An investigation into the variability between different shear wave elastography systems in muscle

被引:64
作者
Alfuraih, Abdulrahman M. [1 ,2 ,3 ]
O'Connor, Phil [3 ]
Tan, Ai Lyn [2 ,3 ]
Hensor, Elizabeth M. A. [2 ,3 ]
Emery, Paul [2 ,3 ]
Wakefield, Richard J. [2 ,3 ]
机构
[1] Prince Sattam bin Abdulaziz Univ, Coll Appl Med Sci, Radiol & Med Imaging Dept, Kharj, Saudi Arabia
[2] Univ Leeds, Chapel Allerton Hosp, Leeds Inst Rheumat & Musculoskeletal Med, Leeds, W Yorkshire, England
[3] Leeds Teaching Hosp NHS Trust, NIHR Leeds Biomed Res Ctr, Leeds, W Yorkshire, England
基金
美国国家卫生研究院;
关键词
elastography; ultrasound; muscles; reliability; SEQUENTIAL TRACKING TAST; SKELETAL-MUSCLE; ELASTIC PROPERTIES; WFUMB GUIDELINES; CLINICAL-USE; CONTRACTION; TISSUE; RECOMMENDATIONS; RELIABILITY; AGREEMENT;
D O I
10.11152/mu-1113
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Aims: The reliability and agreement between shear wave elastography (SWE) systems using different acquisition methods in muscles is not yet established. The objectives were to determine the reliability of a new SWE system on normal resting muscles using different acquisition methods and to compare its performance to an established state-of-the-art system. Material and methods: Small, medium and large ROI sizes in addition to longitudinal, oblique and transverse orientations over five different locations within the rectus femoris muscle were tested using the new system. Results were compared to the established system to test for inter-system reproducibility. Results: Lowest within-subject coefficient of variance (4.3%) and shear wave velocity (1.83 m/s) were associated with the medium ROI and longitudinal orientation from the lateral location. This combination resulted in a strong internal agreement of intra-class correlation of 0.76 (0.57-0.89) for the new system and an almost perfect agreement of 0.92 (0.82-0.97) for the established. Inter-system reproducibility for the best combination was 0.71 (0.48-1) with a mean velocity difference +/- 95% limits of agreement of 0.07 +/- 0.49 m/s. Conclusions: Altering SWE acquisition methods can produce variable results. The new system produced reliable results that are comparable with but not as reliable as the established.
引用
收藏
页码:392 / 400
页数:9
相关论文
共 32 条
[1]   Acute Effect of Static Stretching on Hardness of the Gastrocnemius Muscle [J].
Akagi, Ryota ;
Takahashi, Hideyuki .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2013, 45 (07) :1348-1354
[2]   Muscle shear elastic modulus is linearly related to muscle torque over the entire range of isometric contraction intensity [J].
Ates, Filiz ;
Hug, Francois ;
Bouillard, Killian ;
Jubeau, Marc ;
Frappart, Thomas ;
Couade, Mathieu ;
Bercoff, Jeremy ;
Nordez, Antoine .
JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2015, 25 (04) :703-708
[3]   WFUMB GUIDELINES AND RECOMMENDATIONS FOR CLINICAL USE OF ULTRASOUND ELASTOGRAPHY: PART 2: BREAST [J].
Barr, Richard G. ;
Nakashima, Kazutaka ;
Amy, Dominique ;
Cosgrove, David ;
Farrokh, Andre ;
Schafer, Fritz ;
Bamber, Jeffrey C. ;
Castera, Laurent ;
Choi, Byung Ihn ;
Chou, Yi-Hong ;
Dietrich, Christoph F. ;
Ding, Hong ;
Ferraioli, Giovanna ;
Filice, Carlo ;
Friedrich-Rust, Mireen ;
Hall, Timothy J. ;
Nightingale, Kathryn R. ;
Palmeri, Mark L. ;
Shiina, Tsuyoshi ;
Suzuki, Shinichi ;
Sporea, Ioan ;
Wilson, Stephanie ;
Kudo, Masatoshi .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (05) :1148-1160
[4]   Supersonic shear imaging: A new technique for soft tissue elasticity mapping [J].
Bercoff, J ;
Tanter, M ;
Fink, M .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (04) :396-409
[5]  
Bland JM, 1996, BRIT MED J, V313, P744
[6]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[7]   Statistics notes: Measurement error proportional to the mean .23. [J].
Bland, JM ;
Altman, DG .
BRITISH MEDICAL JOURNAL, 1996, 313 (7049) :106-106
[8]   Skeletal Muscle in Healthy Subjects versus Those with GNE-Related Myopathy: Evaluation with Shear-Wave US-A Pilot Study [J].
Carpenter, Elizabeth L. ;
Lau, Heather A. ;
Kolodny, Edwin H. ;
Adler, Ronald S. .
RADIOLOGY, 2015, 277 (02) :546-554
[9]   Reliability, Validity, and Responsiveness of Myotonoinetric Measurement of Muscle Tone, Elasticity, and Stiffness in Patients With Stroke [J].
Chuang, Li-ling ;
Wu, Ching-yi ;
Lin, Keh-chung .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2012, 93 (03) :532-540
[10]   Muscle and joint elastic properties during elbow flexion in Duchenne muscular dystrophy [J].
Cornu, C ;
Goubel, F ;
Fardeau, M .
JOURNAL OF PHYSIOLOGY-LONDON, 2001, 533 (02) :605-616