A novel anthropomorphic multimodality phantom for MRI-based radiotherapy quality assurance testing

被引:21
|
作者
Singhrao, Kamal [1 ]
Fu, Jie [1 ]
Wu, Holden H. [2 ]
Hu, Peng [2 ]
Kishan, Amar U. [1 ]
Chin, Robert K. [1 ]
Lewis, John H. [3 ]
机构
[1] Univ Calif Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Radiol, Los Angeles, CA 90095 USA
[3] Cedars Sinai Med Ctr, Dept Radiat Oncol, Los Angeles, CA 90048 USA
关键词
anthropomorphic; MRI; multimodality; phantom; QA; radiotherapy; RADIATION-THERAPY; RELAXATION-TIMES; PROSTATE RADIOTHERAPY; IN-VIVO; CT; TISSUE; SYSTEM; SCANS; 1.5T;
D O I
10.1002/mp.14027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Increased utilization of magnetic resonance imaging (MRI) in radiotherapy has caused a growing need for phantoms that provide tissue-like contrast in both computed tomography (CT) and MRI images. Such phantoms can be used to compare MRI-based processes with CT-based clinical standards. Here, we develop and demonstrate the clinical utility of a three-dimensional (3D)-printed anthropomorphic pelvis phantom containing materials capable of T-1, T-2, and electron density matching for a clinically relevant set of soft tissues and bone. Methods The phantom design was based on a male pelvic anatomy template with thin boundaries separating tissue types. Slots were included to allow insertion of various dosimeters. The phantom structure was created using a 3D printer. The tissue compartments were filled with carrageenan-based materials designed to match the T-1 and T-2 relaxation times and electron densities of the corresponding tissues. CT and MRI images of the phantom were acquired and used to compare phantom T-1 and T-2 relaxation times and electron densities to literature-reported values for human tissue. To demonstrate clinical utility, the phantom was used for end-to-end testing of an MRI-only treatment simulation and planning workflow. Based on a T-2-weighted MRI image, synthetic CT (sCT) images were created using a statistical decomposition algorithm (MRIPlanner, Spectronic Research AB, Sweden) and used for dose calculation of volumetric-modulated arc therapy (VMAT) and seven-field intensity-modulated radiation therapy (IMRT) prostate plans. The plans were delivered on a Truebeam STX (Varian Medical Systems, Palo Alto, CA), with film and a 0.3 cc ion chamber used to measure the delivered dose. Doses calculated on the CT and sCTs were compared using common dose volume histogram metrics. Results T-1 and T-2 relaxation time and electron density measurements for the muscle, prostate, and bone agreed well with literature-reported in vivo measurements. Film analysis resulted in a 99.7% gamma pass rate (3.0%, 3.0 mm) for both plans. The ion chamber-measured dose discrepancies at the isocenter were 0.36% and 1.67% for the IMRT and VMAT plans, respectively. The differences in PTV D98% and D95% between plans calculated on the CT and 1.5T/3.0 T-derived sCT images were under 3%. Conclusion The developed phantom provides tissue-like contrast on MRI and CT and can be used to validate MRI-based processes through comparison with standard CT-based processes.
引用
收藏
页码:1443 / 1451
页数:9
相关论文
共 50 条
  • [1] A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy
    Alzahrani, Meshal
    Broadbent, David A.
    Teh, Irvin
    Al-Qaisieh, Bashar
    Johnstone, Emily
    Speight, Richard
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2025, 33
  • [2] Reproducible phantom for quality assurance in abdominal MRI focussing kidney imaging
    Wolf, Marcos
    Kommer, Stefan
    Fembek, Sebastian
    Droeszler, Uwe
    Koerner, Tito
    Berg, Andreas
    Schmid, Albrecht Ingo
    Moser, Ewald
    Meyerspeer, Martin
    FRONTIERS IN PHYSICS, 2022, 10
  • [3] Assessing suitability and stability of materials for a head and neck anthropomorphic multimodality (MRI/CT) phantoms for radiotherapy
    Alzahrani, Meshal
    Broadbent, David A.
    Teh, Irvin
    Al-Qaisieh, Bashar
    Speight, Richard
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (21):
  • [4] A programmable motion phantom for quality assurance of motion management in radiotherapy
    L. Dunn
    T. Kron
    P. N. Johnston
    L. N. McDermott
    M. L. Taylor
    J. Callahan
    R. D. Franich
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 93 - 100
  • [5] A programmable motion phantom for quality assurance of motion management in radiotherapy
    Dunn, L.
    Kron, T.
    Johnston, P. N.
    McDermott, L. N.
    Taylor, M. L.
    Callahan, J.
    Franich, R. D.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 93 - 100
  • [6] Quality assurance evaluation of spot scanning beam proton therapy with an anthropomorphic prostate phantom
    Iqbal, K.
    Gillin, M.
    Summers, P. A.
    Dhanesar, S.
    Gifford, K. A.
    Buzdar, S. A.
    BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1031):
  • [7] Design of an anthropomorphic intensity modulated radiation therapy quality assurance phantom
    Radford, DA
    Followill, DS
    Hanson, WF
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 197 - 200
  • [8] Preliminary study for a water-paraffin based phantom in MRI quality assurance test
    Lupi, Giulia
    Scorza, Andrea
    Rugiano, Maria Livia
    Sciuto, Salvatore Andrea
    Bini, Fabiano
    2015 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA) PROCEEDINGS, 2015, : 308 - 313
  • [9] Dosimetric study of an indigenous and heterogeneous pelvic phantom for radiotherapy quality assurance
    Singh S.
    Raina P.
    Gurjar O.P.
    Iranian Journal of Medical Physics, 2020, 17 (02) : 120 - 125
  • [10] Quality assurance of clinical MRI scanners using ACR MRI phantom: Preliminary results
    Chen, CC
    Wan, YL
    Wai, YY
    Liu, HL
    JOURNAL OF DIGITAL IMAGING, 2004, 17 (04) : 279 - 284