A novel anthropomorphic multimodality phantom for MRI-based radiotherapy quality assurance testing

被引:21
|
作者
Singhrao, Kamal [1 ]
Fu, Jie [1 ]
Wu, Holden H. [2 ]
Hu, Peng [2 ]
Kishan, Amar U. [1 ]
Chin, Robert K. [1 ]
Lewis, John H. [3 ]
机构
[1] Univ Calif Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Radiol, Los Angeles, CA 90095 USA
[3] Cedars Sinai Med Ctr, Dept Radiat Oncol, Los Angeles, CA 90048 USA
关键词
anthropomorphic; MRI; multimodality; phantom; QA; radiotherapy; RADIATION-THERAPY; RELAXATION-TIMES; PROSTATE RADIOTHERAPY; IN-VIVO; CT; TISSUE; SYSTEM; SCANS; 1.5T;
D O I
10.1002/mp.14027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Increased utilization of magnetic resonance imaging (MRI) in radiotherapy has caused a growing need for phantoms that provide tissue-like contrast in both computed tomography (CT) and MRI images. Such phantoms can be used to compare MRI-based processes with CT-based clinical standards. Here, we develop and demonstrate the clinical utility of a three-dimensional (3D)-printed anthropomorphic pelvis phantom containing materials capable of T-1, T-2, and electron density matching for a clinically relevant set of soft tissues and bone. Methods The phantom design was based on a male pelvic anatomy template with thin boundaries separating tissue types. Slots were included to allow insertion of various dosimeters. The phantom structure was created using a 3D printer. The tissue compartments were filled with carrageenan-based materials designed to match the T-1 and T-2 relaxation times and electron densities of the corresponding tissues. CT and MRI images of the phantom were acquired and used to compare phantom T-1 and T-2 relaxation times and electron densities to literature-reported values for human tissue. To demonstrate clinical utility, the phantom was used for end-to-end testing of an MRI-only treatment simulation and planning workflow. Based on a T-2-weighted MRI image, synthetic CT (sCT) images were created using a statistical decomposition algorithm (MRIPlanner, Spectronic Research AB, Sweden) and used for dose calculation of volumetric-modulated arc therapy (VMAT) and seven-field intensity-modulated radiation therapy (IMRT) prostate plans. The plans were delivered on a Truebeam STX (Varian Medical Systems, Palo Alto, CA), with film and a 0.3 cc ion chamber used to measure the delivered dose. Doses calculated on the CT and sCTs were compared using common dose volume histogram metrics. Results T-1 and T-2 relaxation time and electron density measurements for the muscle, prostate, and bone agreed well with literature-reported in vivo measurements. Film analysis resulted in a 99.7% gamma pass rate (3.0%, 3.0 mm) for both plans. The ion chamber-measured dose discrepancies at the isocenter were 0.36% and 1.67% for the IMRT and VMAT plans, respectively. The differences in PTV D98% and D95% between plans calculated on the CT and 1.5T/3.0 T-derived sCT images were under 3%. Conclusion The developed phantom provides tissue-like contrast on MRI and CT and can be used to validate MRI-based processes through comparison with standard CT-based processes.
引用
收藏
页码:1443 / 1451
页数:9
相关论文
共 50 条
  • [21] Simple, inexpensive, and ergonomic phantom for quality assurance control of MRI guided Focused Ultrasound systems
    Anastasia Antoniou
    Christakis Damianou
    Journal of Ultrasound, 2023, 26 : 401 - 408
  • [22] Simple, inexpensive, and ergonomic phantom for quality assurance control of MRI guided Focused Ultrasound systems
    Antoniou, Anastasia
    Damianou, Christakis
    JOURNAL OF ULTRASOUND, 2023, 26 (02) : 401 - 408
  • [23] An anthropomorphic 3D printed inhomogeneity thorax phantom slab for SBRT commissioning and quality assurance
    How, Stephen
    Banjade, Dilli
    Crowe, Scott
    Dillon, Greg
    Skimmings, Andrew
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (02) : 575 - 583
  • [24] An anthropomorphic 3D printed inhomogeneity thorax phantom slab for SBRT commissioning and quality assurance
    Stephen How
    Dilli Banjade
    Scott Crowe
    Greg Dillon
    Andrew Skimmings
    Physical and Engineering Sciences in Medicine, 2023, 46 : 575 - 583
  • [25] Simple phantom fabrication for MRI-based HDR brachytherapy applicator commissioning
    Fagerstrom, Jessica M.
    Kaur, Sukhjit
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 283 - 287
  • [26] An MRI-based mid-ventilation approach for radiotherapy of the liver
    van de Lindt, Tessa N.
    Schubert, Gerald
    van der Heide, Uulke A.
    Sonke, Jan-Jakob
    RADIOTHERAPY AND ONCOLOGY, 2016, 121 (02) : 276 - 280
  • [27] Development of a Novel Breast MRI Phantom for Quality Control
    Tuong, Betty
    Gardiner, Ian
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 201 (03) : W511 - W515
  • [28] Clinical implementation of MRI-based organs-at-risk auto-segmentation with convolutional networks for prostate radiotherapy
    Savenije, Mark H. F.
    Maspero, Matteo
    Sikkes, Gonda G.
    van der Voort van Zyp, Jochem R. N.
    T. J. Kotte, Alexis N.
    Bol, Gijsbert H.
    T. van den Berg, Cornelis A.
    RADIATION ONCOLOGY, 2020, 15 (01)
  • [29] Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy
    Letourneau, Daniel
    Publicover, Julia
    Kozelka, Jakub
    Moseley, Douglas J.
    Jaffray, David A.
    MEDICAL PHYSICS, 2009, 36 (05) : 1813 - 1821
  • [30] Quality Assurance based on Deep Learning for Pelvic OARs Delineation in Radiotherapy
    Yu, Hang
    He, Yisong
    Fu, Yuchuan
    Li, Xia
    Zhang, Jun
    Liu, Huan
    CURRENT MEDICAL IMAGING, 2023, 19 (04) : 373 - 381