Pulse Oximeter Performance during Rapid Desaturation

被引:2
|
作者
Horakova, Lenka [1 ]
Roubik, Karel [1 ]
机构
[1] Czech Tech Univ, Fac Biomed Engn, Dept Biomed Technol, Kladno 27201, Czech Republic
关键词
pulse oximetry; hypoxemia; oxygen saturation; safety limits; avalanche victim; outdoor experiments; CLINICAL-EVALUATION; OXYGEN-SATURATION; SNOW BURIAL; AIR POCKET; ACCURACY; HYPOXIA; RESPIRATION; HYPERCAPNIA; INTUBATION; HYPOXEMIA;
D O I
10.3390/s22114236
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The reliability of pulse oximetry is crucial, especially in cases of rapid changes in body oxygenation. In order to evaluate the performance of pulse oximeters during rapidly developing short periods of concurrent hypoxemia and hypercapnia, 13 healthy volunteers underwent 3 breathing phases during outdoor experiments (39 phases in total), monitored simultaneously by five different pulse oximeters. A significant incongruity in values displayed by the tested pulse oximeters was observed, even when the accuracy declared by the manufacturers were considered. In 28.2% of breathing phases, the five used devices did not show any congruent values. The longest uninterrupted congruent period formed 74.4% of total recorded time. Moreover, the congruent periods were rarely observed during the critical desaturation phase of the experiment. The time difference between the moments when the first and the last pulse oximeter showed the typical study endpoint values of SpO(2) 85% and 75% was 32.1 +/- 23.6 s and 24.7 +/- 19.3 s, respectively. These results suggest that SpO(2) might not be a reliable parameter as a study endpoint, or more importantly as a safety limit in outdoor experiments. In the design of future studies, more parameters and continuous clinical assessment should be included.
引用
收藏
页数:13
相关论文
共 50 条
  • [41] MONITORING THE FETUS WITH A PULSE OXIMETER DURING A CESAREAN-SECTION
    JOHNSON, N
    JOHNSON, VA
    BANNISTER, J
    LYONS, G
    LILFORD, RJ
    GRIFFITHSJONES, M
    TUFFNELL, D
    ONWUDE, JL
    BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1990, 97 (07): : 653 - 654
  • [42] Pulse Oximetry Recorded from the Phone Oximeter for Detection of Obstructive Sleep Apnea Events With and Without Oxygen Desaturation in Children
    Garde, Ainara
    Dehkordi, Parastoo
    Wensley, David
    Ansermino, J. Mark
    Dumont, Guy A.
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 7692 - 7695
  • [43] AN ERRONEOUS PULSE OXIMETER READING
    HOPKINS, PM
    ANAESTHESIA, 1989, 44 (10) : 868 - 868
  • [44] THE PULSE OXIMETER AND ALLEN TEST
    PERSSON, E
    ANAESTHESIA, 1992, 47 (05) : 451 - 451
  • [45] Pulse Oximeter for Cyclists in Smartphone
    Martins, L.
    Gaidos, O.
    dos Santos, I.
    10TH INTERNATIONAL SYMPOSIUM ON MEDICAL INFORMATION PROCESSING AND ANALYSIS, 2015, 9287
  • [46] Pulse oximeter in a comparative test
    Hartmut Gehring
    Intensive Care Medicine, 2006, 32 : 1287 - 1289
  • [47] A NEW USE FOR THE PULSE OXIMETER
    INTRONA, RPS
    SILVERSTEIN, PI
    ANESTHESIOLOGY, 1986, 65 (03) : 342 - 342
  • [48] Pulse oximeter in a comparative test
    Gehring, Hartmut
    INTENSIVE CARE MEDICINE, 2006, 32 (09) : 1287 - 1289
  • [49] ANOTHER USE OF THE PULSE OXIMETER
    PILLAI, EV
    THAS, SK
    ANAESTHESIA, 1989, 44 (08) : 705 - 706
  • [50] PULSE OXIMETER AND ALLEN TEST
    GLAVIN, RJ
    ANAESTHESIA, 1992, 47 (10) : 917 - 917