Simultaneous quantification of fentanyl, sufentanil, cefazolin, doxapram and keto-doxapram in plasma using liquid chromatography-tandem mass spectrometry
被引:14
|
作者:
Flint, Robert B.
论文数: 0引用数: 0
h-index: 0
机构:
Erasmus Univ, Med Ctr, Dept Pharm, Rotterdam, Netherlands
Erasmus Univ, Med Ctr Sophia, Dept Pediat, Div Neonatol, Rotterdam, Netherlands
Dept Pharm, Nijmegen, Netherlands
Radboud Inst Hlth Sci, Nijmegen, NetherlandsErasmus Univ, Med Ctr, Dept Pharm, Rotterdam, Netherlands
Flint, Robert B.
[1
,2
,3
,4
]
Bahmany, Soma
论文数: 0引用数: 0
h-index: 0
机构:
Erasmus Univ, Med Ctr, Dept Pharm, Rotterdam, NetherlandsErasmus Univ, Med Ctr, Dept Pharm, Rotterdam, Netherlands
Bahmany, Soma
[1
]
van der Nagel, Bart C. H.
论文数: 0引用数: 0
h-index: 0
机构:
Erasmus Univ, Med Ctr, Dept Pharm, Rotterdam, NetherlandsErasmus Univ, Med Ctr, Dept Pharm, Rotterdam, Netherlands
van der Nagel, Bart C. H.
[1
]
Koch, Birgit C. P.
论文数: 0引用数: 0
h-index: 0
机构:
Erasmus Univ, Med Ctr, Dept Pharm, Rotterdam, NetherlandsErasmus Univ, Med Ctr, Dept Pharm, Rotterdam, Netherlands
Koch, Birgit C. P.
[1
]
机构:
[1] Erasmus Univ, Med Ctr, Dept Pharm, Rotterdam, Netherlands
[2] Erasmus Univ, Med Ctr Sophia, Dept Pediat, Div Neonatol, Rotterdam, Netherlands
A simple and specific UPLC-MS/MS method was developed and validated for simultaneous quantification of fentanyl, sufentanil, cefazolin, doxapram and its active metabolite keto-doxapram. The internal standard was fentanyl-d5 for all analytes. Chromatographic separation was achieved with a reversed-phase Acquity UPLC HSS T3 column with a run-time of only 5.0 min per injected sample. Gradient elution was performed with a mobile phase consisting of ammonium acetate or formic acid in Milli-Q ultrapure water or in methanol with a total flow rate of 0.4 mL min(-1). A plasma volume of only 50 mu L was required to achieve adequate accuracy and precision. Calibration curves of all five analytes were linear. All analytes were stable for at least 48 h in the autosampler. The method was validated according to US Food and Drug Administration guidelines. This method allows quantification of fentanyl, sufentanil, cefazolin, doxapram and keto-doxapram, which is useful for research as well as therapeutic drug monitoring, if applicable. The strength of this method is the combination of a small sample volume, a short run-time, a deuterated internal standard, an easy sample preparation method and the ability to simultaneously quantify all analytes in one run.