Low-picomolar, label-free procalcitonin analytical detection with an electrolyte-gated organic field-effect transistor based electronic immunosensor

被引:92
作者
Seshadri, Preethi [1 ]
Manoli, Kyriaki [1 ]
Schneiderhan-Marra, Nicole [2 ]
Anthes, Uwe [3 ]
Wierzchowiec, Piotr [3 ]
Bonrad, Klaus [3 ]
Di Franco, Cinzia [4 ]
Torsi, Luisa [1 ,5 ]
机构
[1] Univ Bari A Moro, Dipartimento Chim, Via Orabona 4, I-70125 Bari, Italy
[2] Univ Tubingen, Nat & Med Sci Inst NMI, Markwiesenstr 55, D-72770 Reutlingen, Germany
[3] Merck KGaA, Frankfurter Str 250, D-64271 Darmstadt, Germany
[4] CNR, Ist Foton & Nanotecnol, Sede Bari, Bari, Italy
[5] Abo Akad Univ, Fac Sci & Engn, Biskopsgatan 8 Abo, SF-20500 Turku, Finland
关键词
Electrolyte-gated organic field-effect transistors; Immunosensor; Procalcitonin; Label-free detection; POINT; SENSORS; ANTIBODIES; BIOSENSOR;
D O I
10.1016/j.bios.2017.12.041
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Herein a label-free immunosensor based on electrolyte-gated organic field-effect transistor (EGOFET) was developed for the detection of procalcitonin (PCT), a sepsis marker. Antibodies specific to PCT were immobilized on the poly-3-hexylthiophene (P3HT) organic semiconductor surface through direct physical adsorption followed by a post-treatment with bovine serum albumin (BSA) which served as the blocking agent to prevent nonspecific adsorption. Antibodies together with BSA (forming the whole biorecognition layer) served to selectively capture the procalcitonin target analyte. The entire immunosensor fabrication process was fast, requiring overall 45 min to be completed before analyte sensing. The EGOFET immunosensor showed excellent electrical properties, comparable to those of bare P3HT based EGOFET confirming reliable biosensing with bio-functional EGOFET immunosensor. The detection limit of the immunosensor was as low as 2.2 pM and within a range of clinical relevance. The relative standard deviation of the individual calibration data points, measured on immunosensors fabricated on different chips (reproducibility error) was below 7%. The developed immunosensor showed high selectivity to the PCT analyte which was evident through control experiments. This report of PCT detection is first of its kind among the electronic sensors based on EGOFETs. The developed sensor is versatile and compatible with low-cost fabrication techniques.
引用
收藏
页码:113 / 119
页数:7
相关论文
共 44 条
  • [1] Structural and Functional Characteristics of Chimeric Avidins Physically Adsorbed onto Functionalized Polythiophene Thin Films
    Albers, Willem M.
    Pelto, Jani M.
    Suspene, Clement
    Maatta, Juha A.
    Yassar, Abderrahim
    Hytonen, Vesa P.
    Vikholm-Lundin, Inger M.
    Tappura, Kirsi
    [J]. ACS APPLIED MATERIALS & INTERFACES, 2012, 4 (08) : 4067 - 4077
  • [2] Effects of Polythiophene Surface Structure on Adsorption and Conformation of Bovine Serum Albumin: A Multivariate and Multitechnique Study
    Awsiuk, K.
    Budkowski, A.
    Marzec, M. M.
    Petrou, P.
    Rysz, J.
    Bernasik, A.
    [J]. LANGMUIR, 2014, 30 (46) : 13925 - 13933
  • [3] A new procalcitonin optical immunosensor for POCT applications
    Baldini, F.
    Bolzoni, L.
    Giannetti, A.
    Kess, M.
    Kraemer, P. M.
    Kremmer, E.
    Porro, G.
    Senesi, F.
    Trono, C.
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 393 (04) : 1183 - 1190
  • [4] Baldini F., 2009, INT SOC OPTICS PHOTO
  • [5] Water-gated organic field effect transistors - opportunities for biochemical sensing and extracellular signal transduction
    Cramer, T.
    Campana, A.
    Leonardi, F.
    Casalini, S.
    Kyndiah, A.
    Murgia, M.
    Biscarini, F.
    [J]. JOURNAL OF MATERIALS CHEMISTRY B, 2013, 1 (31) : 3728 - 3741
  • [6] Dinelli Franco, 2005, J PHYS CHEM B
  • [7] Fortunato A., 2016, PRACT LAB MED, V6, P1, DOI DOI 10.1016/j.plabm.2016.06.002
  • [8] Point of Care Diagnostics: Status and Future
    Gubala, Vladimir
    Harris, Leanne F.
    Ricco, Antonio J.
    Tan, Ming X.
    Williams, David E.
    [J]. ANALYTICAL CHEMISTRY, 2012, 84 (02) : 487 - 515
  • [9] A practical guide to analytical method validation, including measurement uncertainty and accuracy profiles
    Gustavo Gonzalez, A.
    Angeles Herrador, M.
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2007, 26 (03) : 227 - 238
  • [10] Recent advances in immobilization methods of antibodies on solid supports
    Jung, Yongwon
    Jeong, Jin Young
    Chung, Bong Hyun
    [J]. ANALYST, 2008, 133 (06) : 697 - 701