Safety and Feasibility Study of a Novel Stent-Graft for Thoracic Endovascular Aortic Repair: a Canine Model Experiment

被引:1
|
作者
Yang, Fan [1 ]
Qiu, Jiehua [2 ]
Fu, Zeliang [3 ]
Qiu, Yuchen [3 ]
Luo, Junrong [1 ]
Xiao, Qingyang [1 ]
Cao, Huabin [1 ]
机构
[1] Jiangxi Agr Univ, Coll Anim Sci & Technol, Inst Anim Populat Hlth, Nanchang, Jiangxi, Peoples R China
[2] Nanchang Univ, Affiliated Hosp 2, Dept Vasc Surg, Nanchang, Jiangxi, Peoples R China
[3] APT Med Inc, Shenzhen, Hunan, Peoples R China
关键词
Stent-graft; Thoracic Endovascular Aortic Repair; Dogs; Models; Animal; Safety; Feasibility Studies; Aortic Diseases; Endovascular Procedures; Stents; DISSECTION; ANEURYSMS; ARCH;
D O I
10.21470/1678-9741-2016-0058
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To evaluate the safety and feasibility of a novel stent-graft for thoracic endovascular aortic repair (TEVAR) in a canine model, 9 adult hybrid dogs were used for the experiment. Methods: All animals were implanted with a novel thoracic aortic stent-graft via femoral artery. Blood sample was collected at pre-operation and 1, 2, 4, 8 and 12 weeks after implantation for hematological examination. Moreover, tissues from randomly selected 4 dogs were subjected to histopathological analysis with the optical microscope after stent-grafts were implanted for 3, 6, 9, and 12 months respectively. The experimental period lasted for more than 2 years. Results: A total of 9 stent-grafts were successfully implanted in the canine thoracic aortas and no migration or deformation occurred. Related indicators of blood routine, inflammatory factors, and immunology changes were not significantly (P>0.05), except the white blood cell (WBC) counts in the first week. Moreover, abnormal morphology was not found in all thoracic aortas via histopathological examination. Additionally, all stent-grafts were patent and did not migrate, and there was no thrombus in the lumens of stent-grafts. Conclusion: The novel thoracic aortic stent-graft made in China was safe and feasible for thoracic endovascular aortic repair in a canine model.
引用
收藏
页码:401 / 407
页数:7
相关论文
共 50 条
  • [1] Type A dissection following endovascular thoracic aortic stent-graft repair
    Neuhauser, B
    Czermak, BV
    Fish, J
    Perkmann, R
    Jaschke, W
    Chemelli, A
    Fraedrich, G
    JOURNAL OF ENDOVASCULAR THERAPY, 2005, 12 (01) : 74 - 81
  • [2] Homemade fenestrated stent-graft for thoracic endovascular aortic repair of zone 2 aortic lesions
    Canaud, Ludovic
    Morishita, Kiyofumi
    Gandet, Thomas
    Sfeir, Julien
    Bommart, Sebastien
    Alric, Pierre
    Mandelli, Marcello
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2018, 155 (02) : 488 - 493
  • [3] Clinical safety and efficacy of a next-generation stent-graft device for thoracic endovascular aortic repair
    Eggebrecht, Holger
    Schmermund, Axel
    Voigtlaender, Thomas
    EXPERT REVIEW OF CARDIOVASCULAR THERAPY, 2013, 11 (01) : 27 - 30
  • [4] Thoracic endovascular stent-graft therapy in aortic dissection
    Akin, Ibrahim
    Kische, Stephan
    Rehders, Tim C.
    Ince, Hueseyin
    Nienaber, Christoph A.
    CURRENT OPINION IN CARDIOLOGY, 2010, 25 (06) : 552 - 559
  • [5] Endovascular stent-graft repair of ruptured aortic aneurysms
    Scharrer-Pamler, R
    Kotsis, T
    Kapfer, X
    Görich, J
    Sunder-Plassmann, L
    JOURNAL OF ENDOVASCULAR THERAPY, 2003, 10 (03) : 447 - 452
  • [6] Hemodynamic effects of stent-graft introducer sheath during thoracic endovascular aortic repair
    Qiao, Yonghui
    Mao, Le
    Wang, Yan
    Luan, Jingyang
    Chen, Yanlu
    Zhu, Ting
    Luo, Kun
    Fan, Jianren
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2022, 21 (02) : 419 - 431
  • [7] In vitro quantification of stent-graft behaviour during chimney thoracic endovascular aortic repair
    Kwiecinski, Jakub
    Uberoi, Raman
    Hadi, Mohammed
    Cheng, Christopher P.
    You, Zhong
    VASCULAR, 2024,
  • [8] Descending thoracic aortic diseases: Stent-graft repair
    Fattori, R
    Napoli, G
    Lovato, L
    Grazia, C
    Piva, T
    Rocchi, G
    Angeli, E
    Di Bartolomeo, R
    Gavelli, G
    RADIOLOGY, 2003, 229 (01) : 176 - 183
  • [9] Hemodynamic effects of stent-graft introducer sheath during thoracic endovascular aortic repair
    Yonghui Qiao
    Le Mao
    Yan Wang
    Jingyang Luan
    Yanlu Chen
    Ting Zhu
    Kun Luo
    Jianren Fan
    Biomechanics and Modeling in Mechanobiology, 2022, 21 : 419 - 431
  • [10] Endovascular stent-graft repair of acute thoracic aortic dissection - Early clinical experiences
    Tiesenhausen, K
    Amann, W
    Koch, G
    Hausegger, KA
    Oberwalder, P
    Rigler, B
    THORACIC AND CARDIOVASCULAR SURGEON, 2001, 49 (01) : 16 - 20