On-line rheological characterization of semi-solid formulations

被引:13
作者
Qwist, Pernille Kjaergaard [1 ,2 ]
Sander, Camilla [2 ]
Okkels, Fridolin [3 ]
Jessen, Vibeke [2 ]
Baldursdottir, Stefania [1 ,4 ]
Rantanen, Jukka [1 ]
机构
[1] Univ Copenhagen, Fac Hlth & Med Sci, Dept Pharm, Univ Pk 2, Copenhagen, Denmark
[2] LEO Pharma AS, Ind Pk 55, Ballerup, Denmark
[3] Fluidan ApS, Diplomvej Bldg 381, Lyngby, Denmark
[4] Prof Claraplein 14, Antwerp, Belgium
关键词
Rheology; On-line; Hydroxyethyl cellulose gel; Semi-solid; Process analytical technology (PAT); Pressure difference apparatus (PDA); POLYMER MELTS; SPECTROSCOPY; FLOW; VISCOMETER; VISCOSITY; BATCH; TIME;
D O I
10.1016/j.ejps.2018.11.014
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The rheological profile of a semi-solid product is a critical quality attribute. To monitor changes of this attribute during manufacturing, it would be beneficial to measure the rheological parameters in an on-line or in-line mode and implement this as a part of a control strategy for manufacturing of semi-solids. None of the process analytical technology (PAT) tools for measuring the rheological parameters have yet been widely accepted in the pharmaceutical area, as most of the equipment can only measure viscosity. Therefore, an automated system based on the measurement of pressure difference across both a topology optimized channel and a tube geometry (capillary viscometer) was investigated. The Pressure Difference Apparatus (PDA) can sample from the bulk intermediate/product stream and press the sample through the apparatus at different flow rates to yield a frequency sweep (G' and G '') and a flow curve (viscosity). A calibration model was successfully prepared and verified with hydroxyethyl cellulose gels with polymer content varying from 1.0 to 1.5% (w/w) resulting in gels of different viscosities. The calibration model was used on-line during manufacturing of a gel and manufacturing changes related to dilution of the product were clearly reflected in the batch evolution profiles. The measurements with the PDA reflected the shear rate and frequency ranges relevant for manufacturing and thereby complemented the rheology measurements obtained with a standard rheometer with real time data.
引用
收藏
页码:36 / 42
页数:7
相关论文
共 32 条
[21]  
Okkels F., 2012, A flow measurement device and method Pending, Patent No. [EP 2 724 142 B1, 2724142]
[22]  
Okkels F., 2017, ANN T, V25
[23]   Application of in-line viscometer for in-process monitoring of microcrystalline cellulose-carboxymethylcellulose hydrogel formation during batch manufacturing [J].
Pu, Yu ;
Chaudhry, Saeed ;
Parikh, Maider ;
Berry, Julianne .
DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 2015, 41 (01) :28-34
[24]   Real-time monitoring of active ingredient dispersion in a pharmaceutical aqueous suspension using Raman spectroscopy [J].
Pu, Yu ;
Medendorp, Joseph P. ;
Yonzon, Chanda R. .
JOURNAL OF RAMAN SPECTROSCOPY, 2011, 42 (11) :1994-1999
[25]  
Rao M.A., 2014, MEASUREMENT FLOW VIS, P63
[26]   Introducing Diffusing Wave Spectroscopy as a Process Analytical Tool for Pharmaceutical Emulsion Manufacturing [J].
Reufer, Mathias ;
Machado, Alexandra H. E. ;
Niederquell, Andreas ;
Bohnenblust, Katharina ;
Mueller, Beat ;
Voelker, Andreas Charles ;
Kuentz, Martin .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2014, 103 (12) :3902-3913
[27]   Embedded Doppler System for Industrial In-Line Rheometry [J].
Ricci, Stefano ;
Liard, Maxime ;
Birkhofer, Beat ;
Lootens, Didier ;
Bruehwiler, Armin ;
Tortoli, Piero .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2012, 59 (07) :1395-1401
[28]  
Roberts I., 2003, Texture in food. Volume 1: Semi-solid foods, P161, DOI 10.1533/9781855737082.1.161
[29]   Quality by Design Approach of a Pharmaceutical Gel Manufacturing Process, Part 2: Near Infrared Monitoring of Composition and Physical Parameters [J].
Rosas, Juan G. ;
Blanco, Marcel ;
Gonzalez, Josep M. ;
Alcala, Manel .
JOURNAL OF PHARMACEUTICAL SCIENCES, 2011, 100 (10) :4442-4451
[30]  
Shanley A., 2016, PHARM TECHNOL, P26