Effect of sutureless implantation of the Perceval S aortic valve bioprosthesis on intraoperative and early postoperative outcomes

被引:145
作者
Flameng, Willem [1 ]
Herregods, Marie-Christine [1 ]
Hermans, Hadewich [1 ]
Van der Mieren, Gerry [1 ]
Vercalsteren, Monique [1 ]
Poortmans, Gert [1 ]
Van Hemelrijck, Jan [1 ]
Meuris, Bart [1 ]
机构
[1] Katholieke Univ Leuven, Dept Cardiovasc Dis, Louvain, Belgium
关键词
HIGH-RISK PATIENTS; REPLACEMENT; STENOSIS; PROSTHESIS;
D O I
10.1016/j.jtcvs.2011.02.021
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Prolonged aortic crossclamping can increase mortality and morbidity after aortic valve replacement in elderly and high-risk patients. Sutureless implantation of the prosthesis has the potential to shorten aortic crossclamp time. Methods: The Perceval S valve (Sorin Biomedica Cardio Srl, Sallugia, Italy), a sutureless implantable aortic bioprosthesis, was used in 32 patients (median age, 78 years; median logistic euroSCORE, 9.99) requiring aortic valve replacement with or without concomitant coronary artery bypass grafting. Hemodynamic parameters and clinical outcome were obtained at discharge, at 6 months, and up to 1 year postoperatively. Results: Aortic crossclamp time needed for aortic valve replacement was 18 +/- 6 minutes. Hemodynamics at discharge showed good function of all Perceval S valves with low transvalvular pressure gradients (mean, 12 +/- 5 mm Hg and peak, 23 +/- 9 mm Hg) and low incidence of paravalvular or valvular leakage. Operative mortality was 0%. Follow-up at 1 year showed 3 non-valve-related deaths. Survivors showed good clinical outcome and stable hemodynamic function of the valve prosthesis, except for 1 patient in whom endocarditis developed. Despite a moderate decrease in platelet counts persisting up to 12 months, freedom of bleeding and thromboembolic events was 100%. Conclusions: It is possible to implant a well-functioning sutureless stent-mounted valve in the aortic position in less than 20 minutes of aortic crossclamping. This is associated with excellent early clinical and hemodynamic outcome in high-risk patients. Moderate changes in hematologic parameters persisted but were not related to clinical events. (J Thorac Cardiovasc Surg 2011;142:1453-7)
引用
收藏
页码:1453 / 1457
页数:5
相关论文
共 11 条
[1]   Clinical experience with the second-generation 3f Enable sutureless aortic valve prosthesis [J].
Aymard, Thierry ;
Kadner, Alexander ;
Walpoth, Nazan ;
Goeber, Volkhart ;
Englberger, Lars ;
Stalder, Mario ;
Eckstein, Friedrich ;
Zobrist, Claudia ;
Carrel, Thierry .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 140 (02) :313-316
[2]   Sutureless aortic valve replacement with the Trilogy Aortic Valve System: Multicenter experience [J].
Breitenbach, Ingo ;
Wimmer-Greinecker, Gerhard ;
Bockeria, Leo A. ;
Sadowski, Jerzy ;
Schmitz, Christoph ;
Kapelak, Boguslaw ;
Bartus, Krzysztof ;
Muratov, Ravil ;
Harringer, Wolfgang .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 140 (04) :878-+
[3]   Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[4]   Determinants of early and late results of combined valve operations and coronary artery bypass grafting [J].
Flameng, WJ ;
Herijgers, P ;
Szecsi, J ;
Sergeant, PT ;
Daenen, WJ ;
Scheys, I .
ANNALS OF THORACIC SURGERY, 1996, 61 (02) :621-628
[5]  
Kallenbach K, 2009, HERZ, V34, P130, DOI 10.1007/s00059-009-3201-1
[6]   Sutureless Aortic Valve Replacement With the 3f Enable Aortic Bioprosthesis [J].
Martens, Sven ;
Ploss, Anja ;
Sirat, Sami ;
Miskovic, Alexandra ;
Moritz, Anton ;
Doss, Mirko .
ANNALS OF THORACIC SURGERY, 2009, 87 (06) :1914-1918
[7]   Thrombocytopenia After Aortic Valve Replacement With Freedom Solo Bioprosthesis: A Propensity Study [J].
Piccardo, Alessandro ;
Rusinaru, Dan ;
Petitprez, Benoit ;
Marticho, Paul ;
Vaida, Ioana ;
Tribouilloy, Christophe ;
Caus, Thierry .
ANNALS OF THORACIC SURGERY, 2010, 89 (05) :1425-1431
[8]  
Shrestha M, 2009, J HEART VALVE DIS, V18, P698
[9]   Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI) [J].
Vahanian, Alec ;
Alfieri, Ottavio R. ;
At-Attar, Nawwar ;
Antunes, Manuel J. ;
Bax, Jeroen ;
Cormier, Bertrand ;
Cribier, Alain ;
De Jaegere, Peter ;
Fournial, Gerard ;
Kappetein, Arie Pieter ;
Kovac, Jan ;
Ludgate, Susanne ;
Maisano, Francesco ;
Moat, Neil ;
Mohr, Friedrich -Wihelm ;
Nataf, Patrick ;
Pierard, Luc ;
Pomar, Jose Luis ;
Schofer, Joachim ;
Tornos, Pilar ;
Tuzcu, Murat ;
van Hout, Ben ;
von Segesser, Ludwig K. ;
Walther, Thomas .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 34 (01) :1-8
[10]   Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis [J].
Webb, John G. ;
Pasupati, Sanjeevan ;
Humphries, Karin ;
Thompson, Christopher ;
Altwegg, Lukas ;
Moss, Robert ;
Sinhal, Ajay ;
Carere, Ronald G. ;
Munt, Brad ;
Ricci, Donald ;
Ye, Jian ;
Cheung, Anson ;
Lichtenstein, Sam V. .
CIRCULATION, 2007, 116 (07) :755-763