Virtual Monochromatic Spectral Imaging with Fast Kilovoltage Switching: Improved Image Quality as Compared with That Obtained with Conventional 120-kVp CT

被引:401
作者
Matsumoto, Kazuhiro [1 ]
Jinzaki, Masahiro [1 ]
Tanami, Yutaka [1 ]
Ueno, Akihisa [1 ]
Yamada, Minoru [2 ]
Kuribayashi, Sachio [1 ]
机构
[1] Keio Univ, Sch Med, Dept Diagnost Radiol, Shinjuku Ku, Tokyo 1608582, Japan
[2] Keio Univ, Sch Med, Multi Dimens Biomed Imaging & Informat Lab Res Pk, Shinjuku Ku, Tokyo 1608582, Japan
关键词
DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; APPARATUS; PHANTOM;
D O I
10.1148/radiol.11100978
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare image quality obtained in phantoms with virtual monochromatic spectral (VMS) imaging with that obtained with conventional 120-kVp computed tomography (CT) for a given radiation dose. Materials and Methods: Three syringes were filled with a diluted contrast medium (each syringe contained a contrast medium with a different iodine concentration [5, 10, or 15 mg of iodine per milliliter]), and a fourth syringe was filled with water. These syringes were placed in a torso phantom meant to simulate the standard human physique. The phantom was examined with a CT system and use of the fast kilovoltage switching (80 and 140 kVp) and conventional (120 kVp) modes. Image noise and contrast-to-noise (CNR) ratio were analyzed on VMS images and 120-kVp CT images. Results: Image noise on VMS images in the range of 67-72 keV was significantly lower than that on the 120-kVp CT images (P < .014). Image noise was lowest at 69 keV and was 12% lower when compared with that on 120-kVp CT images. CNR on the VMS images was highest at 68 keV. CNR on the VMS images obtained at 68 keV in the syringes filled with diluted contrast material (5, 10, and 15 mg of iodine per milliliter) was 28%, 31%, and 30% higher, respectively, compared with that on the 120-kVp CT images (P < .001). Conclusion: VMS imaging at approximately 70 keV yielded lower image noise and higher CNR than did 120-kVp CT for a given radiation dose. VMS imaging has the potential to replace 120-kVp CT as the standard CT imaging modality, since optimal VMS imaging may be expected to yield improved image quality in a patient with standard body habitus. (C) RSNA, 2011
引用
收藏
页码:257 / 262
页数:6
相关论文
共 17 条
  • [1] ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY
    ALVAREZ, RE
    MACOVSKI, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) : 733 - 744
  • [2] Image Fusion in Dual Energy Computed Tomography Effect on Contrast Enhancement, Signal-to-Noise Ratio and Image Quality in Computed Tomography Angiography
    Behrendt, Florian F.
    Schmidt, Bernhard
    Plumhans, Cedric
    Keil, Sebastian
    Woodruff, Seth G.
    Ackermann, Diana
    Muehlenbruch, Georg
    Flohr, Thomas
    Guenther, Rolf W.
    Mahnken, Andreas H.
    [J]. INVESTIGATIVE RADIOLOGY, 2009, 44 (01) : 1 - 6
  • [3] Single- and dual-energy CT with monochromatic synchrotron x-rays
    Dilmanian, FA
    Wu, XY
    Parsons, EC
    Ren, B
    Kress, J
    Button, TM
    Chapman, LD
    Coderre, JA
    Giron, F
    Greenberg, D
    Krus, DJ
    Liang, Z
    Marcovici, S
    Petersen, MJ
    Roque, CT
    Shleifer, M
    Slatkin, DN
    Thomlinson, WC
    Yamamoto, K
    Zhong, Z
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (02) : 371 - 387
  • [4] DILMANIAN FA, 1995, 1994 IEEE NUCL SCI S, V3, P1392
  • [5] GRANILLO GAR, 2010, INT J CARDIOVAS IMAG, V26, P345
  • [6] Dual energy CT characterization of urinary calculi: Initial in vitro and clinical experience
    Graser, Anno
    Johnson, Thorsten R. C.
    Bader, Markus
    Staehler, Michael
    Haseke, Nicolas
    Nikolaou, Konstantin
    Reiser, Maximilian F.
    Stief, Christian G.
    Becker, Christoph R.
    [J]. INVESTIGATIVE RADIOLOGY, 2008, 43 (02) : 112 - 119
  • [7] AN ALGORITHM FOR NOISE SUPPRESSION IN DUAL ENERGY CT MATERIAL DENSITY IMAGES
    KALENDER, WA
    KLOTZ, E
    KOSTARIDOU, L
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (03) : 218 - 224
  • [8] EVALUATION OF A PROTOTYPE DUAL-ENERGY COMPUTED TOMOGRAPHIC APPARATUS .1. PHANTOM STUDIES
    KALENDER, WA
    PERMAN, WH
    VETTER, JR
    KLOTZ, E
    [J]. MEDICAL PHYSICS, 1986, 13 (03) : 334 - 339
  • [9] GENERALIZED IMAGE COMBINATIONS IN DUAL KVP DIGITAL RADIOGRAPHY
    LEHMANN, LA
    ALVAREZ, RE
    MACOVSKI, A
    BRODY, WR
    PELC, NJ
    RIEDERER, SJ
    HALL, AL
    [J]. MEDICAL PHYSICS, 1981, 8 (05) : 659 - 667
  • [10] Reduction of beam-hardening artifacts in X-ray CT
    Menvielle, Nathalie
    Goussard, Yves
    Orban, Dominique
    Soulez, Gilles
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 1865 - 1868