraction in conjunction with liquid chromatography/tandem mass spectrometry
    Basileo, G
    Breda, M
    Fonte, G
    Pisano, R
    James, CA
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2003, 32 (4-5) : 591 - 600
  • [38] Determination of acrinathrin in water samples by micro liquid-liquid extraction and gas chromatography mass spectrometry
    Vilchez, JL
    Espinosa, P
    Arrebola, FJ
    GonzalezCasado, A
    ANALYTICAL SCIENCES, 1997, 13 (05) : 817 - 819
  • [39] Isolation, preconcentration and determination of rhamnolipids in aqueous samples by dispersive liquid-liquid microextraction and liquid chromatography with tandem mass spectrometry
    Zgola-Grzeskowiak, Agnieszka
    Kaczorek, Ewa
    TALANTA, 2011, 83 (03) : 744 - 750
  • [40] A Validated Liquid Chromatography-Tandem Mass Spectrometry Coupled with Liquid-Liquid Extraction for Indacaterol Quantitation in Human Plasma
    Ammari, Wesam G.
    Al-Qadhi, Zainab
    Khalil, Mohammad
    Tayyem, Rabab
    Qammaz, Samir
    Oriquat, Ghaleb
    Basheti, Iman A.
    Chrystyn, Henry