A 3.7 MHz phased array probe using 0.91Pb(Zn1/3Nb2/3)O3-0.09PbTiO3 single crystal

被引:59
作者
Saitoh, S [1 ]
Takeuchi, T
Kobayashi, T
Harada, K
Shimanuki, S
Yamashita, Y
机构
[1] Toshiba Corp, Ctr Res & Dev, Mat Lab, Kawasaki, Kanagawa 2108582, Japan
[2] Toshiba Corp, Ctr Res & Dev, Device Res Lab, Kawasaki, Kanagawa 2108582, Japan
[3] Toshiba Corp, Nasu Works, Otawara 3248850, Japan
关键词
D O I
10.1109/58.753031
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A novel 128-channel phased array probe for echocardiography with a center frequency of 3.7 MHz using 0.91Pb(Zn1/3Nb2/3)O-3-0.09PbTiO(3) (PZN-9%PT) single crystal has been fabricated to realize greater sensitivity and broader bandwidth properties. The echo amplitude of the PZN-9%PT single-crystal probe is about 5 dB higher than that of the conventional lead zirconate titanate (PZT) ceramic probe, and the fractional bandwidth is about 25 percentage points broader. The quality of B mode images obtained by the PZN-9%PT probe satisfies the performance of the two types of conventional PZT ceramic probes that have center frequencies of 2.5 and 3.75 MHz. At the reference frequency of 3 MHz, the Doppler sensitivity of the PZN-9%PT probe is about 5 dB higher than that of the 3.75 MHz PZT probe; the blood flow of a pulmonary vein in a hard-to-image patient is much more clearly imaged than in the case of using the PZT probe. These superior images are attributable to the use of sufficiently large PZN-9%PT single crystals obtained by the serf-flux method.