A contactless conductivity detection (CCD) system is used for capillary zone electrophoresis (CZE) with non-aqueous solvents of the buffering background electrolyte, which exhibit strong UV absorbance below 230 nm. It is found that the CCD characteristics with such solvents (propylene carbonate, NN-dimethylformamide and N,N-dimethylacetamide as examples) is the same as with aqueous solutions: the same signal and noise is obtained for a given electric conductance of the background electrolyte, independent of the kind of the solvent. Therefore CCD enables the extension of the application range to solvents with restricted use for common UV detection in CZE due to their unfavourable or even unfitting optical properties. The applicability of CCD is demonstrated by CZE of aliphatic ammonium compounds in these solvents. (C) 2001 Elsevier Science BV All rights reserved.
引用
收藏
页码:147 / 154
页数:8
相关论文
共 50 条
  • [1] Nonaqueous electrophoresis microchip separations: Conductivity detection in UV-absorbing solvents
    Wang, J
    Pumera, M
    ANALYTICAL CHEMISTRY, 2003, 75 (02) : 341 - 345
  • [2] SEPARATION OF NON UV-ABSORBING CATIONS BY CAPILLARY ELECTROPHORESIS
    BECK, W
    ENGELHARDT, H
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (6-9): : 618 - 621
  • [3] Contactless conductivity detector for microchip capillary electrophoresis
    Pumera, M
    Wang, J
    Opekar, F
    Jelínek, I
    Feldman, J
    Löwe, H
    Hardt, S
    ANALYTICAL CHEMISTRY, 2002, 74 (09) : 1968 - 1971
  • [4] Contactless conductivity detector for capillary electrophoresis chip
    Liu, Xiao-Wei
    Zhang, Hai-Feng
    Shao, Xian-Hui
    Li, Ling
    Peng, Zhi-Cheng
    Nami Jishu yu Jingmi Gongcheng/Nanotechnology and Precision Engineering, 2009, 7 (02): : 123 - 126
  • [5] Contactless conductivity detector array for capillary electrophoresis
    Stojkovic, Marko
    Koenka, Israel Joel
    Thormann, Wolfgang
    Hauser, Peter C.
    ELECTROPHORESIS, 2014, 35 (04) : 482 - 486
  • [6] Application of an external contactless conductivity detector for the analysis of beverages by microchip capillary electrophoresis
    Kubán, P
    Hauser, PC
    ELECTROPHORESIS, 2005, 26 (16) : 3169 - 3178
  • [7] CAPILLARY ZONE ELECTROPHORESIS WITH INDIRECT UV DETECTION APPLYING A UV-ABSORBING COUNTER ION
    COLLET, J
    GAREIL, P
    JOURNAL OF CHROMATOGRAPHY A, 1995, 716 (1-2) : 115 - 122
  • [8] Design of Movable Contactless Conductivity Detector for Microchip Capillary Electrophoresis
    Zhang, Hai-feng
    Liui, Xiao-wei
    Peng, Zhi-cheng
    Wang, Xi-lian
    Wang, Wei
    Chen, Yu-feng
    2009 4TH IEEE INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS, VOLS 1 AND 2, 2009, : 190 - +
  • [9] Integrated Capacitively Coupled Contactless Conductivity Detector for Capillary Electrophoresis
    Wang Lixin
    Fu Chonggang
    PROCEEDINGS OF THE SECOND INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOL. 3, 2008, : 1493 - 1495
  • [10] Analyzing the response of a contactless conductivity detector in capillary electrophoresis by a resonant method
    Kang, Qi
    Zhang, Qing
    Li, Yao Long
    Li, Dong Dong
    Shen, Da Zhong
    CHINESE CHEMICAL LETTERS, 2012, 23 (07) : 831 - 834